| Name | Class |
|---|---|
| Juvenile Diabetes Research Foundation | OTHER |
| Becton, Dickinson and Company | INDUSTRY |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
The purpose of this study is to assess the accuracy of the Becton Dickenson (BD) Technologies Glucose Binding Protein-Based Continuous Glucose Monitor (GBP CGM) in patients with Type 1 diabetes during low (goal glucose 55 mg/dL), normal (80-140 mg/dL) and high (>180mg/dL) glucose levels over a 24 hour period. This will be achieved by monitoring blood sugar levels when a regular dinner meal is given, when a liquid breakfast meal (BOOST Original containing 41 grams Carbohydrates (CHO), 4 grams fat, 10 grams protein) is given, when subcutaneous insulin is dosed to induce hypoglycemia to a goal of 55 mg/dL, and when a regular lunch meal is given. A Continuous Glucose Monitor (CGM) is an electronic device that measures and displays blood sugar (glucose) levels in the body throughout the day and night. The method being used to detect blood sugar in the investigational Glucose Binding Protein-Based Continuous Glucose Monitor (GBP CGM) is different than the method that is currently in use by commercially available models.
Some sensors cannot tell the difference between glucose (sugar) and other substances such as Tylenol, aspirin or citric acid etc. Because they cannot tell the difference, they may give false readings. The GBP CGM is made to only recognize glucose in the body rather than other substances (e.g., Tylenol, aspirin, citric acid, etc.). As a result, the investigators expect the new GBP CGM to be more accurate at detecting low blood sugar levels than the current devices.
Development of accurate continuous glucose monitoring devices is critically important for the maintenance of strict glycemic control without increasing the risk for hypoglycemia and for the successful implementation of an artificial pancreas. Glucose binding protein-based (GBP) continuous glucose monitors offer several potential advantages over commercially available glucose oxidase based sensors. Unlike glucose oxidase based sensors, GBP sensors do not require establishment of an equilibrium reduction/oxidation reaction and are not as susceptible to interference from other compounds because they are engineered to recognize only glucose. As a result, GBP-based glucose sensors are exceptionally accurate at low blood glucose levels and have a faster warm-up time than glucose oxidase based systems. The principal idea of this proposal is to evaluate the accuracy and performance of a second generation glucose binding protein-based continuous glucose monitor (GBP CGM) developed by BD Technologies in subjects with type 1 diabetes, with emphasis on warm-up period after insertion, accuracy in hypoglycemic and hyperglycemic ranges, and lag time over a 24 hour period of sensor use. In order to accomplish this, the investigators will conduct a clinical trial consisting of 25.5-hour studies involving 15 subjects, each wearing one active and one mock (no GBP) device simultaneously during meals and induced hypoglycemia to observe a wide range of glucose values. The investigators envision that the glucose binding protein-based continuous glucose monitor will demonstrate minimal warm-up time after insertion, overall accuracy comparable to commercial glucose oxidase sensors with excellent performance in the hypoglycemic range and a nominal lag time to sensing a change in blood glucose.
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| GBP-CGM | Experimental | All participants will wear one active GBP-CGM and one inactive GBP-CGM |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| GBP CGM | Device | Visit 1: Screening visit to determine if subject qualifies for the study. Visit 2: Inpatient admission requiring a 25.5-hour hospital stay. Each subject will wear one active & one mock device simultaneously during hyperglycemic & hypoglycemic challenge conditions to observe a wide range of glucose values. Visit 3 & 4: Subjects will return to the research center approximately 24 & 48 hours after sensor removal, respectively, for evaluation of the postimplantation sensor site. Visit 5: Subjects will return to the research center approximately 28 days post inpatient admission. Blood samples for future testing of GBP and polyethylene Glycol neutralizing antibodies will be taken at Visit 1 & 5. |
| Measure | Description | Time Frame |
|---|---|---|
| Absolute Relative Difference (ARD) | ARD=100*(G_sensor-G_reference)/G_reference Calculated for when patient's G_ref was Normal (70-180 mg/dl), Hyperglycemic (>180 mg/dl) and Hypoglycemic (<70 mg/dl) The study data includes 208 paired sensor-YSI plasma glucose readings (G_reference) for each GBP CGM sensor (G_sensor) inserted for 24 hours during hyperglycemic and hypoglycemic challenge conditions. Data pairs will permit the detailed evaluation of sensor performance parameters, including static accuracy metrics such as median and mean absolute deviations and median and mean absolute relative deviation and Point CG-EGA, as well as dynamic parameters, such as warm-up time, trend accuracy (Rate CG-EGA), and sensor lag. | 25.5 hours |
Not provided
Not provided
Inclusion Criteria:
Clinical diagnosis of type 1 diabetes mellitus for ≥1 year. For an individual to be enrolled at least one criterion from each list must be met.
Criteria for documented hyperglycemia (at least 1 must be met):
Criteria for requiring insulin at diagnosis (1 must be met):
Use of an insulin pump to treat his or her diabetes for at least six months prior to the study.
Actively using a bolus calculator function with the current insulin pump with pre-defined parameters for glucose goal(s), carbohydrate ratio(s), and insulin sensitivity factor(s).
Signed informed consent
Age ≥21 and <65 years old
Body mass index between 19 and 30 kg/m2, inclusive
HbA1c ≤11%
Exclusion Criteria:
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Stacey Anderson, MD | University of Virginia, Center for Diabetes Technology | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| University of Virginia | Charlottesville | Virginia | 22908 | United States |
Not provided
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| FG000 | GBP-CGM | All participants will wear one active GBP-CGM and one inactive GBP-CGM GBP CGM: Visit 1: Screening visit to determine if subject qualifies for the study. Visit 2: Inpatient admission requiring a 25.5-hour hospital stay. Each subject will wear one active & one mock device simultaneously during hyperglycemic & hypoglycemic challenge conditions to observe a wide range of glucose values. Visit 3 & 4: Subjects will return to the research center approximately 24 & 48 hours after sensor removal, respectively, for evaluation of the postimplantation sensor site. Visit 5: Subjects will return to the research center approximately 28 days post inpatient admission. Blood samples for future testing of GBP and polyethylene Glycol neutralizing antibodies will be taken at Visit 1 & 5. |
| Title | Milestones | Reasons Not Completed | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Overall Study |
|
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| BG000 | GBP-CGM | All participants will wear one active GBP-CGM and one inactive GBP-CGM GBP CGM: Visit 1: Screening visit to determine if subject qualifies for the study. Visit 2: Inpatient admission requiring a 25.5-hour hospital stay. Each subject will wear one active & one mock device simultaneously during hyperglycemic & hypoglycemic challenge conditions to observe a wide range of glucose values. Visit 3 & 4: Subjects will return to the research center approximately 24 & 48 hours after sensor removal, respectively, for evaluation of the postimplantation sensor site. Visit 5: Subjects will return to the research center approximately 28 days post inpatient admission. Blood samples for future testing of GBP and polyethylene Glycol neutralizing antibodies will be taken at Visit 1 & 5. |
| Units | Counts |
|---|---|
| Participants |
|
| Title | Description | Population Description | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Denominator Units Selected | Denominators | Classes |
|---|---|---|---|---|---|---|---|---|---|
| Age, Continuous | Mean |
| Type | Title | Description | Population Description | Reporting Status | Anticipated Posting Date | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Time Frame | Units Analyzed | Denominator Units Selected | Arm/Group Information | Denominators | Classes | Analyses | |||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Primary | Absolute Relative Difference (ARD) | ARD=100*(G_sensor-G_reference)/G_reference Calculated for when patient's G_ref was Normal (70-180 mg/dl), Hyperglycemic (>180 mg/dl) and Hypoglycemic (<70 mg/dl) The study data includes 208 paired sensor-YSI plasma glucose readings (G_reference) for each GBP CGM sensor (G_sensor) inserted for 24 hours during hyperglycemic and hypoglycemic challenge conditions. Data pairs will permit the detailed evaluation of sensor performance parameters, including static accuracy metrics such as median and mean absolute deviations and median and mean absolute relative deviation and Point CG-EGA, as well as dynamic parameters, such as warm-up time, trend accuracy (Rate CG-EGA), and sensor lag. | Posted | Number | percentage of error | 25.5 hours |
|
Not provided
Not provided
Not provided
| ID | Title | Description | Deaths (Affected) | Deaths (At Risk) | Serious Events (Affected) | Serious Events (At Risk) | Other Events (Affected) | Other Events (At Risk) |
|---|---|---|---|---|---|---|---|---|
| EG000 | GBP-CGM | All participants will wear one active GBP-CGM and one inactive GBP-CGM GBP CGM: Visit 1: Screening visit to determine if subject qualifies for the study. Visit 2: Inpatient admission requiring a 25.5-hour hospital stay. Each subject will wear one active & one mock device simultaneously during hyperglycemic & hypoglycemic challenge conditions to observe a wide range of glucose values. Visit 3 & 4: Subjects will return to the research center approximately 24 & 48 hours after sensor removal, respectively, for evaluation of the postimplantation sensor site. Visit 5: Subjects will return to the research center approximately 28 days post inpatient admission. Blood samples for future testing of GBP and polyethylene Glycol neutralizing antibodies will be taken at Visit 1 & 5. |
Not provided
| Term | Organ System | Source Vocabulary | Assessment Type | Notes | Statistical Information |
|---|---|---|---|---|---|
| Erythema (score=2) | Skin and subcutaneous tissue disorders | Systematic Assessment |
Not provided
| Title | Organization | Phone | Extension | |
|---|---|---|---|---|
| Stacey Anderson, MD | University of Virginia | sg4c@virginia.edu |
Not provided
| ID | Term |
|---|---|
| D003922 | Diabetes Mellitus, Type 1 |
| ID | Term |
|---|---|
| D003920 | Diabetes Mellitus |
| D044882 | Glucose Metabolism Disorders |
| D008659 | Metabolic Diseases |
| D009750 | Nutritional and Metabolic Diseases |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
|
|
| years |
|
| Sex: Female, Male | Count of Participants | Participants |
|
| Duration of Diabetes | Time since diagnosis of Type 1 Diabetes | Mean | Standard Deviation | years |
|
|
|
| 0 |
| 16 |
| 4 |
| 16 |
Not provided
Not provided
Not provided
| D004700 | Endocrine System Diseases |
| D001327 | Autoimmune Diseases |
| D007154 | Immune System Diseases |