Accuracy of RuiXin-FFR by Comparing With Pressure-wire-based FFR in Detecting Hemodynamically Significant Stenosis: A Prospective Multicenter Study
Accuracy of RuiXin-FFR by Comparing With Pressure-wire-based FFR in Detecting Hemodynamically Significant Stenosis: A Prospective Multicenter Study
Coronary Artery Disease (CAD) is the top killer nowadays. Pressure-wire-based Fractional Flow Reserve (FFR) is the gold standard for measuring ischemia in coronary arteries. CFD-based RuiXin-FFR, which is noninvasive, is developed recently. But its accuracy is not verified. This is a multi-center and prospective study to evaluate the sensitivity, specificity, and accuracy of CFD-based RuiXin-FFR compared with wire-based FFR.
The participants are entered with suspended CAD and angiography ischemia between 30%-90% from 6 hospitals in China. Pressure-wire-based FFR is conducted in the standard protocol by each hospital. FFR>0.8 is chosen as the threshold to indicate non-ischemia. CFD-based RuiXin-FFR is performed by Raysight Medical (Shenzhen, China). CT images are acquired in a standard protocol by each hospital. Images with poor quality are excluded in this trial. Based on these images, RuiXin-FFR reconstructs the 3D model of coronary trees and obtains RuiXin-FFR values by conducting CFD simulations in a blinded fashion. RuiXin-FFR>0.8 is chosen as the threshold to indicate non-ischemia as well. By comparing the clinical diagnosis outcome of pressure-wire-based FFR and CFD-based RuiXin-FFR, the sensitivity and specificity of the RuiXin-FFR are obtained, which is the primary end point of this trial. The secondary end point includes the measurement of accuracy, NPV, PPV and ROC of the RuiXin-FFR.
Inclusion Criteria:
Exclusion Criteria: