According to Section 7.3 and Section 7.4, ISO 15197(2003), it adopts the whole blood of fingertips and vein as samples to evaluate the accuracy of portable BGM.
The clinical medical technologists collected blood samples from 308 volunteers including 87 diabetics and 221 healthy people. The range of the blood glucose concentration was from 41.2mg/dL to 521mg/dL. The result of clinic evaluation shows that the accuracy of BGM matches the requirement of ISO15197(2003).
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Measure | Description | Time Frame |
|---|---|---|
| glucose | 10 min |
Not provided
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
The glucose concentrations of volunteer shall be distributed as specified in ISO15197 Table 3
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Feng-Huei Lin, PhD | Biomedical Engineering Department of National Taiwan University Hospital | Study Director |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| National Taiwan University Hospital | Taipei | Taiwan |