Respiratory Rate Validation Study - ChroniSense Polso
Respiratory Rate Validation Study - ChroniSense Polso
20 volunteer test subjects entered into an accuracy study that is designed to validate accuracy of the respiratory rate of the ChroniSense Polso
20 volunteer test subjects entered into an accuracy study that is designed to validate accuracy of the respiratory rate of the ChroniSense Polso.
A range of stable respiratory rates were elicited from each volunteer test subject. The rates will be approximately 5, 10, 15, 20, 25, 30, 35, 40 and 45 breaths per minute; with some natural variation from these exact numbers.
Each subject was instrumented with an open system mask that allows for measurement of the end tidal carbon dioxide (EtCO2) respiratory rate and tidal volume. The EtCO2 monitor was used as the Accuracy Reference Device (Reference) in this study.
Inclusion Criteria:
Exclusion Criteria:
Compromised circulation, injury, or physical malformation of fingers, toes, hands, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study. Tattoo in the optical path which would limit the ability to test sites needed for the study. (Note: Certain malformations may still allow subjects to participate if the condition is noted and would not affect the particular sites utilized.)
Subjects with known respiratory conditions such as:
Subjects with self-reported heart or cardiovascular conditions such as:
Self-reported health conditions as identified in the Health Assessment Form
Other known health condition, should be considered upon disclosure in health assessment form