Evaluation of electromagnetic interference between Bioelectrical Impedance Analysis (BIA) and implantable cardioverter defibrillators (ICDs).
Electrograms of ICDs will be analyzed during wireless telemetry and during body composition measurement using BIA. ICD therapy will have been temporarily inactivated before the measurement to avoid inappropriate shocks, and the sensitivity level of the ICD will be set to maximum. Body composition will be measured at three frequencies (5 khz, 50 khz and 100 khz).
Interference will be assessed by the presence of visible artefacts on the real-time electrogram, with ICD marker annotations showing atrial and/or ventricular sensed episodes. The outcome measure will be the number and proportion of patients who have electromagnetic interferences of the ICD when BIA is done.
The procedure will take approximately 5-10minutes, with reprogramming of the device at the end of the test. No further follow-up in the context of the study will be performed.
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Electromagnetic Interference |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| Exposure to BIA interference with the ICD | Other | Exposure to BIA interference with the ICD |
|
| Measure | Description | Time Frame |
|---|---|---|
| Number of patients who have electromagnetic interference of the ICD when exposed to BIA. | 5-10 minutes |
Not provided
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Patients with ICD.
Not provided
Not provided
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| University Hospital Geneva | Recruiting | Geneva | 1211 | Switzerland |
| PubMed Identifier | Type | Citation | Retractions |
|---|---|---|---|
| 22452409 | Background | Buch E, Bradfield J, Larson T, Horwich T. Effect of bioimpedance body composition analysis on function of implanted cardiac devices. Pacing Clin Electrophysiol. 2012 Jun;35(6):681-4. doi: 10.1111/j.1540-8159.2012.03377.x. Epub 2012 Mar 27. | |
| 26903305 | Derived | Meyer P, Makhlouf AM, Mondouagne Engkolo LP, Trentaz F, Thibault R, Pichard C, Burri H. Safety of Bioelectrical Impedance Analysis in Patients Equipped With Implantable Cardioverter Defibrillators. JPEN J Parenter Enteral Nutr. 2017 Aug;41(6):981-985. doi: 10.1177/0148607116633823. Epub 2016 Feb 22. |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided