Purpose of this study was to evaluate the effect of presbyopic correction lens on Humphrey visual field (HVF) testing in patients with 2 models of multifocal intraocular lens (MIOL) both with and without glaucoma. All participants performed HVF testing with presbyopic collection lenses and without presbyopic collection lenses on the same occasion (5-10 minutes between the test). The mean deviation (MD) and pattern standard deviation (PSD) were compared between with and without presbyopic collection lenses, between glaucoma and non-glaucoma and between 2 models of MIOL.
This was a prospective comparative study in patients who underwent uneventful phacoemulsification with MIOL implantation at the Department of Ophthalmology, Faculty of Medicine, Siriraj Hospital, Mahidol University between June 2009 to July 2012. The study protocol was approved by the Siriraj Institutional Review Board complied with all of the principles set forth in the Declaration of Helsinki (1964) and its subsequent amendments. (certificate of approval number Si 003/2010). The written informed consent was obtained from each participant prior to participation. Each patient received a comprehensive eye examination with slit-lamp biomicroscopy, distant and near visual acuity, and auto-refraction. The eligible participants performed HVF 24-2 or 30-2 on the enrollment day with and without presbyopic correction lens. The participants were randomized to perform HVF with presbyopic collection lenses followed by without presbyopic collection lenses or without presbyopic collection lenses followed by with presbyopic collection lenses by simple randomization. Data collection included demographic data, distant (logMar) and near (Jaeger reading chart) BCVA, and HVF test result including mean deviation (MD), pattern standard deviation (PSD), test duration (minute), fixation loss, false positive, and false negative.
Multifocal intraocular lens characteristics There were two types of aspheric diffractive MIOL included in the study: TECNIS ZM900 (Abbott Medical Optics Inc, Santa Ana, CA, USA) and AcrySof IQ ReSTOR SN6AD1 (Alcon Laboratories, Inc., Irvine, CA, USA). The TECNIS ZM900 is a silicone 3-piece aspheric diffractive lens. The power of the add is +4.00 D with a 50/50 distance/near light distribution. The AcrySof IQ ReSTOR SN6AD1is an acrylic single-piece aspheric diffractive lens. The power of the add is +3.00 D with a 65/35 light distribution.
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Glaucomatous arm | Experimental | Mild stage glaucomatous patients age more than 18 years old, a history of uneventful phacoemulsification with MIOL implantation at least 1 month before participation, distant best-corrected visual acuity (BCVA) equal to or better than 20/30, near BCVA at least Jaeger 2. |
|
| Non-glaucomatous arm | Active Comparator | Pseudophakic participants without glaucoma, age more than 18 years old, a history of uneventful phacoemulsification with MIOL implantation at least 1 month before participation, distant best-corrected visual acuity (BCVA) equal to or better than 20/30, near BCVA at least Jaeger 2. |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| Humphrey visual field testing | Diagnostic Test | Perimetry was performed with the Humphrey Field Analyzer (Carl Zeiss Meditec Inc). Participants performed a 30-2 or a 24-2 Swedish Interactive Threshold Algorithm (SITA) standard test with stimulus size III, and standard Humphrey background luminance of 31.5 apostilbs. The participants were selected by simple random sampling to perform the HVF test with presbyopic collection lenses followed by without presbyopic collection lenses, or without presbyopic collection lenses followed by with presbyopic collection lenses. The power of presbyopic correction lens at the testing distance of 33 cm (the bowl radius) was auto-calculated according to the recommendation by the Humphrey Field Analyzer's manufacturer, depending on the current refractive error and the patient's age. |
| Measure | Description | Time Frame |
|---|---|---|
| Mean deviation | Mean deviation from Humphrey visual field testing | at 1 day after enrollment |
| Pattern standard deviation | Pattern standard deviation from Humphrey visual field testing | at 1 day after enrollment |
| Measure | Description | Time Frame |
|---|---|---|
| Fixation loss (percent) | Fixation loss from Humphrey visual field testing | at 1 day after enrollment |
| False negative (percent) | False negative from Humphrey visual field testing |
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Naris Kitnarong, M.D., M.B.A. | Mahidol University | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| Naris Kitnarong | Bangkoknoi | Bangkok | 10700 | Thailand |
| PubMed Identifier | Type | Citation | Retractions |
|---|---|---|---|
| 23430147 | Result | Aychoua N, Junoy Montolio FG, Jansonius NM. Influence of multifocal intraocular lenses on standard automated perimetry test results. JAMA Ophthalmol. 2013 Apr;131(4):481-5. doi: 10.1001/jamaophthalmol.2013.2368. | |
| 24784872 | Result | Farid M, Chak G, Garg S, Steinert RF. Reduction in mean deviation values in automated perimetry in eyes with multifocal compared to monofocal intraocular lens implants. Am J Ophthalmol. 2014 Aug;158(2):227-231.e1. doi: 10.1016/j.ajo.2014.04.017. Epub 2014 Apr 29. |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| ID | Term |
|---|---|
| D005901 | Glaucoma |
| D002386 | Cataract |
| D011305 | Presbyopia |
| ID | Term |
|---|---|
| D009798 | Ocular Hypertension |
| D005128 | Eye Diseases |
| D007905 | Lens Diseases |
| D012030 | Refractive Errors |
Not provided
Not provided
Non-randomized, prospective comparative study
Not provided
Not provided
Not provided
Not provided
|
| at 1 day after enrollment |
| False positive (percent) | False positive from Humphrey visual field testing | at 1 day after enrollment |
| Test duration (minutes0 | Test duration from Humphrey visual field testing | at 1 day after enrollment |
| 28676280 | Result | Yohannan J, Wang J, Brown J, Chauhan BC, Boland MV, Friedman DS, Ramulu PY. Evidence-based Criteria for Assessment of Visual Field Reliability. Ophthalmology. 2017 Nov;124(11):1612-1620. doi: 10.1016/j.ophtha.2017.04.035. Epub 2017 Jul 1. |
| 17159450 | Result | Kumar BV, Phillips RP, Prasad S. Multifocal intraocular lenses in the setting of glaucoma. Curr Opin Ophthalmol. 2007 Feb;18(1):62-6. doi: 10.1097/ICU.0b013e328011d108. |