The OXISITE-ILD study is a prospective, multicenter, observational study designed to evaluate the agreement between finger (digital) and earlobe (auricular) pulse oximetry for measuring oxygen saturation (SpO2) during the six-minute walk test (6MWT) in patients with interstitial lung disease (ILD). In routine clinical practice, exercise SpO2 is usually measured at the finger; however, the finger reading can be unreliable in some patients, and there is currently no recommendation on the best sensor location in ILD. This study compares the two sensor locations, recorded at the same time, and evaluates whether any disagreement changes how exercise desaturation is classified and whether ambulatory oxygen is indicated.
All patients undergoing a 6MWT as part of routine ILD care are included consecutively to ensure a pragmatic, real-world representation of the ILD population. The primary objective is to measure the agreement between the two locations in the lowest SpO2 reached during the test, including the size and direction of any difference. Secondary objectives include the reclassification of patients at the clinical desaturation thresholds, the comparison between autoimmune and non-autoimmune ILD, the rate of invalid readings at each location, and the clinical, vascular and functional factors associated with disagreement.
Inclusion Criteria:
Exclusion Criteria:
jbordas@fphag.org938 42 50 00
Adults with an established diagnosis of interstitial lung disease (ILD) undergoing a six-minute walk test (6MWT) as part of routine care. Each participant is assessed once, with oxygen saturation (SpO2) and heart rate recorded simultaneously at the finger and at the earlobe. Pre-specified subgroups for the analysis are autoimmune-associated versus non-autoimmune ILD.
Agreement Between 5 Different Pulse Oximeters Used Simultaneously During a Six-minute Walk Test
A Clinical Evaluation of the Oxitone Non-invasive Oximeter
Titrated Ambulatory Oxygen in Fibrotic ILD and COPD With Isolated Exertional Hypoxemia
Comparison of Upper and Lower Limb Maximal Exercise Capacities and Muscle Oxygenation in Patients With ILD
Continuous Versus Demand-Based Oxygen in Patients With Fibrotic Interstitial Lung Disease and Chronic Obstructive Pulmonary Disease
Can High-Flow Oxygen Therapy Improve Oxygenation During Exercise in ILD Patients?
Automated Oxygen Titration During Walking in Patients With Lung Fibrosis (OXYWILD)
Comparative Study of 3 Portable Oxygen Concentrators During a 6-minute Walk Test in Patients With Chronic Lung Disease