The purpose of this study is to evaluate the safety and effectiveness of the G4 System when worn for up to 7-days by subjects >18 years-old with diabetes mellitus requiring insulin.
Device performance will be primarily evaluated in terms of the proportion of G4 System values that are within ±20% of YSI reference value for glucose levels >80 mg/dL and ±20 mg/dL at meter glucose levels <80 mg/dL. These G4-YSI matched pairs will be collected during in-clinic sessions on Days 1, 4, and 7 of Sensor wear. The proportion of matched pairs from this study (PTL-900360) meeting the above criteria will be compared to the proportion of such points observed in the DexCom clinical study that supported approval of the SEVEN PLUS System (PTL 300012).
Safety data of the G4 System will also be collected and safety will be characterized by the incidence of Adverse Device Effects, Serious Adverse Device Events, and Unanticipated Adverse Device Effects experienced by study participants.
This will be a non-randomized, single-armed study conducted in the United States. Following screening, a minimum of 60 to a maximum of 80 subjects at up to 4 clinical sites will be enrolled. All participants will wear the G4 Sensor for one 7-day wear period (~168 hours). A subset of 20-30 subjects will wear two G4 Systems (one of which will be blinded) during the 7-day wear period for the purposes of assessing precision. During home use, subjects will be asked to use the meter provided to them to take fingersticks for calibration and diabetes management. Subjects will also be asked to confirm high and low G4 System glucose alerts/alarms by taking a comparative fingerstick immediately after receiving an alert/alarm.
All subjects will participate in one, 8-hour in-clinic session on Day 1, 4, or 7 of the study to collect accuracy information against a laboratory standard (YSI) and against SMBG. During the in-clinic session, subjects will have venous blood draws for evaluation of the YSI blood glucose measurements. Carbohydrate consumption and insulin dosing may be manipulated during the in-clinic session to ensure a wide range of glucose values. For the duration of this study, subjects will be instructed to use G4 System information as an adjunct to (and not as a replacement for) standard SMBG guidance of diabetes self-management.
Subjects will perform G4 Sensor insertions at the clinic under supervision of study staff. Sensor insertion sites will be examined after removal of the Sensor(s). Study staff will document any irritation and/or Adverse Device Effects.
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| CGM | Experimental | continuous glucose monitoring system |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| DexCom™ G4 Continuous Glucose Monitoring System | Device | Continuous Glucose Monitoring System |
|
| Measure | Description | Time Frame |
|---|---|---|
| the Proportion of G4 CGM System in Agreement With the Reference Standard | The proportion of G4 System values within (±) 20% of YSI reference value for glucose levels >80 mg/dL and (±) 20 mg/dL at YSI glucose levels <80 mg/dL. This is primarayly laboratory measurement outcome,and it is not measured by clinical outcome, for example, diagnosis, treatement and complication, and clinical effectiveness, etc. | Assessment done on either Day 1, 4 or 7 of the sensor wear period |
Not provided
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Andrew Balo | DexCom, Inc. | Study Chair |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Prospective study- the investigator does not assign specific interventions to the subjects of the study in terms of their diabetes management.
Clinical Research Centers during 12-Mar-2010 to 09-Apr-2010.
Not provided
| ID | Title | Description |
|---|---|---|
| FG000 | Continuous Glucose Monitoring System | DexCom™ G4 Continuous Glucose Monitoring System : Continuous Glucose Monitoring System, that is a glucose monitoring device indicated for detecting trends and tracking patterns in persons with diabetes. The system is intended for single patient use and requires a prescription. |
| Title | Milestones | Reasons Not Completed | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Overall Study |
|
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| BG000 | Continuous Glucose Monitoring System | The Dexcom G4 Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracking patterns in persons with diabetes. The system is intended for single patient use and requires a prescription. |
| Units | Counts |
|---|---|
| Participants |
|
| Title | Description | Population Description | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Denominator Units Selected | Denominators | Classes |
|---|---|---|---|---|---|---|---|---|---|
| Age, Categorical | Count of Participants |
| Type | Title | Description | Population Description | Reporting Status | Anticipated Posting Date | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Time Frame | Units Analyzed | Denominator Units Selected | Arm/Group Information | Denominators | Classes | Analyses | |||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Primary | the Proportion of G4 CGM System in Agreement With the Reference Standard | The proportion of G4 System values within (±) 20% of YSI reference value for glucose levels >80 mg/dL and (±) 20 mg/dL at YSI glucose levels <80 mg/dL. This is primarayly laboratory measurement outcome,and it is not measured by clinical outcome, for example, diagnosis, treatement and complication, and clinical effectiveness, etc. | Randomly enrolled | Posted | Mean | 95% Confidence Interval | %20/20 | Assessment done on either Day 1, 4 or 7 of the sensor wear period |
|
|
7 days
Not provided
Not provided
| ID | Title | Description | Deaths (Affected) | Deaths (At Risk) | Serious Events (Affected) | Serious Events (At Risk) | Other Events (Affected) | Other Events (At Risk) |
|---|---|---|---|---|---|---|---|---|
| EG000 | Real Time Continuous Glucose Monitoring System | continuous glucose monitoring system DexCom™ G4 Continuous Glucose Monitoring System: Continuous Glucose Monitoring System |
Not provided
Not provided
Not provided
| Title | Organization | Phone | Extension | |
|---|---|---|---|---|
| Andy Balo/Sr. VP | Dexcom, Inc | 8582000203 | abalo@dexcom.com |
Not provided
| ID | Term |
|---|---|
| D003920 | Diabetes Mellitus |
| ID | Term |
|---|---|
| D044882 | Glucose Metabolism Disorders |
| D008659 | Metabolic Diseases |
| D009750 | Nutritional and Metabolic Diseases |
| D004700 | Endocrine System Diseases |
Not provided
Not provided
| ID | Term |
|---|---|
| D000095583 | Continuous Glucose Monitoring |
| ID | Term |
|---|---|
| D001774 | Blood Chemical Analysis |
| D019963 | Clinical Chemistry Tests |
| D019411 | Clinical Laboratory Techniques |
| D019937 | Diagnostic Techniques and Procedures |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Participants |
|
| Age, Continuous | Mean | Standard Deviation | years |
|
| Sex: Female, Male | Count of Participants | Participants |
|
| Region of Enrollment | Number | participants |
|
| Counts |
|---|
| Participants |
|
|
| 0 |
| 60 |
| 0 |
| 60 |
Not provided
| D003933 | Diagnosis |
| D003940 | Diagnostic Techniques, Endocrine |
| D008991 | Monitoring, Physiologic |
| D008919 | Investigative Techniques |