The goal of this clinical trial is to test how accurately a new chest-worn device called the Perin Health Patch measures breathing rate in adults aged 22 and older. The main question it aims to answer is:
How accurately does the Perin Health Patch measure breathing rate compared to a standard medical device that measures carbon dioxide in exhaled breath?
Participants will:
The study will include both healthy adults and people with conditions like asthma, COPD, heart failure, diabetes, high blood pressure, or obesity to test how well the device works for different types of people. Researchers will compare the breathing rate numbers from the new chest patch to the standard medical device to see if the patch is accurate enough for medical use.
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| Chest Worn Respiratory Rate Monitor | Device | The Perin Health Patch is a non-invasive, chest-worn wearable device designed to continuously monitor physiological signals, including respiratory rate and other health metrics. In this study, the device will be evaluated for its ability to accurately measure respiratory rate during controlled breathing exercises in healthy adult volunteers. The respiratory rate values recorded by the Perin Health Patch will be compared against respiratory rate measurements obtained via manually scored end tidal carbon dioxide (EtCO2) waveforms from a GE Datex-Ohmeda monitor, following Element Laboratory protocols aligned with clinical validation standards. The intervention in this study is limited to the application and use of the device for data collection during controlled respiratory rate conditions, where participants perform paced breathing exercises at various rates (5-50 breaths per minute) using a breathing app while positioned sitting or lying supine. |
| Measure | Description | Time Frame |
|---|---|---|
| Accuracy of Perin Health Patch Respiratory Rate Measurements Compared to End Tidal COâ‚‚ Reference | The primary outcome is the accuracy of respiratory rate (RR) values reported by the Perin Health Patch compared to respiratory rate measured by manually scored end tidal carbon dioxide (EtCOâ‚‚) waveforms from a GE Datex-Ohmeda monitor. Accuracy will be assessed using mean absolute error (MAE), accuracy root mean square (ARMS), bias (mean difference), and standard deviation. Measurements will be collected during multiple stable respiratory rate plateaus (approximately 5-50 breaths per minute) achieved through paced breathing exercises. Each comparison will be made during 60-second intervals at steady-state conditions, with the reference EtCOâ‚‚ waveform scored by blinded reviewers counting respiratory peaks per minute. | During a single in-lab session lasting approximately 1 hour per participant |
Not provided
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
The study population will consist of a minimum of 35 adult participants and maximum of 80 adult participants, ages 22 years and older. The study population will be comprised of diversity in race, age, gender, BMI, skin tone, and health conditions. The health conditions will constitute certain proportions of healthy controls (to allow testing over the full range of breaths per minute) and disease comorbidities (to represent comorbidities such as COPD, CHF, Asthma, Diabetes, Hypertension, Diabetes and Obesity). An attempt will be made to enroll a variety of skin tones. The participants must understand the study and consent to participate by signing the Informed Consent Form. Participant enrollment and participation in this clinical study is based on meeting the inclusion criteria and none of the exclusion criteria, a satisfactory health screen, and the participant and data demographics needed for the study.