A Clinical Trial to Verify the Accuracy of Seers Technology's Pulse Oximeter (mobiCARE +Pulse, MP100W) in Healthy Volunteers
A Clinical Trial to Verify the Accuracy of Seers Technology's Pulse Oximeter (mobiCARE +Pulse, MP100W) in Healthy Volunteers
It is important to evaluate the accuracy of the pulse oximeter, which is commonly used in clinical practice. The US FDA only allows the use of devices that meet the ISO 80601-2-61:2017 standard in the United States. According to the ISO 80601-2-61:2017 standard, the accuracy of the oxygen saturation (SpO2) of the pulse oximeter should have an error range of less than 4.0% compared to the actual arterial blood oxygen saturation (SaO2) in the range of 70-100%. To prove compliance with this requirement, it is necessary to conduct a clinical trial that induces hypoxemia in healthy adults and compares and evaluates the arterial blood oxygen saturation values of the pulse oximeter's SpO2 and the arterial blood oxygen saturation values of the carbon monoxide-oxygen meter (CO-oximeter).
Inclusion Criteria:
Exclusion Criteria: