| Name | Class |
|---|---|
| Philips Medical Systems | INDUSTRY |
Not provided
Not provided
Not provided
Evaluation of Philips FAST technology with Nellcor sensors
Evaluate saturation accuracy claims for the Philips picoSAT X SpO2 board with FAST algorithm, paired with Nellcor™ sensors in a diverse subject population, over a specified saturation range
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Healthy Volunteers | All subjects who meet the inclusion criteria are enrolled into the test group and participate in data collection have the Philips FAST SpO2 with Nellcor Pulse Oximetry Sensors |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| Philips FAST SpO2 with Nellcor Pulse Oximetry Sensors | Device | Noninvasive pulse oximeter |
|
| Measure | Description | Time Frame |
|---|---|---|
| Evaluation of Saturation Accuracy | Performance will be determined by calculating the accuracy in terms of root mean square difference (ARMS) comparing the noninvasive oxygen saturation measurement (SpO2) to the arterial oxygen saturation (SaO2) value obtained from a reference blood sample. | 2 - 3 hours |
Not provided
Not provided
Inclusion Criteria:
Exclusion Criteria:
Subject is considered as being morbidly obese (defined as BMI >39.5)
Compromised circulation, injury, or physical malformation of fingers, hands, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study (Note: Certain malformations may still allow subjects to participate if the condition is noted and would not affect the sites utilized)
Participants of childbearing potential who are pregnant, who are trying to get pregnant, or who have a urine test positive for pregnancy on the day of the study
Subjects with COHb levels >3% as assessed by CO-Oximetry during the procedure
tHb < 10 g/dl as assessed by CO-Oximetry during the procedure
MetHb ≥ 2% as assessed by CO-Oximetry during the procedure
Subjects with known respiratory conditions such as:
Subjects with known heart or cardiovascular conditions such as:
Self-reported health conditions as identified in the Health Assessment Form
Not provided
Not provided
Healthy subjects ≥ 18 and < 51 years of age, with varying skin tones and diversity across ethnicity, race, gender, and sex assigned at birth.
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Randall M Clark, M.D. | Medtronic Clinical Physiology Lab | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| Medtronic Clinical Physiology Lab | Denver | Colorado | 80218 | United States |
Not provided
| ID | Term |
|---|---|
| D000860 | Hypoxia |
| ID | Term |
|---|---|
| D012818 | Signs and Symptoms, Respiratory |
| D012816 | Signs and Symptoms |
| D013568 | Pathological Conditions, Signs and Symptoms |
Not provided