The purpose of this study is to conduct a Respiratory Rate accuracy validation comparing the RS001.2 Respiree Cardio-Respiratory Monitor to the Reference, an FDA cleared End Tidal Carbon Dioxide monitor (GE Datex-Ohmeda).
After IRB approval, a minimum of 30 volunteer test subjects will be enrolled for the final validation study population. A maximum of 60 subjects will be enrolled. All subjects will be 18 or older and selected to represent a range of body types including small, average, muscular, and large with a range of BMI's.
Each subject will be connected to a commonly used End Tidal Carbon Dioxide monitor (GE Datex-Ohmeda) and the RS001.2 Respiree Cardio-Respiratory Monitor. The end tidal carbon dioxide (EtCO2) monitor will determine performance of respiratory rate metrics (Reference). Each subject will be instrumented with a mouthpiece or nasal cannula that allows for measurement of the EtCO2. Subjects will also be connected to a 3-lead ECG and pulse oximeter in order to monitor the subject's vital signs for safety purposes.
Subjects will be seated for the study. A range of stable respiratory rates will be elicited from each volunteer test subject. After a baseline data set at the subject natural respiratory rate, the rates will be approximately 5, 10, 15, 20, 25, 30, 35, 40, 45, and 50 breaths per minute; with some natural variation from these exact numbers and tailored to the subject's capabilities. A paced breathing app will be used. It is expected that some of the subjects may not be able to be evaluated over the entire range as some subjects may not be able to breathe at the lower and higher respiratory rates. In this case, additional data may be recorded at the baseline range and other ranges that are comfortable for the subject. Once stable breathing at specified rate is achieved data will be collected for one to three minutes per respiratory rate plateau. Between plateaus, the subjects using a mouthpiece are allowed to take the mouthpiece out and relax.
The respiratory rate will be measured simultaneously with the Reference and the Device Under Test. During the stable plateaus at each level, the data will be averaged in approximately 60 second intervals. For final validation, the Reference EtCO2 waveform will be scored by counting the respiratory peaks per minute. The Accuracy root-mean-square (Arms) will be the basis for evaluation and acceptance.
To 'Pass' this test the RS001.2 Respiree Cardio-Respiratory Monitor (Device Under Test) must demonstrate an Accuracy root-mean-square (Arms) of ≤3 breaths per minute when compared to the Reference EtCO2 monitor.
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| RS001.2 Respiree Cardio-Respiratory Monitor | Device | The RS001.2 Respiree Cardio-Respiratory Monitor is intended as a noninvasive multiparameter monitor on the adult population |
| Measure | Description | Time Frame |
|---|---|---|
| Accuracy of the device to Reference | The primary objective of this study is to compare the accuracy of the device under test for the measurement of respiratory rate to the Reference, which is an End Tidal Carbon Dioxide Monitor (EtCO2). For final validation, the EtCO2 waveform will be scored by counting the respiratory peaks per minute. | 1 to 3 minutes at a given level |
Not provided
Not provided
Inclusion Criteria:
Exclusion Criteria:
Subjects who meet any of the following criteria will be excluded from the study:
Subjects who refuse or are unable to provide to sign an informed written consent for study
Subject is evaluated by the Investigator and found to be medically unsuitable for participation in this study
Compromised circulation, injury, or physical malformation of fingers, toes, hands, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study. Tattoo in the optical path which would limit the ability to test sites needed for the study. (Note: Certain malformations may still allow subjects to participate if the condition is noted and would not affect the particular sites utilized.)
Subjects with known respiratory conditions such as:
Subjects with self-reported heart or cardiovascular conditions such as:
Self-reported health conditions as identified in the Health Assessment Form
Subjects with severe contact allergies to standard adhesives, latex or other materials found in pulse oximetry sensors, ECG electrodes, or other medical sensors (self-reported)
Other known health condition, should be considered upon disclosure in health assessment form
Not provided
Not provided
Not provided
Not provided
The study population will include a minimum of 30 healthy competent adults, ages 18 years and older. A maximum of 60 subjects will be enrolled.
The subject selection will be a mix of males and females with small to large physiques. An attempt will be made to enroll a variety of skin tones. The goal is to include a range of skin pigmentations, including at least 5 darkly pigmented subjects or 15% of the subject pool, whichever is larger.
The subjects must understand the study and consent to participate by signing the Informed Consent Form. Subject enrollment and participation in this clinical study is based on meeting the inclusion criteria and none of the exclusion criteria, a satisfactory health screen, and the subject and data demographics needed for the study.