| Primary | Time Spent <70 mg/dL | Difference in time <70 mg/dL between intervention and control group assessed in days 194 to 208 adjusting for baseline (days 1 to 15). | 1 subject from the standard blood glucose monitoring group had no baseline sensor data and could not be included in the analysis of sensor data. | Posted | | Mean | Standard Deviation | hours per day | | Baseline and Days 194 to 208 | | | | ID | Title | Description |
|---|
| OG000 | Sensor Based Glucose Monitoring System | Standard sensing system use for 6 months. Sensor Based Glucose Monitoring System: Subjects will wear the Abbott Sensor Based Glucose Monitoring System (unmasked) for 6 months to monitor their glucose levels. All subjects will wear a masked Abbott Sensor Based Glucose Monitoring System, for 14 days prior to randomisation (Sensor glucose measurements not visible during this time). | | OG001 | Standard Blood Glucose Monitoring | Subjects randomised to the control group will be given blood glucose meters for monitoring for the 6 months study duration. Standard Blood Glucose Monitoring: Subjects will use an Abbott Blood Glucose Monitoring System (standard blood glucose meter) for 6 months to monitor their glucose levels. A 14-day masked wear of the Abbott Sensor Based Glucose Monitoring System is included for these subjects at the 3 and 6 month time point, to collect glycaemic variability data for comparison to the intervention group of the study (Sensor glucose measurements not visible during this time). All subjects will wear a masked Abbott Sensor Based Glucose Monitoring System, for 14 days prior to randomisation (Sensor glucose measurements not visible during this time). |
| | | Title | Denominators | Categories |
|---|
| Baseline | | | Title | Measurements |
|---|
| - OG0003.38± 2.31
- OG0013.44± 2.62
|
| | Days 194-208 | | |
| | Group IDs | Group Description | Statistical Method | Statistical Comment | P-Value | P-Value Comment | Parameter Type | Parameter Value | Dispersion Type | Dispersion Value | Confidence Interval Sides | Confidence Interval % | CI Lower Limit | CI Upper Limit | CI Lower Limit Comment | CI Upper Limit Comment | Estimate Comment | Tested Non-Inferiority | Non-Inferiority Type | Non-Inferiority Comment | Other Analysis Description |
|---|
| | ANCOVA | | <0.0001 | | Mean Difference (Final Values) | -1.24 | Standard Error of the Mean | 0.239 | 2-Sided | | | | | | | | Superiority or Other | | |
|
| Secondary | HbA1c at 6 Months | Difference in HbA1c between intervention and control group at day 208 adjusting for baseline HbA1c at day 1 | | Posted | | Mean | Standard Deviation | percentage of Glycated Haemoglobin | | Baseline and Day 208 | | | | ID | Title | Description |
|---|
| OG000 | Sensor Based Glucose Monitoring System | Standard sensing system use for 6 months. Sensor Based Glucose Monitoring System: Subjects will wear the Abbott Sensor Based Glucose Monitoring System (unmasked) for 6 months to monitor their glucose levels. All subjects will wear a masked Abbott Sensor Based Glucose Monitoring System, for 14 days prior to randomisation (Sensor glucose measurements not visible during this time). | | OG001 | Standard Blood Glucose Monitoring | Subjects randomised to the control group will be given blood glucose meters for monitoring for the 6 months study duration. Standard Blood Glucose Monitoring: Subjects will use an Abbott Blood Glucose Monitoring System (standard blood glucose meter) for 6 months to monitor their glucose levels. A 14-day masked wear of the Abbott Sensor Based Glucose Monitoring System is included for these subjects at the 3 and 6 month time point, to collect glycaemic variability data for comparison to the intervention group of the study (Sensor glucose measurements not visible during this time). All subjects will wear a masked Abbott Sensor Based Glucose Monitoring System, for 14 days prior to randomisation (Sensor glucose measurements not visible during this time). |
|
| Secondary | Time Spent <55 mg/dL and <40 mg/dL | Difference in time <55 mg/dL & <40 mg/dL (hours per day) between intervention and control group assessed in days 194 to 208 adjusting for baseline (days 1 to 15). | 1 subject from the standard blood glucose monitoring group had no baseline sensor data and could not be included in the analysis of sensor data. | Posted | | Mean | Standard Deviation | hours per day | | Baseline and Days 194 to 208 | | | | ID | Title | Description |
|---|
| OG000 | Sensor Based Glucose Monitoring System | Standard sensing system use for 6 months. Sensor Based Glucose Monitoring System: Subjects will wear the Abbott Sensor Based Glucose Monitoring System (unmasked) for 6 months to monitor their glucose levels. All subjects will wear a masked Abbott Sensor Based Glucose Monitoring System, for 14 days prior to randomisation (Sensor glucose measurements not visible during this time). | | OG001 | Standard Blood Glucose Monitoring | Subjects randomised to the control group will be given blood glucose meters for monitoring for the 6 months study duration. Standard Blood Glucose Monitoring: Subjects will use an Abbott Blood Glucose Monitoring System (standard blood glucose meter) for 6 months to monitor their glucose levels. A 14-day masked wear of the Abbott Sensor Based Glucose Monitoring System is included for these subjects at the 3 and 6 month time point, to collect glycaemic variability data for comparison to the intervention group of the study (Sensor glucose measurements not visible during this time). All subjects will wear a masked Abbott Sensor Based Glucose Monitoring System, for 14 days prior to randomisation (Sensor glucose measurements not visible during this time). |
|
| Secondary | Frequency of Episodes <70 mg/dL, <55 mg/dL and <40 mg/dL | Difference in frequency of episodes <70 mg/dL, <55 mg/dL and <40 mg/dL (number per day) between intervention and control group assessed in days 194 to 208 adjusting for baseline (days 1 to 15). | 1 subject from the standard blood glucose monitoring group had no baseline sensor data and could not be included in the analysis of sensor data. | Posted | | Mean | Standard Deviation | number of episodes per day | | Baseline and Days 194-208 | | | | ID | Title | Description |
|---|
| OG000 | Sensor Based Glucose Monitoring System | Standard sensing system use for 6 months. Sensor Based Glucose Monitoring System: Subjects will wear the Abbott Sensor Based Glucose Monitoring System (unmasked) for 6 months to monitor their glucose levels. All subjects will wear a masked Abbott Sensor Based Glucose Monitoring System, for 14 days prior to randomisation (Sensor glucose measurements not visible during this time). | | OG001 | Standard Blood Glucose Monitoring | Subjects randomised to the control group will be given blood glucose meters for monitoring for the 6 months study duration. Standard Blood Glucose Monitoring: Subjects will use an Abbott Blood Glucose Monitoring System (standard blood glucose meter) for 6 months to monitor their glucose levels. A 14-day masked wear of the Abbott Sensor Based Glucose Monitoring System is included for these subjects at the 3 and 6 month time point, to collect glycaemic variability data for comparison to the intervention group of the study (Sensor glucose measurements not visible during this time). All subjects will wear a masked Abbott Sensor Based Glucose Monitoring System, for 14 days prior to randomisation (Sensor glucose measurements not visible during this time). |
|
| Secondary | Time Spent >180 mg/dL and >240 mg/dL | Difference in time >180 mg/dL and >240 mg/dL (hours per day) between intervention and control group assessed in days 194 to 208 adjusting for baseline (days 1 to 15). | 1 subject from the standard blood glucose monitoring group had no baseline sensor data and could not be included in the analysis of sensor data. | Posted | | Mean | Standard Deviation | hours per day | | Baseline and Days 194 to 208 | | | | ID | Title | Description |
|---|
| OG000 | Sensor Based Glucose Monitoring System | Standard sensing system use for 6 months. Sensor Based Glucose Monitoring System: Subjects will wear the Abbott Sensor Based Glucose Monitoring System (unmasked) for 6 months to monitor their glucose levels. All subjects will wear a masked Abbott Sensor Based Glucose Monitoring System, for 14 days prior to randomisation (Sensor glucose measurements not visible during this time). | | OG001 | Standard Blood Glucose Monitoring | Subjects randomised to the control group will be given blood glucose meters for monitoring for the 6 months study duration. Standard Blood Glucose Monitoring: Subjects will use an Abbott Blood Glucose Monitoring System (standard blood glucose meter) for 6 months to monitor their glucose levels. A 14-day masked wear of the Abbott Sensor Based Glucose Monitoring System is included for these subjects at the 3 and 6 month time point, to collect glycaemic variability data for comparison to the intervention group of the study (Sensor glucose measurements not visible during this time). All subjects will wear a masked Abbott Sensor Based Glucose Monitoring System, for 14 days prior to randomisation (Sensor glucose measurements not visible during this time). |
|
| Secondary | Time in Range | Difference in time in range 70-180 mg/dL between intervention and control group assessed in days 194 to 208 adjusting for baseline (days 1 to 15 time in range). | 1 subject from the standard blood glucose monitoring group had no baseline sensor data and could not be included in the analysis of sensor data. | Posted | | Mean | Standard Deviation | hours per day | | Baseline and Days 194 to 208 | | | | ID | Title | Description |
|---|
| OG000 | Sensor Based Glucose Monitoring System | Standard sensing system use for 6 months. Sensor Based Glucose Monitoring System: Subjects will wear the Abbott Sensor Based Glucose Monitoring System (unmasked) for 6 months to monitor their glucose levels. All subjects will wear a masked Abbott Sensor Based Glucose Monitoring System, for 14 days prior to randomisation (Sensor glucose measurements not visible during this time). | | OG001 | Standard Blood Glucose Monitoring | Subjects randomised to the control group will be given blood glucose meters for monitoring for the 6 months study duration. Standard Blood Glucose Monitoring: Subjects will use an Abbott Blood Glucose Monitoring System (standard blood glucose meter) for 6 months to monitor their glucose levels. A 14-day masked wear of the Abbott Sensor Based Glucose Monitoring System is included for these subjects at the 3 and 6 month time point, to collect glycaemic variability data for comparison to the intervention group of the study (Sensor glucose measurements not visible during this time). All subjects will wear a masked Abbott Sensor Based Glucose Monitoring System, for 14 days prior to randomisation (Sensor glucose measurements not visible during this time). |
|
| Secondary | Number of Glucose Measurements Performed | Number of blood glucose fingerstick tests per day by intervention and control group during baseline (days 1 to 15) and days 194 to 208. The number of sensor scans performed performed by the intervention group during days 15 to 208. | | Posted | | Mean | Standard Deviation | number of measurements per day | | Days 1 to 208 | | | | ID | Title | Description |
|---|
| OG000 | Sensor Based Glucose Monitoring System | Standard sensing system use for 6 months. Sensor Based Glucose Monitoring System: Subjects will wear the Abbott Sensor Based Glucose Monitoring System (unmasked) for 6 months to monitor their glucose levels. All subjects will wear a masked Abbott Sensor Based Glucose Monitoring System, for 14 days prior to randomisation (Sensor glucose measurements not visible during this time). | | OG001 | Standard Blood Glucose Monitoring | Subjects randomised to the control group will be given blood glucose meters for monitoring for the 6 months study duration. Standard Blood Glucose Monitoring: Subjects will use an Abbott Blood Glucose Monitoring System (standard blood glucose meter) for 6 months to monitor their glucose levels. A 14-day masked wear of the Abbott Sensor Based Glucose Monitoring System is included for these subjects at the 3 and 6 month time point, to collect glycaemic variability data for comparison to the intervention group of the study (Sensor glucose measurements not visible during this time). All subjects will wear a masked Abbott Sensor Based Glucose Monitoring System, for 14 days prior to randomisation (Sensor glucose measurements not visible during this time). |
|
| Secondary | System Utilisation | System utilisation assessed by percentage of sensor glucose data collected by the intervention group | 112 subjects included in the analysis, 7 were not included due to missing data. | Posted | | Mean | Standard Deviation | percentage of sensor glucose collected | | Days 15 to 208 | | | | ID | Title | Description |
|---|
| OG000 | Sensor Based Glucose Monitoring System | Standard sensing system use for 6 months. Sensor Based Glucose Monitoring System: Subjects will wear the Abbott Sensor Based Glucose Monitoring System (unmasked) for 6 months to monitor their glucose levels. All subjects will wear a masked Abbott Sensor Based Glucose Monitoring System, for 14 days prior to randomisation (Sensor glucose measurements not visible during this time). |
| |
| Secondary | Change in Diabetes Treatment Satisfaction Questionnaire (DTSQc) Scores From Day 1 to Day 208 | The Diabetes Treatment Satisfaction Questionnaire change (DTSQc) score is used to assess relative change in participant satisfaction from baseline. The questionnaire consists of 8 items, 6 of which (1 and 4 through 8) assess treatment satisfaction. Each item is rated on a 7-point Likert scale (which ranges from -3 (much less satisfied) to +3 (much more satisfied). The scores from the 6 treatment satisfaction items are summed to a Total Treatment Satisfaction Score, which ranges from -18 (much less satisfied) to +18 (much more satisfied). There is one question to assess the change in perceived frequency of Hypoglycaemia and one question to assess change in perceived frequency of Hyperglycaemia. Each question is rated on a 7-point Likert scale (-3 to +3), -3 (much less of the time now) to +3 (much more of the time now). The ANCOVA adjusts for baseline DTSQs (status version). | | Posted | | Mean | Standard Deviation | units on a scale | | Baseline and Day 208 | | | | ID | Title | Description |
|---|
| OG000 | Sensor Based Glucose Monitoring System | Standard sensing system use for 6 months. Sensor Based Glucose Monitoring System: Subjects will wear the Abbott Sensor Based Glucose Monitoring System (unmasked) for 6 months to monitor their glucose levels. All subjects will wear a masked Abbott Sensor Based Glucose Monitoring System, for 14 days prior to randomisation (Sensor glucose measurements not visible during this time). | | OG001 | Standard Blood Glucose Monitoring | |
|