| Name | Class |
|---|---|
| Clinimark, LLC | OTHER |
Not provided
Not provided
Pulse oximetry monitoring is considered a standard physiological measurement and is used by clinicians in everyday situations to estimate arterial oxygen saturation. There are two devices used in this investigational study: BabySat v.1.0 and Owlet Baby Care, Inc. Smart SockTM 2 (currently available over-the-counter), with specific emphasis on its pulse oximetry accuracy. The BabySat and Smart SockTM 2 are non-invasive home care devices for use with infants. The purpose of this study was to validate the SpO2 accuracy of BabySat v1.0 and Owlet Smart SockTM 2, OSS v1.1 sensors pulse oximetry, during non-motion conditions over the range of 70-100% SaO2 as compared to arterial blood samples assessed by CO-Oximetry.
Pulse oximetry monitoring is considered a standard physiological measurement and is used by clinicians in everyday situations to estimate arterial oxygen saturation. A pulse oximeter is a device that measures the oxygen saturation of arterial blood non-invasively.
The purpose of this study was to evaluate the SpO2 accuracy performance of the Owlet BabySat v1.0 and Smart SockTM 2 pulse oximetry OSS v1.1 sensors during non-motion conditions over the range of 70-100% SaO2 to arterial blood samples assessed by CO-Oximetry for a SpO2 validation.
The goal, in its entirety, was to show the SpO2 accuracy performance for the investigational device Owlet Baby Care, Inc. Smart SockTM V. 2.
It was expected that the Accuracy Root Mean Square (ARMS) performance would meet the required specification of ARMS 3% or lower in non-motion conditions for the range of 70 - 100% SaO2 thereby demonstrating an acceptable SpO2 accuracy performance specification.
No risks or adverse device effects were expected. There were no contraindications for use in the proposed study population.
The study was conducted in accordance to the code of federal regulations for non-significant risk medical device studies and applicable ISO 14155 (2nd edition 2011-02-01), applicable sections of ISO80601-2-61 (1st edition 2011-04- 01), and Pulse Oximeters - Premarket Notifications Submissions [510(k)s] Guidance For Industry and Food and Drug Administration Staff (issued: March 4, 2013)
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Reference CO-Oximetry | Active Comparator | Reference Co-Oximetry |
|
| Owlet BabySat v1.0 | Experimental | Owlet BabySat v1.0 |
|
| Owlet Smart Sock V2 v1.1 | Experimental | Owlet Smart SockTM 2 , OSS v1.1 Sensor and Custom Adult Thumb Sock |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| Pulse oximetry with Owlet BabySat v1.0 sensor | Diagnostic Test | A comparative, single-center, non-randomized study to evaluate the SpO2 accuracy guidelines for pulse oximetry over the range of 70-100% SaO2 under non-motion conditions. Arterial blood sampling measured by functional SaO2 CO-Oximetry, was used as the basis for comparison with Owlet BabySat v1.0 and Owlet Smart Sock 2 sensors |
| Measure | Description | Time Frame |
|---|---|---|
| It Was Expected That the Accuracy Root Mean Square (ARMS) Between Measured SpO2 of Owlet Smart Sock V2 v1.1 and Reference SpO2 Would Meet the Required Specification of ARMS 3% or Lower in Non-motion Conditions for the Range of 70 - 100% | The purpose of the study was to evaluate the accuracy of the blood oxygen saturation (SpO2) at rest over the range of 70-100% compared to arterial blood samples assessed by standard methods [Co-Oximetry]. A standard pulse oximeter accuracy is evaluated by the Accuracy Root Mean Square (ARMS). The results of accurate performance of the oximetry system require ARMS of 3% or less in non-motion conditions for the range of 70-100% SpO2 per standard. | Acute immediate assessment of the sensor accuracy compared to CO-oximetry |
Not provided
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Milena Adamian, MD | Owlet Baby Care, Inc. | Study Director |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| Owlet Baby Care, Inc. | Lehi | Utah | 84043 | United States |
Not provided
Not provided
Not provided
Not provided
Not provided
Eleven subjects were enrolled into the study with one subject later withdrawn due to poor sensor placement.
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| FG000 | Owlet Smart Sock V2 v1.1 | Pulse oximetry with Owlet BabySat v1.0 sensor: A comparative, single-center, non-randomized study to evaluate the SpO2 accuracy guidelines for pulse oximetry over the range of 70-100% SaO2 under non-motion conditions. Arterial blood sampling measured by functional SaO2 CO-Oximetry, was used as the basis for comparison with Owlet BabySat v1.0 and Owlet Smart Sock 2 sensors |
| Title | Milestones | Reasons Not Completed | ||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Overall Study |
|
|
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| BG000 | Owlet Smart Sock V2 v1.1 | Pulse oximetry with Owlet BabySat v1.0 sensor: A comparative, single-center, non-randomized study to evaluate the SpO2 accuracy guidelines for pulse oximetry over the range of 70-100% SaO2 under non-motion conditions. Arterial blood sampling measured by functional SaO2 CO-Oximetry, was used as the basis for comparison with Owlet BabySat v1.0 and Owlet Smart Sock 2 sensors |
| Units | Counts |
|---|---|
| Participants |
|
| Title | Description | Population Description | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Denominator Units Selected | Denominators | Classes |
|---|---|---|---|---|---|---|---|---|---|
| Age, Categorical | Count of Participants |
| Type | Title | Description | Population Description | Reporting Status | Anticipated Posting Date | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Time Frame | Units Analyzed | Denominator Units Selected | Arm/Group Information | Denominators | Classes | Analyses | |||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Primary | It Was Expected That the Accuracy Root Mean Square (ARMS) Between Measured SpO2 of Owlet Smart Sock V2 v1.1 and Reference SpO2 Would Meet the Required Specification of ARMS 3% or Lower in Non-motion Conditions for the Range of 70 - 100% | The purpose of the study was to evaluate the accuracy of the blood oxygen saturation (SpO2) at rest over the range of 70-100% compared to arterial blood samples assessed by standard methods [Co-Oximetry]. A standard pulse oximeter accuracy is evaluated by the Accuracy Root Mean Square (ARMS). The results of accurate performance of the oximetry system require ARMS of 3% or less in non-motion conditions for the range of 70-100% SpO2 per standard. | Posted | Number | Percent SpO2 | Acute immediate assessment of the sensor accuracy compared to CO-oximetry | Number of Data Points | Number of Data Points |
|
Throughout study duration and a follow-up with each subject will be conducted within 7 days following participation in the study and will be conducted via telephone, text or email.
Not provided
Not provided
| ID | Title | Description | Deaths (Affected) | Deaths (At Risk) | Serious Events (Affected) | Serious Events (At Risk) | Other Events (Affected) | Other Events (At Risk) |
|---|---|---|---|---|---|---|---|---|
| EG000 | Owlet Smart Sock V2 v1.1 | Pulse oximetry with Owlet BabySat v1.0 sensor: A comparative, single-center, non-randomized study to evaluate the SpO2 accuracy guidelines for pulse oximetry over the range of 70-100% SaO2 under non-motion conditions. Arterial blood sampling measured by functional SaO2 CO-Oximetry, was used as the basis for comparison with Owlet BabySat v1.0 and Owlet Smart Sock 2 sensors |
Not provided
Not provided
Not provided
| Title | Organization | Phone | Extension | |
|---|---|---|---|---|
| Madison Kenley, CCRP, BS | Owlet Baby Care, Inc. | 801-552-8383 | mkenley@owletcare.com |
Not provided
| Type | Includes Protocol | Includes SAP | Includes ICF | Document Label | Document Date | Document Uploaded Date | Document File Name |
|---|---|---|---|---|---|---|---|
| Prot_SAP | Yes | Yes | No | Study Protocol and Statistical Analysis Plan | Nov 30, 2017 | May 9, 2019 | Prot_SAP_000.pdf |
Not provided
| ID | Term |
|---|---|
| D000860 | Hypoxia |
| ID | Term |
|---|---|
| D012818 | Signs and Symptoms, Respiratory |
| D012816 | Signs and Symptoms |
| D013568 | Pathological Conditions, Signs and Symptoms |
Not provided
Not provided
| ID | Term |
|---|---|
| D010092 | Oximetry |
| ID | Term |
|---|---|
| D001784 | Blood Gas Analysis |
| D001774 | Blood Chemical Analysis |
| D019963 | Clinical Chemistry Tests |
| D019411 | Clinical Laboratory Techniques |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
|
| Participants |
|
| Age, Continuous | Mean | Standard Deviation | years |
|
| Sex: Female, Male | Count of Participants | Participants |
|
| Ethnicity (NIH/OMB) | Count of Participants | Participants |
|
| Race (NIH/OMB) | Count of Participants | Participants |
|
| Region of Enrollment | Number | participants |
|
| Weight | Mean | Standard Deviation | lbs |
|
| Height | Mean | Standard Deviation | inches |
|
|
|
| 0 |
| 11 |
| 0 |
| 11 |
| 0 |
| 11 |
Not provided
Not provided
| D019937 | Diagnostic Techniques and Procedures |
| D003933 | Diagnosis |
| D006334 | Heart Function Tests |
| D003935 | Diagnostic Techniques, Cardiovascular |
| D012129 | Respiratory Function Tests |
| D003948 | Diagnostic Techniques, Respiratory System |
| D008919 | Investigative Techniques |