The main objective of this study is to determine whether home use of fully closed-loop glucose control applying ultra-rapid Lispro insulin is superior to standard insulin pump therapy with continuous glucose monitoring (CGM) in adults with type 1 diabetes on insulin pump therapy with sub-optimal glycaemic control (HbA1c ≥ 8.0%).
This is an open-label, single centre, randomised, crossover design study, involving a run-in period followed by two study periods during which glucose levels will be controlled either by an automated closed-loop system using ultra-rapid Lispro insulin or by participants usual insulin pump therapy with continuous glucose monitoring in random order. A total of up to 30 adults (aiming for 24 completed participants) with T1D on insulin pump therapy will be recruited through diabetes clinics and other established methods. Participants who drop out of the study within the first 4 weeks of the first intervention arm will be replaced.
Participants will receive appropriate training in the safe use of the closed-loop devices. Participants will have access to the study team during the home study phase with 24/7 telephone support.
The primary outcome is time spent in target range between 3.9 and 10.0 mmol/L as recorded by CGM over the 8 week period. Secondary outcomes are HbA1c, time spent with glucose levels above and below target as recorded by CGM, and other CGM-based metrics in addition to insulin requirements. Safety evaluation comprises severe hypoglycaemic episodes, diabetic ketoacidosis (DKA) events and other adverse and serious adverse events.
Purpose of the clinical trial:
To determine whether home use of fully closed-loop applying ultra-rapid Lispro insulin is superior to insulin pump therapy with continuous glucose monitoring (CGM).
Study objective:
The study objective is to compare home use of fully closed-loop glucose control applying ultra-rapid Lispro insulin with standard insulin pump therapy with CGM.
Participating clinical centres:
Addenbrooke's Hospital, Cambridge University Hospitals NHS Foundation Trust, Cambridge
Sample Size:
24 adults completing the study. Up to 30 participants will be recruited to allow for dropouts.
Maximum duration of study for a participant:
20 weeks (5 months)
Recruitment The participants will be recruited through the adult diabetes outpatient clinics or other established methods.
Consent:
Participants will be asked to provide written informed consent.
Baseline Assessment:
Eligible participants will undergo a baseline evaluation including a blood sample for the measurement of HbA1c, renal, liver functions, full blood count, thyroid functions and coeliac antibody screen (if not done in the previous 3 months). Urine pregnancy test will be done in females of child bearing age. Human factor questionnaires will be administered.
Run-in Period During the 2-3 week run-in period, participants will use their own insulin pump and wear a masked CGM system. At the end of the run-in period, for compliance, at least 10 days of CGM data need to be recorded. CGM data during the run-in period will be used to assess baseline glucose control before the start of the first home study phase.
Randomisation:
Eligible participants will be randomised using randomisation software to the use of closed-loop glucose control or to standard pump therapy with CGM. There will be no washout period between the two intervention periods.
Automated closed-loop:
Training on the use of closed-loop will be provided by the research team during a 2 to 3 hour session in an outpatient setting (clinical research facility) or may be done remotely. Competency on the use of study insulin pump, study CGM and closed-loop system will be evaluated using a competency assessment tool developed by the research team. Further training may be delivered as required. Participants will be advised to use the closed-loop system with ultra-rapid Lispro insulin for next 8 weeks.
Conventional insulin pump therapy with CGM:
Participants will use their own insulin pump and study CGM. Training on the use of real-time CGM and how to interpret real-time will be provided. Participants will use standard insulin pump therapy, with their usual insulin, and real-time CGM for the next 8 weeks.
Cross-over Assessment:
At the end of the first intervention period, a blood sample for the measurement of HbA1c will be taken and human factor questionnaires will be administered.
End of study assessments:
A blood sample will be taken for measurement of HbA1c and human factor questionnaires will be administered. Study devices will be returned and participants will resume usual care.
Procedures for safety monitoring during trial:
Standard operating procedures for monitoring and reporting of all adverse events and adverse device events will be in place, including serious adverse events (SAE), serious adverse device effects (SADE) and specific adverse events (AE) such as severe hypoglycaemia.
A data monitoring and ethics committee (DMEC) will be informed of all serious adverse events and any unanticipated adverse device/method effects that occur during the study and will review compiled adverse event data at periodic intervals.
Criteria for withdrawal of patients on safety grounds:
A participant may terminate participation in the study at any time without necessarily giving a reason and without any personal disadvantage. An investigator can stop the participation of a subject after consideration of the benefit/risk ratio. Possible reasons are:
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Fully closed-loop system with ultra-rapid Lispro insulin | Experimental | The fully closed-loop system (CamAPS HX) will consist of:
Participants will use ultra-rapid Lispro insulin in the closed-loop system |
|
| Standard insulin pump therapy with CGM | Active Comparator | Participants will use their own insulin pump and usual insulin throughout this study period. The CGM will be the Dexcom G6 real-time CGM sensor (Dexcom, Northridge, CA, USA) |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| CamAPS HX | Device | Fully automated closed-loop system (CamAPS HX) with ultra-rapid Lispro insulin |
|
| Measure | Description | Time Frame |
|---|---|---|
| Time in target glucose range | Time spent in the target glucose range from 3.9 to 10.0 mmol/l (70 to 180mg/dl) based on continuous glucose monitoring (CGM) | 8-week home use |
| Measure | Description | Time Frame |
|---|---|---|
| Time spent above the target glucose range | Time spent above target glucose (10.0 mmol/l) (180 mg/dl) based on CGM | 8-week home use |
| Time spent below the target glucose range | Time spent below target glucose (3.9mmol/l) (70mg/dl) based on CGM |
| Measure | Description | Time Frame |
|---|---|---|
| Safety evaluation | The number of episodes of hypoglycaemia, DKA and / or significant ketonemia (> 3.0mmol/l) as well as nature and severity of any other adverse events including SADEs and SAEs. | 8-week home use |
| Utility evaluation |
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| Cambridge University Hospitals NHS Foundation Trust | Cambridge | CB2 0QQ | United Kingdom |
| PubMed Identifier | Type | Citation | Retractions |
|---|---|---|---|
| 38426909 | Derived | Lakshman R, Hartnell S, Ware J, Allen JM, Wilinska ME, Nwokolo M, Evans ML, Hovorka R, Boughton CK. Lived Experience of Fully Closed-Loop Insulin Delivery in Adults with Type 1 Diabetes. Diabetes Technol Ther. 2024 Apr;26(4):211-221. doi: 10.1089/dia.2023.0394. Epub 2024 Feb 28. | |
| 37616583 | Derived | Boughton CK, Hartnell S, Lakshman R, Nwokolo M, Wilinska ME, Ware J, Allen JM, Evans ML, Hovorka R. Fully Closed-Loop Glucose Control Compared With Insulin Pump Therapy With Continuous Glucose Monitoring in Adults With Type 1 Diabetes and Suboptimal Glycemic Control: A Single-Center, Randomized, Crossover Study. Diabetes Care. 2023 Nov 1;46(11):1916-1922. doi: 10.2337/dc23-0728. |
Not provided
Not provided
Study protocol, statistical analysis plan and fully anonymised individual participant data that underlie the results reported in the manuscript will be available 6 months following publication and ending 36 months following manuscript publication to investigators whose proposed use of the data has been approved by an independent review committee identified for this purpose, to achieve aims in the approved proposal. Proposals should be directed to rh347@cam.ac.uk and may be submitted up to 36 months following article publication. To gain access, data requestors will need to sign a data access agreement.
Fully anonymised data may be shared with third parties (EU or non-EU based) for the purposes of advancing management and treatment of diabetes.
Study protocol, statistical analysis plan and fully anonymised individual participant data that underlie the results reported in the manuscript will be available 6 months following publication and ending 36 months following manuscript publication to investigators whose proposed use of the data has been approved by an independent review committee identified for this purpose, to achieve aims in the approved proposal. Proposals should be directed to rh347@cam.ac.uk and may be submitted up to 36 months following article publication.
Study protocol, statistical analysis plan and fully anonymised individual participant data that underlie the results reported in the manuscript will be available 6 months following publication and ending 36 months following manuscript publication to investigators whose proposed use of the data has been approved by an independent review committee identified for this purpose, to achieve aims in the approved proposal. Proposals should be directed to rh347@cam.ac.uk and may be submitted up to 36 months following article publication.
Not provided
Not provided
| ID | Term |
|---|---|
| D003922 | Diabetes Mellitus, Type 1 |
| ID | Term |
|---|---|
| D003920 | Diabetes Mellitus |
| D044882 | Glucose Metabolism Disorders |
| D008659 | Metabolic Diseases |
| D009750 | Nutritional and Metabolic Diseases |
Not provided
Not provided
| ID | Term |
|---|---|
| D000095583 | Continuous Glucose Monitoring |
| ID | Term |
|---|---|
| D001774 | Blood Chemical Analysis |
| D019963 | Clinical Chemistry Tests |
| D019411 | Clinical Laboratory Techniques |
| D019937 | Diagnostic Techniques and Procedures |
Not provided
Not provided
An open-label, single-centre, randomised, two-period, crossover study comparing closed-loop glucose control compared to standard insulin pump therapy combined with continuous glucose monitoring
Not provided
Not provided
Not provided
Not provided
| Standard insulin pump therapy with CGM | Device | Participants usual insulin pump therapy with Dexcom G6 CGM |
|
| 8-week home use |
| Mean glucose | Average of sensor glucose levels | 8-week home use |
| Standard deviation and coefficient of variation of glucose | Standard deviation and coefficient of variation of CGM glucose levels | 8-week home use |
| Time spent in hypoglycaemia | Time with glucose levels < 3.0 mmol/l (54mg/dl) based on CGM | 8-week home use |
| Time spent in hyperglycaemia | Time with glucose levels >13.9 mmol/l (250mg/dl) and > 16.7 mmol/l) (300mg/dl) | 8-week home use |
| HbA1c | Glycated haemoglobin measured at the end of the treatment period | after 8-week home use |
| Total, basal, and bolus insulin dose | Total, basal, and bolus insulin dose | 8-week home use |
Frequency and duration of use of the closed-loop system at home
| 8-week home use |
| Human Factor assessment | Expectations, attitudes and responses to the closed-loop system will be assessed using questionnaires. INSPIRE questionnaire PAID questionnaire Hypoglycaemia Confidence Scale | 8-week home use |
| D004700 | Endocrine System Diseases |
| D001327 | Autoimmune Diseases |
| D007154 | Immune System Diseases |
| D003933 | Diagnosis |
| D003940 | Diagnostic Techniques, Endocrine |
| D008991 | Monitoring, Physiologic |
| D008919 | Investigative Techniques |