Not provided
Not provided
| ID | Type | Description | Link |
|---|---|---|---|
| BTS-TRANS01 | Other Identifier | BTSInternational |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Class |
|---|---|
| West China Hospital | OTHER |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
The clinical protocol of the clinical testing of this device:
Hospital Information The data was collected by West China Medical College Clinical Investigator Team at Wuhou District Hospital, No. 9 Tongzilin Road, Wuhou District, Chengdu 610041, P.R. China.
Investigator: Catharine Zhang, Nurse A; Vivian Luo, Nurse B; Dr. James Zhu, Sponsor.
Contact Dr. James Zhu Tel: +8613308036568 Used Equipments DUT (Device Under Test): Transtek Blood Pressure Monitor, TMB-986 Reference Device: Yuyue medical BP meter, BP212, accuracy: ±1mmHg and range: 0-300mmHg.
Test Protocol
Test Purpose:
The aim of clinical test is to determine, for the purpose of design qualification, the overall performance of the system meeting the following requirement: For systolic and diastolic pressures, treated separately, the mean difference of the paired measurement of the test system and the comparison system shall be ± 5mmHg or less, with standard deviation of 8mmHg or less.
Target Subject:
1) The subject database shall contain at least 85 subjects. 2) The distribution of blood pressure of the subjects required to be: At least 10 percent below 100mmHg systolic; at least 10 percent above 160mmHg systolic; at least 10 percent below 60mmHg diastolic; at least 10 percent above 100mmHg diastolic; with the remainder distributed between these outer limits.
3. Identification and precision of equipment used According to ANSI/AAMI SP10, the maximum measurement error of the non-automatic sphygmomanometers used in the comparing test shall not exceed 1.0mmHg at the temperature of the test. The manometer was used to verify its accuracy before test.
The clinical study was conducted according to the protocol described in the ANSI/AAMI SP10 with 95 subjects.
4. Test procedures (auscultatory method, same-limb sequential measurement was chosen)
Fig. 0 Method of auscultatory validation 4) As shown in Fig. 0, two observers shall make simultaneous, blinded, blood pressure determination on the subject's left arm. Each nurse will write down the reading on a small piece of paper, make a tick if there is an auscultatory gap and then submit to the recorder.
5) The two readings separately from two observers will be recorded, and if two observers agree that they find an auscultatory gap this time, it will also be recorded. The arm cuff will be take off, after 90 seconds rest, TMB-986 cuff will be taken on and tested, and the readings (systolic, diastolic, mean BP and Heart rate) be recorded.
6) For each subject, repeat 3) and 5) to get 3 measurements. 7) Repeat step 1) to 6) for every subject. 5. Note:
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Blood Pressure, Heart Rate, Monitor |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| Comparison test | Device | Yuyue medical BP meter, accuracy: ±1mmHg and range: 0-300mmHg. |
|
| Measure | Description | Time Frame |
|---|---|---|
| Verify the accuracy of measure functions of device | Verify the accuracy of device if meet the requirements of ANSI/AAMI SP10 when it measure Systolic & Diastolic blood pressure. | 10 days |
Not provided
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
The subject population included patients suffering from hypertension, labile hypertension, normal, and hypotension. The 95 patients had a wide range in age, weight, height, heart rate, systolic and diastolic blood pressures.
18~99 years old. Male or Female. Arm circumference in 22-42cm.
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Guoqing Li, Director | Wuhou District Hospital | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| Wuhou District Hospital | Chengdu | Sichuan | 610041 | China |
| PubMed Identifier | Type | Citation | Retractions |
|---|---|---|---|
| 20559140 | Background | Alpert BS. Validation of the Tiba Medical Ambulo 2400 ambulatory blood pressure monitor to the ISO Standard and BHS protocol. Blood Press Monit. 2010 Oct;15(5):275-7. doi: 10.1097/MBP.0b013e32833c8b39. |
Not provided
Not provided
| Type | Date | Date Unknown |
|---|---|---|
| Release | Feb 12, 2011 | |
| Reset | Mar 7, 2011 | |
| Release | Sep 25, 2011 | |
| Reset | Oct 26, 2011 | |
| Release | Dec 19, 2011 | |
| Unrelease | Yes | |
| Release | Dec 21, 2011 | |
| Unrelease | Yes | |
| Release | Dec 22, 2011 | |
| Unrelease | Yes | |
| Release | Dec 23, 2011 | |
| Reset | Jan 30, 2012 | |
| Release | Jul 10, 2012 | |
| Reset | Aug 17, 2012 | |
| Release | Aug 13, 2013 | |
| Reset | Oct 16, 2013 | |
| Release | Jan 26, 2015 | |
| Reset | Feb 10, 2015 | |
| Release | Jul 25, 2017 | |
| Reset | Feb 5, 2018 | |
| Release | Jun 23, 2018 | |
| Unrelease | Jun 24, 2018 | |
| Release | Jun 24, 2018 | |
| Reset | Jan 11, 2019 |
Not provided
Not provided
| Release Date | Unrelease Date | Unrelease Date Unknown | Reset Date | MCP Release Number |
|---|---|---|---|---|
| Feb 12, 2011 | Mar 7, 2011 | |||
| Sep 25, 2011 |
| ID | Term |
|---|---|
| D006973 | Hypertension |
| ID | Term |
|---|---|
| D014652 | Vascular Diseases |
| D002318 | Cardiovascular Diseases |
Not provided
Not provided
Not provided
Not provided
Not provided
| Oct 26, 2011 |
| Dec 19, 2011 | Yes |
| Dec 21, 2011 | Yes |
| Dec 22, 2011 | Yes |
| Dec 23, 2011 | Jan 30, 2012 |
| Jul 10, 2012 | Aug 17, 2012 |
| Aug 13, 2013 | Oct 16, 2013 |
| Jan 26, 2015 | Feb 10, 2015 |
| Jul 25, 2017 | Feb 5, 2018 |
| Jun 23, 2018 | Jun 24, 2018 |
| Jun 24, 2018 | Jan 11, 2019 |