The purpose of this study is to conduct a Respiratory Rate accuracy validation on pediatrics population comparing the Gabi SmartCare Gabi Band to the Reference, an FDA cleared End Tidal Carbon Dioxide monitor (GE Datex- Ohmeda) by manually scoring the collected waveform for data analysis.
Each participant is expected to take 1-2 hours, including paperwork, sensor application, and a maximum of 60 minutes of data collection. Each participant will be connected to a commonly used End Tidal Carbon Dioxide monitor (GE Datex-Ohmeda) and the Gabi SmartCare Gabi Band. The end tidal carbon dioxide (EtCO2) monitor will determine performance of respiratory rate metrics (GE Healthcare S5 Compact Monitor with M-COVX (K001814) or E-CaiO (K051092) modules). Each participant will be instrumented with neonatal or pediatric nasal cannula that allows for measurement of the EtCO2. SpO2 sensors may be placed on the feet and/or toes for safety monitoring purposes.
Data will be simultaneously and continuously recorded from the Reference EtCO2 monitor and the Device Under Test (DUT). Periods of quiet will be encouraged. Data will be marked for stable quiet periods that are useable for analysis. There is no additional follow-up required for the investigation.
One minute epochs will be generated from the stable data periods for analysis. For verification, the Reference EtCO2 waveform will be scored by counting the respiratory peaks per minute.
The reviewers counting the raw EtCO2 waveform will be blinded to the results from the test device. The DUT data will be averaged over the corresponding one minute period. The Accuracy root- mean-square (Arms) will be the basis for evaluation and acceptance.
To 'Pass' this test the Gabi SmartCare Gabi Band (Device Under Test) must demonstrate an Accuracy root-mean- square (Arms) of ≤3.0 breaths per minute when compared to the Reference EtCO2 monitor.
The study population will include a minimum of 23 neonates, infants, and pediatrics from 0-12 years of age. A maximum of 60 participants will be enrolled. The participant selection will be an equitable distribution of males and females of any race with varying skin tones. An attempt will be made to include 3 darkly pigmented participants (Fitzpatrick 5 or 6) or 15% of the participant pool (whichever is larger).
The parents or guardian of the participants must understand the study and consent to their child's participation in the study by signing the Informed Consent Form. Subjects 7 to 12 years of age will be provided an Assent. The study will be explained within his/her ability to understand. Participant enrollment and participation in this clinical study is based on meeting the inclusion criteria and none of the exclusion criteria, a satisfactory screening, and the participant and data demographics needed for the study.
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Pediatrics population | Experimental |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| Respiratory rate measure based on PPG | Device | Testing was conducted under normal office environment conditions. Each subject was connected to the reference EtCO2 monitor (GE Datex-Ohmeda) and the Device Under Test, the Gabi SmartCare Gabi Band. A waveform collected from an EtCO2 measurement was used to determine performance of respiratory rate metrics. Each subject was given a nasal cannula that allowed for measurement of the EtCO2 respiratory rate. The reference respiratory rate was measured simultaneously with the Device Under Test. Data was marked for stable quiet periods that were useable for analysis. |
| Measure | Description | Time Frame |
|---|---|---|
| Accuracy of Device Under Test for the Measurement of Respiratory Rate, Measured in Breath/Min | Arms or rms error, given in breath per minute, is the calculation of root mean square between Respiratory rate of the subject device and the manually scored reference Respiratory Rate EtCO2 waveform, for all stable respiratory periods. It is a way of averaging the absolute values of errors over the full measurement range and is a similar measurement to a "rms error". It indicates how far (accurate) the provided value is from the reference used. A passing result required an Arms of ≤3 breaths/min. | Every 60 seconds, over a total duration of maximum 60 minutes |
Not provided
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| Clinimark LLC | Louisville | Colorado | 80027 | United States |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| FG000 | Pediatrics Population | Respiratory rate measure based on PPG: Testing was conducted under normal office environment conditions. Each subject was connected to the reference EtCO2 monitor (GE Datex-Ohmeda) and the Device Under Test, the Gabi SmartCare Gabi Band. A waveform collected from an EtCO2 measurement was used to determine performance of respiratory rate metrics. Each subject was given a nasal cannula that allowed for measurement of the EtCO2 respiratory rate. The reference respiratory rate was measured simultaneously with the Device Under Test. Data was marked for stable quiet periods that were useable for analysis. |
| Title | Milestones | Reasons Not Completed | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Overall Study |
|
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| BG000 | Pediatrics Population | Respiratory rate measure based on PPG: Testing was conducted under normal office environment conditions. Each subject was connected to the reference EtCO2 monitor (GE Datex-Ohmeda) and the Device Under Test, the Gabi SmartCare Gabi Band. A waveform collected from an EtCO2 measurement was used to determine performance of respiratory rate metrics. Each subject was given a nasal cannula that allowed for measurement of the EtCO2 respiratory rate. The reference respiratory rate was measured simultaneously with the Device Under Test. Data was marked for stable quiet periods that were useable for analysis. |
| Units | Counts |
|---|---|
| Participants |
|
| Title | Description | Population Description | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Denominator Units Selected | Denominators | Classes |
|---|---|---|---|---|---|---|---|---|---|
| Age, Categorical | Count of Participants |
| Type | Title | Description | Population Description | Reporting Status | Anticipated Posting Date | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Time Frame | Units Analyzed | Denominator Units Selected | Arm/Group Information | Denominators | Classes | Analyses | |||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Primary | Accuracy of Device Under Test for the Measurement of Respiratory Rate, Measured in Breath/Min | Arms or rms error, given in breath per minute, is the calculation of root mean square between Respiratory rate of the subject device and the manually scored reference Respiratory Rate EtCO2 waveform, for all stable respiratory periods. It is a way of averaging the absolute values of errors over the full measurement range and is a similar measurement to a "rms error". It indicates how far (accurate) the provided value is from the reference used. A passing result required an Arms of ≤3 breaths/min. | Posted | Number | breaths/min | Every 60 seconds, over a total duration of maximum 60 minutes |
|
During the period of measurement and 15 minutes after, up to 75 minutes
Not provided
Not provided
| ID | Title | Description | Deaths (Affected) | Deaths (At Risk) | Serious Events (Affected) | Serious Events (At Risk) | Other Events (Affected) | Other Events (At Risk) |
|---|---|---|---|---|---|---|---|---|
| EG000 | Pediatrics Population | Respiratory rate measure based on PPG: Testing was conducted under normal office environment conditions. Each subject was connected to the reference EtCO2 monitor (GE Datex-Ohmeda) and the Device Under Test, the Gabi SmartCare Gabi Band. A waveform collected from an EtCO2 measurement was used to determine performance of respiratory rate metrics. Each subject was given a nasal cannula that allowed for measurement of the EtCO2 respiratory rate. The reference respiratory rate was measured simultaneously with the Device Under Test. Data was marked for stable quiet periods that were useable for analysis. |
Not provided
Not provided
Not provided
| Title | Organization | Phone | Extension | |
|---|---|---|---|---|
| Associate Researcher | Clinimark | 833 373 0998 | 1 | Carolyn.Cohen@element.com |
Not provided
| Type | Includes Protocol | Includes SAP | Includes ICF | Document Label | Document Date | Document Uploaded Date | Document File Name |
|---|---|---|---|---|---|---|---|
| Prot | Yes | No | No | Study Protocol | Sep 22, 2022 | Oct 2, 2024 | Prot_000.pdf |
| SAP | No | Yes | No | Statistical Analysis Plan | Nov 21, 2022 | Oct 2, 2024 | SAP_001.pdf |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
|
| Participants |
|
| Sex: Female, Male | Count of Participants | Participants |
|
| Ethnicity (NIH/OMB) | Count of Participants | Participants |
|
| Race (NIH/OMB) | Count of Participants | Participants |
|
| Region of Enrollment | Count of Participants | Participants |
|
|
|
| 0 |
| 13 |
| 0 |
| 13 |
| 0 |
| 13 |
Not provided
Not provided