The mBCA 515 was compared to the gold standard methods MRI, DXA, ADP, D2O- and NaBr-dilution in previous clinical investigations (BCA-01, BCA-02, BCA-20) to generate prediction equations for calculating body composition (FFM, SMM, ECW, TBW and VAT) in different populations. Furthermore, normal ranges for body composition were generated based on the previous study BCA-03 conducted at the UKE.
The mBCA 555 is a new body composition analyzer measuring bioimpedance in standing position developed by seca.
In order to apply prediction equations and normal ranges from previous clinical investigations BCA-01, BCA-02, BCA-03 and BCA-20 to the new mBCA 555 it shall be compared to the mBCA 515.
The use of bioelectrical impedance analysis (BIA) is widespread both in healthy subjects and patients. BIA allows the determination of the fat-free mass (FFM), skeletal muscle mass (SMM), fat mass (FM), extracellular water (ECW) and total body water (TBW) when using appropriate population, age or pathology-specific BIA equations and established procedures. BIA equations are based on validation studies where BIA is compared with reference methods such as Air Displacement Plethysmography (ADP), Dual-Energy X-Ray Absorptiometry (DXA), Magnetic Resonance Imaging (MRI) and isotope dilution.
The predicate device seca mBCA 515 was compared to the gold standard methods MRI, DXA, ADP, D2O- and NaBr-dilution in previous clinical investigations in Kiel (Germany) and New York City (USA) to generate prediction equations for calculating body composition (FFM, SMM, ECW, TBW and VAT) in different populations. Furthermore, normal ranges for body composition were generated in a previous study conducted at the University Hospital Hamburg-Eppendorf.
The mBCA 555 is a new body composition analyzer measuring bioimpedance in standing position developed by seca.
Primary objective is to test whether the mBCA 515 prediction equations can be applied to the new mBCA 555. If necessary, to proof substantial equivalence, prediction equations should be adjusted for the mBCA 555.
Secondary objective is to compare Impedance, phase angle, reactance and resistance for different frequencies in order to clarify if normal ranges for BIVA and phase angle generated with the mBCA 515 need to be adjusted for the new mBCA 555.
Bioimpedance was measured using the seca mBCA 555 and mBCA 515. Body height was measured using the seca 274.
Not provided
Not provided
Not provided
Not provided
Not provided
| Measure | Description | Time Frame |
|---|---|---|
| SMM | Skeletal muscle mass, bias of mBCA 555 in comparison to mBCA 515 | 10min |
| Measure | Description | Time Frame |
|---|---|---|
| Phase Angle | Phase angle at 50 kHz, bias of mBCA 555 in comparison to mBCA 515 | 10 minutes |
| Measure | Description | Time Frame |
|---|---|---|
| R50RS | resistance right body side, 50 kHz, bias of mBCA 555 in comparison to mBCA 515 | 5 minutes |
| R5RS | resistance right body side, 5 kHz, bias of mBCA 555 in comparison to mBCA 515 |
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Not provided
healthy adults
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Sven Peine, Dr. | Universitätsklinikum Hamburg-Eppendorf | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| University Medical Center Hamburg-Eppendorf - Institute of Transfusion Medicine | Hamburg | 20246 | Germany |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Subjects were recruited at the blood donor service of the Institute of Transfusion Medicine. All subjects needed to be suitable for blood donation according to "Hämotherapie-Richtlinien nach §§ 12 a und 18 TFG", chapter 2.1.4. Blood donation was conducted after study examinations.
Not provided
| ID | Title | Description |
|---|---|---|
| FG000 | Healthy Blood Donors | single arm study with healthy blood donors |
| Title | Milestones | Reasons Not Completed | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Overall Study |
|
healthy male and female subjects 18years and older
1 subject of the planed 120 subjects did not complete the study
Not provided
| ID | Title | Description |
|---|---|---|
| BG000 | Healthy Blood Donors | single arm study with healthy blood donors |
| Units | Counts |
|---|---|
| Participants |
|
| Title | Description | Population Description | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Denominator Units Selected | Denominators | Classes |
|---|---|---|---|---|---|---|---|---|---|
| Age, Continuous | No valid measurement is available for ID 113, resulting in 119 valid measurements |
| Type | Title | Description | Population Description | Reporting Status | Anticipated Posting Date | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Time Frame | Units Analyzed | Denominator Units Selected | Arm/Group Information | Denominators | Classes | Analyses | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Primary | SMM | Skeletal muscle mass, bias of mBCA 555 in comparison to mBCA 515 | healthy blood donors | Posted | Mean | Standard Deviation | kg | 10min |
|
|
2 month
No AE or SAE were observed
Not provided
| ID | Title | Description | Deaths (Affected) | Deaths (At Risk) | Serious Events (Affected) | Serious Events (At Risk) | Other Events (Affected) | Other Events (At Risk) |
|---|---|---|---|---|---|---|---|---|
| EG000 | Single Arm Study | healthy blood donors | 0 |
Not provided
Not provided
Not provided
| Title | Organization | Phone | Extension | |
|---|---|---|---|---|
| Kristin Klückmann | seca gmbh & co. kg | +49(40)200000 | 232 | kristin.klueckmann@seca.com |
Not provided
| Type | Includes Protocol | Includes SAP | Includes ICF | Document Label | Document Date | Document Uploaded Date | Document File Name |
|---|---|---|---|---|---|---|---|
| Prot_SAP | Yes | Yes | No | Study Protocol and Statistical Analysis Plan | Dec 7, 2018 | Oct 5, 2021 | Prot_SAP_000.pdf |
Not provided
Not provided
Not provided
Not provided
| 5 minutes |
| Mean |
| Standard Deviation |
| years |
|
| Sex: Female, Male | No valid measurement is available for ID 113, resulting in 119 valid measurements | Count of Participants | Participants |
|
| Ethnicity (NIH/OMB) | No valid measurement is available for ID 113, resulting in 119 valid measurements | Count of Participants | Participants |
|
| weight | No valid measurement is available for ID 113, resulting in 119 valid measurements | Mean | Standard Deviation | kg |
|
|
| Secondary | Phase Angle | Phase angle at 50 kHz, bias of mBCA 555 in comparison to mBCA 515 | healthy blood donors | Posted | Mean | Standard Error | Degree | 10 minutes |
|
|
|
| Other Pre-specified | R50RS | resistance right body side, 50 kHz, bias of mBCA 555 in comparison to mBCA 515 | Posted | Least Squares Mean | Standard Deviation | Ohm | 5 minutes |
|
|
|
| Other Pre-specified | R5RS | resistance right body side, 5 kHz, bias of mBCA 555 in comparison to mBCA 515 | Posted | Least Squares Mean | Standard Deviation | Ohm | 5 minutes |
|
|
|
| 120 |
| 0 |
| 120 |
| 0 |
| 120 |
Not provided
Not provided