Verify the Functions of BPM LS-802 to Comply With ANSI/AAMI SP10
Verify the Functions of BPM LS-802 to Comply With ANSI/AAMI SP10
The clinical protocol of the clinical testing of this device:
Hospital Information The data was collected by Zhongshan City People's Hospital Clinical Investigator Team at Zhongshan City People's Hospital, No.2 Sunwen Dong Road, Zhongshan, Guangdong 528403, P. R. China.
Investigator: Dr. Cao Yuedong, Leader; Li Qidong, Nurse A; Chen Wenqian, Nurse B.
Contact Dr. Cao Yuedong Tel: +86 760 88823818
Used Equipments DUT (Device Under Test): Transtek Blood Pressure Monitor, Model: LS-802. Cuff size: 22-32 cm and 22-42 cm.
Reference Device: Yuyue Medical Blood Pressure Meter, YYBP-212, accuracy: ±1 mmHg and range: 0-300 mmHg.
Test Protocol
Test Purpose:
The aim of clinical test is to determine, for the purpose of design qualification, the overall performance of the system meeting the following requirement: For systolic and diastolic pressures, treated separately, the mean difference of the paired measurement of the test system and the comparison system shall be ± 5 mmHg or less, with standard deviation of 8 mmHg or less.
Target Subject:
1) The subject database shall contain at least 85 subjects. 2) The distribution of blood pressure of the subjects required to be: At least 10 percent below 100 mmHg systolic; at least 10 percent above 160 mmHg systolic; at least 10 percent below 60 mmHg diastolic; at least 10 percent above 100 mmHg diastolic; with the remainder distributed between these outer limits.
3. Identification and precision of equipment used According to ANSI/AAMI SP10, the maximum measurement error of the non-automatic sphygmomanometers used in the comparing test shall not exceed 1.0mmHg at the temperature of the test. The manometer was used to verify its accuracy before test.
The clinical study was conducted according to the protocol described in the ANSI/AAMI SP10 with 95 subjects.
4. Test procedures (auscultatory method, same-limb sequential measurement was chosen)
1) No motion and speaking are allowed during the measurement. 2) DUT uses the 22-42 cm Cuff when the arm circumference of patients above 32 cm and the other patients use 22-32 cm Cuff.
3) The reference device's cuff release at the rate about 3 mmHg/s so to ensure an accurate reading of observer.
4) The arm of the same height and heart when measurement. 5) Test environment: Temperature: 20±1 ℃; Relative humidity: 40~50%.
Inclusion Criteria:
Exclusion Criteria: