The purpose of this study is to evaluate the effectiveness and safety of the Dexcom G4 System when used as an adjuvant to blood glucose testing.
Device performance will be primarily evaluated in terms of the proportion of glucose values within a pre-specified range compared to reference values.
Safety data of the G4 System will also be collected vis-a-vis adverse event reporting characterized by the incidence and severity of Serious Adverse Device Events, and Adverse Device Events experienced by study participants.
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Type 1 Diabetes Mellitus (TIDM)Type 2 Diabetes Mellitus (T2DM) | Adults that have been diagnosed with insulin-requiring diabetes and are on multiple daily injections (MDI) or Continuous Subcutaneous Insulin Infusion (CSII) insulin therapy; |
Not provided
| Measure | Description | Time Frame |
|---|---|---|
| Absolute Relative Difference (%) to Reference Standard | The primary objective is to characterize the System performance with respect to laboratory reference venous sample measurements. The device performance will be primarily evaluated in terms of point and rate accuracy of the G4 System in reference to YSI. The point accuracy is measured as the proportion of G4 System values that are within ±20% of YSI reference value for glucose levels >80 mg/dL and ±20 mg/dL at YSI glucose levels <80 mg/dL. | one time measure (day 1) |
Not provided
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Subjects that fit the inclusion criteria will be chosen from the general population
Not provided
| Name | Affiliation | Role |
|---|---|---|
| David Price, MD | DexCom, Inc. | Study Director |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| Rocky Mountain Diabetes and Osteoporosis Center | Idaho Falls | Idaho | 83404 | United States |
Not provided
| ID | Term |
|---|---|
| D003920 | Diabetes Mellitus |
| ID | Term |
|---|---|
| D044882 | Glucose Metabolism Disorders |
| D008659 | Metabolic Diseases |
| D009750 | Nutritional and Metabolic Diseases |
| D004700 | Endocrine System Diseases |