The purpose of the study is to test the performance and safety of the Bioserenity pulse oxymeter EOXY. For that, and in agreement with ISO60601-2-61 norm and FDA regulation, measures of heart rate, with a gold-standard, and SaO2 measures generated by blood samples will be compared with Sp02 and heart rate measures that are generated by our EOXY.
Thirteen healthy subjects will participate in the study in total. Three subjects will be enrolled during a primary phase .They will participate at a secondary phase as well as ten others subjects. The purpose of the primary phase is to test the body area where the measurement can be done. The purpose of the secondary phase is to measure SpO2 and HR in desaturation conditions.
The primary objective is to evaluate and document the SpO2 accuracy of the pulse oximeter equipment, evaluating the performance of the device compared to invasive method blood sampling (SaO2) by CO-Oximeter in healthy subjects during invasive controlled desaturation study at different ranges (100 % to 70% SaO2).
Subject safety during desaturation and saturation period, will be evaluated by reviewing vital sign: heart rate (HR) continuously and blood pressure/temperature between each desaturation plateau.
There are two secondaries objectives, the measurement of the variation of the fraction of inspired oxygen (FiO2), and the validation of the algorithm for data review and analysis.
For the study primary phase, the purpose of this phase is to test the body area where the SpO2 is better detected and define 3 measurement areas. For this, three healthy volunteers will be placed in normal O2 saturation, the SpO2 will be measured using the Bioserenity oximeter (noninvasive method) and will be compared to a gold-standard oximeter CE-marked. The investigator will determine the best area in the body where the SpO2 could be taken for the following subjects.
During a second phase, the measurement of Sp02 and HR with EOXY, will be performed in O2 saturation conditions up to 100% and then in desaturation by plateau from 100 % to 70%. 3 EOXY pulse oximeter recorded the SpO2 at the same time in 3 different areas of the body. As required by the standard, Sp02 measures will be simultaneously compared to SaO2 values measured by blood sampling taken from an indwelling arterial catheter. The HR obtained with EOXY will be compared simultaneously to a gold standard oximeter CE-marked.
As the example in the standard, it was choosen to do the test on the 70-100% range, with 5 plateaus (100%-97%, 97%-92%, 92%-85%, 84%-78%, 77%-70%) and to collect five Sp02 and Sa02 values (blood samples) and five HR measures, per plateau, for each of 13 subjects.
The desaturation will be controlled thanks' the FiO2 measures, that should allow to bring subjects near target levels of desaturation plateau.
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| EOXY device and Gold standard oximter and SaO2 measures | Experimental | Heart rate and SPO2 measures provided from EOXY device Heart rate measures provided from gold standard oximeter. SaO2 measures provided from blood sampling There is only one arm: all subjects have simultaneously three interventions (as required in the European standard ISO 80601-2-61), to qualify a pulse oximeter |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| EOXY device | Device | Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard |
|
| Measure | Description | Time Frame |
|---|---|---|
| Percentage SPO2 Accuracy Root-Mean Square (ARMS) at Forehead Position | For each desaturation plateau, SpO2 ARMS are calculated by the difference between the investigational pulse oximeter SpO2 measurement and the co-oximeter's SaO2 measurements | 1 day |
| Percentage SPO2 Accuracy Root-Mean Square (ARMS) at Torso Position | For each desaturation plateau, SpO2 ARMS (Accuracy Root-Mean Square) are calculated by the difference between the investigational pulse oximeter SpO2 measurement and the co-oximeter's SaO2 measurements | 1 day |
| Percentage SPO2 Accuracy Root-Mean Square (ARMS) at BACK Position | For each desaturation plateau, SpO2 ARMS are calculated by the difference between the investigational pulse oximeter SpO2 measurement and the co-oximeter's SaO2 measurements | 1 day |
| HR Accuracy Root-Mean Square (ARMS) at Forehead Position | For each desaturation plateau, HR ARMS are calculated by the difference between the investigational pulse oximeter HR measurement and the HR gold standard measurements | 1 day |
| Percentage of HR Accuracy Root-Mean Square (ARMS) at TORSO Position | For each desaturation plateau, HR ARMS are calculated by the difference between the investigational pulse oximeter HR measurement and the HR gold standard measurements | 1 day |
| Percentage of HR Accuracy Root-Mean Square (ARMS) at BACK Position | For each desaturation plateau, HR ARMS are calculated by the difference between the investigational pulse oximeter HR measurement and the HR gold standard measurements |
Not provided
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Affiliation | Role |
|---|---|---|
| marc FRANCAUX | Université Catholique de Louvain | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| Université catholique louvain la neuve, centre sportif | Louvain-la-Neuve | 1348 | Belgium |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| FG000 | EOXY Device and Gold Standard Oximeter and SaO2 Measures | Heart rate and SPO2 measures provided from EOXY device Heart rate measures provided from gold standard oximeter. SaO2 measures provided from blood sampling There is only one arm: all subjects have simultaneously three interventions (as required in the European standard ISO 80601-2-61), to qualify a pulse oximeter EOXY device: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard Gold standard oximeter: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard SaO2 sampling: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard |
| Title | Milestones | Reasons Not Completed | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Overall Study |
|
Healthy subjects (3 patients excluded reducing the number analyzed at 10)
Not provided
| ID | Title | Description |
|---|---|---|
| BG000 | EOXY Device and Gold Standard Oximeter and SaO2 Measures | Heart rate and SPO2 measures provided from EOXY device Heart rate measures provided from gold standard oximeter. SaO2 measures provided from blood sampling There is only one arm: all subjects have simultaneously three interventions (as required in the European standard ISO 80601-2-61), to qualify a pulse oximeter EOXY device: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard Gold standard oximeter: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard SaO2 sampling: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard |
| Units | Counts |
|---|---|
| Participants |
|
| Title | Description | Population Description | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Denominator Units Selected | Denominators | Classes |
|---|---|---|---|---|---|---|---|---|---|
| Age, Categorical | Count of Participants |
| Type | Title | Description | Population Description | Reporting Status | Anticipated Posting Date | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Time Frame | Units Analyzed | Denominator Units Selected | Arm/Group Information | Denominators | Classes | Analyses | |||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Primary | Percentage SPO2 Accuracy Root-Mean Square (ARMS) at Forehead Position | For each desaturation plateau, SpO2 ARMS are calculated by the difference between the investigational pulse oximeter SpO2 measurement and the co-oximeter's SaO2 measurements | Subjects respecting the study conditions (no movements, good irrigation) 3 subjects excluded from analysis(no respect of study conditions) so 10 subjects analyzed | Posted | Number | percentage of ARMS (accuracy) | 1 day |
|
2 days for each subject (days of participation)
Not provided
Not provided
| ID | Title | Description | Deaths (Affected) | Deaths (At Risk) | Serious Events (Affected) | Serious Events (At Risk) | Other Events (Affected) | Other Events (At Risk) |
|---|---|---|---|---|---|---|---|---|
| EG000 | EOXY Device and Gold Standard Oximeter and SaO2 Measures | Heart rate and SPO2 measures provided from EOXY device Heart rate measures provided from gold standard oximeter. SaO2 measures provided from blood sampling There is only one arm: all subjects have simultaneously three interventions (as required in the European standard ISO 80601-2-61), to qualify a pulse oximeter EOXY device: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard Gold standard oximeter: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard SaO2 sampling: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard |
Not provided
Not provided
Not provided
| Title | Organization | Phone | Extension | |
|---|---|---|---|---|
| Eléonore CAILLET | Bioserenity | 01 57 27 40 00. | eleonore.caillet@bioserenity.com |
Not provided
| Type | Includes Protocol | Includes SAP | Includes ICF | Document Label | Document Date | Document Uploaded Date | Document File Name |
|---|---|---|---|---|---|---|---|
| Prot_SAP | Yes | Yes | No | Study Protocol and Statistical Analysis Plan | Jun 5, 2017 | Dec 19, 2019 | Prot_SAP_000.pdf |
Not provided
No randomisation. All subjects will test simultaneously the gold standard and the EOXY device
Not provided
Not provided
Not provided
Not provided
| Gold standard oximeter | Device | Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard |
|
| SaO2 sampling | Procedure | Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard |
|
| 1 day |
| Participants |
|
| Age, Continuous | Mean | Full Range | years |
|
| Sex: Female, Male | Count of Participants | Participants |
|
| Race (NIH/OMB) | Count of Participants | Participants |
|
| Region of Enrollment | Number | participants |
|
|
|
| Primary | Percentage SPO2 Accuracy Root-Mean Square (ARMS) at Torso Position | For each desaturation plateau, SpO2 ARMS (Accuracy Root-Mean Square) are calculated by the difference between the investigational pulse oximeter SpO2 measurement and the co-oximeter's SaO2 measurements | Subjects respecting the study conditions (no movements, good irrigation) 3 subjects excluded from analysis(no respect of study conditions) so 10 subjects analyzed | Posted | Number | percentage of ARMS (accuracy) | 1 day |
|
|
|
| Primary | Percentage SPO2 Accuracy Root-Mean Square (ARMS) at BACK Position | For each desaturation plateau, SpO2 ARMS are calculated by the difference between the investigational pulse oximeter SpO2 measurement and the co-oximeter's SaO2 measurements | Posted | Number | percentage of ARMS (accuracy) | 1 day |
|
|
|
| Primary | HR Accuracy Root-Mean Square (ARMS) at Forehead Position | For each desaturation plateau, HR ARMS are calculated by the difference between the investigational pulse oximeter HR measurement and the HR gold standard measurements | People respecting study conditions (good irrigation and no movements) | Posted | Number | percentage of ARMS (accuracy) | 1 day |
|
|
|
| Primary | Percentage of HR Accuracy Root-Mean Square (ARMS) at TORSO Position | For each desaturation plateau, HR ARMS are calculated by the difference between the investigational pulse oximeter HR measurement and the HR gold standard measurements | People respecting study conditions (good irrigation, no movements) | Posted | Number | percentage of ARMS (accuracy) | 1 day |
|
|
|
| Primary | Percentage of HR Accuracy Root-Mean Square (ARMS) at BACK Position | For each desaturation plateau, HR ARMS are calculated by the difference between the investigational pulse oximeter HR measurement and the HR gold standard measurements | People respecting study conditions (good irrigation and no movements) | Posted | Number | percentage of ARMS (accuracy) | 1 day |
|
|
|
| 0 |
| 10 |
| 0 |
| 10 |
| 0 |
| 10 |
Not provided
Not provided
| Title | Measurements |
|---|---|
|
| 92-85% of O2 |
|
| 84-78% of O2 |
|
| 77-70% of O2 |
|
| Title | Measurements |
|---|---|
|
| 92-85% of O2 |
|
| 84-78% of O2 |
|
| 77-70% of O2 |
|
| Title | Measurements |
|---|---|
|
| 92-85% of O2 |
|
| 84-78% of O2 |
|
| 77-70% of O2 |
|
| Title | Measurements |
|---|---|
|
| 92-85% of O2 |
|
| 84-78% of O2 |
|
| 77-70% of O2 |
|
| Title | Measurements |
|---|---|
|
| 92-85% of O2 |
|
| 84-78% of O2 |
|
| 77-70% of O2 |
|