Glucose Monitor Accuracy Investigation (Ⅱ)
Glucose Monitor Accuracy Investigation (Ⅱ)
According to Section 7.3 and Section 7.4, ISO 15197(2003), it adopts the whole blood of fingertips as samples to evaluate the accuracy of portable BGM.
The clinical medical technologists collected blood samples from 319 volunteers including 90 diabetics and 229 healthy people. The range of the blood glucose concentration was from 30.2mg/dL to 577.5mg/dL. The result of clinic evaluation shows that the accuracy of BGM matches the requirement of ISO15197(2003).
Inclusion Criteria:
Exclusion Criteria: