Performance testing of pulse oximeters. The protocol involves brief stable arterial oxygen desaturation in healthy volunteers and sampling arterial blood when a stable level of hypoxia has been attained. The blood sample is analyzed for oxygen saturation with a gold standard bench CO-oximeter. This protocol is aligned with the latest ISO and FDA guidance documents for pulse oximeter testing.
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| Owlet Smart Sock Sensor 3rd Generation | Diagnostic Test | Pulse Oximetry |
| Measure | Description | Time Frame |
|---|---|---|
| SpO2 Accuracy (Arms) of ≤3 Between 70 and 100% | Determination of Average Root Mean Square SpO2 Accuracy of the Owlet Smart Sock 3rd Generation | 1 hour |
Not provided
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Participants in this study should be healthy, between 18 and 50 years of age and willing to participate in breathing studies with or without blood samples in San Francisco.
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Philip Bickler, MD | UCSF Department of Anesthesia | Principal Investigator |
| John Feiner, MD | UCSF Department of Anesthesia | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| University California San Fransisco, Department of Anesthesia | San Francisco | California | 94133 | United States |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| FG000 | ARMS SpO2 70-100% | Comparison to Reference CO-Oximetry Accuracy of the Owlet Smart Sock Sensor 3rd Generation: The purpose of this study was to validate the SpO2 accuracy of the Owlet Smart Sock Sensor 3rd Generation during non-motion conditions over the range of 70-100% SaO2 as compared to arterial blood samples assessed by CO-Oximetry. It was expected that the Accuracy Root Mean Square (ARMS) performance of the oximetry system would meet the required specification of ARMS of 3.0% or less. |
| Title | Milestones | Reasons Not Completed | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Overall Study |
|
14 Healthy volunteer subjects
| ID | Title | Description |
|---|---|---|
| BG000 | ARMS SpO2 70-100% | Comparison to Reference CO-Oximetry Accuracy of the Owlet Smart Sock Sensor 3rd Generation: The purpose of this study was to validate the SpO2 accuracy of the Owlet Smart Sock Sensor 3rd Generation during non-motion conditions over the range of 70-100% SaO2 as compared to arterial blood samples assessed by CO-Oximetry. It was expected that the Accuracy Root Mean Square (ARMS) performance of the oximetry system would meet the required specification of ARMS of 3.0% or less. |
| Units | Counts |
|---|---|
| Participants |
|
| Title | Description | Population Description | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Denominator Units Selected | Denominators | Classes |
|---|---|---|---|---|---|---|---|---|---|
| Age, Categorical | Count of Participants |
| Type | Title | Description | Population Description | Reporting Status | Anticipated Posting Date | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Time Frame | Units Analyzed | Denominator Units Selected | Arm/Group Information | Denominators | Classes | Analyses | |||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Primary | SpO2 Accuracy (Arms) of ≤3 Between 70 and 100% | Determination of Average Root Mean Square SpO2 Accuracy of the Owlet Smart Sock 3rd Generation | Posted | Number | Percentage | 1 hour | Number of Data Points | Number of Data Points |
|
Study Duration (approximately 1 hour)
All subjects completed the study without events.
Not provided
| ID | Title | Description | Deaths (Affected) | Deaths (At Risk) | Serious Events (Affected) | Serious Events (At Risk) | Other Events (Affected) | Other Events (At Risk) |
|---|---|---|---|---|---|---|---|---|
| EG000 | ARMS SpO2 70-100%, Oximeter #1 (Left) | Comparison to Reference CO-Oximetry on the left thumb. Accuracy of the Owlet Smart Sock Sensor 3rd Generation: The purpose of this study was to validate the SpO2 accuracy of the Owlet Smart Sock Sensor 3rd Generation during non-motion conditions over the range of 70-100% SaO2 as compared to arterial blood samples assessed by CO-Oximetry. It was expected that the Accuracy Root Mean Square (ARMS) performance of the oximetry system would meet the required specification of ARMS of 3.0% or less. |
Not provided
Not provided
Not provided
| Title | Organization | Phone | Extension | |
|---|---|---|---|---|
| Madison Kenley, Clinical Research Director | Owlet Baby Care, Inc | 801-552-8383 | mkenley@owletcare.com |
Not provided
| Type | Includes Protocol | Includes SAP | Includes ICF | Document Label | Document Date | Document Uploaded Date | Document File Name |
|---|---|---|---|---|---|---|---|
| Prot_SAP | Yes | Yes | No | Study Protocol and Statistical Analysis Plan | Jan 22, 2019 | Sep 22, 2020 | Prot_SAP_000.pdf |
Not provided
| ID | Term |
|---|---|
| D000860 | Hypoxia |
| ID | Term |
|---|---|
| D012818 | Signs and Symptoms, Respiratory |
| D012816 | Signs and Symptoms |
| D013568 | Pathological Conditions, Signs and Symptoms |
Not provided
Not provided
Not provided
Not provided
Not provided
| Arterial Blood Samples |
|
| Participants |
| Participants |
|
| Sex: Female, Male | Count of Participants | Participants | Participants |
|
| Race (NIH/OMB) | Count of Participants | Participants | Participants |
|
| Region of Enrollment | Number | participants | Participants |
|
| Skin Tone | Count of Participants | Participants | Participants |
|
Comparison to Reference CO-Oximetry on the right thumb. Accuracy of the Owlet Smart Sock Sensor 3rd Generation: The purpose of this study was to validate the SpO2 accuracy of the Owlet Smart Sock Sensor 3rd Generation during non-motion conditions over the range of 70-100% SaO2 as compared to arterial blood samples assessed by CO-Oximetry. It was expected that the Accuracy Root Mean Square (ARMS) performance of the oximetry system would meet the required specification of ARMS of 3.0% or less. |
|
|
| 0 |
| 14 |
| 0 |
| 14 |
| 0 |
| 14 |
| EG001 | ARMS SpO2 70-100%, Oximeter #2 (Right) | Comparison to Reference CO-Oximetry on the left thumb. Accuracy of the Owlet Smart Sock Sensor 3rd Generation: The purpose of this study was to validate the SpO2 accuracy of the Owlet Smart Sock Sensor 3rd Generation during non-motion conditions over the range of 70-100% SaO2 as compared to arterial blood samples assessed by CO-Oximetry. It was expected that the Accuracy Root Mean Square (ARMS) performance of the oximetry system would meet the required specification of ARMS of 3.0% or less. | 0 | 14 | 0 | 14 | 0 | 14 |
Not provided
Not provided