| Primary | Average Percentage Improvement From Baseline of the 6-minute Walk Test Distance and Minnesota Living With Heart Failure Quality of Life Score at 6-months | Combined, average percentage improvement in 6-minute walk test distance and Minnessota Living With Heart Failure Quality of Life score from baseline to 6-month follow-up for the Protos DR/CLS (Group 1) and Stratos LV (Group 2) compared with the active control (Group 3). The 6-minute walk test is a test that measure how far a patient can walk in 6 minutes in a standardized walking course. Percent Change (0% (worst)-100% (best)) | Patients included in this analysis are those patients with complete six minute walk test and Quality of Life data at both baseline and the six-month follow-up. | Posted | | Mean | Standard Error | Percent Change | | Change from baseline to six months post-procedure | | | | ID | Title | Description |
|---|
| OG000 | biV-pacing With CLS Rate Adaption (Protos CLS) | Patients are implanted with a Protos CLS device, using the atrial port of the device for the left ventricular lead and the ventricular port for the ventricular lead to provide biV pacing. CLS is activated for rate adaptation. | | OG001 | biV Pacing With Accelerometer Based Rate Adaption (Stratos LV) | Patients are implanted with a Stratos LV device, with the atrial port plugged and the left ventricular port for the left ventricular lead and the right ventricular port for the right ventricular lead to provide biV pacing. The acceleromer is activated for rate adaptation. | | OG002 | RV Pacing With Accelerometer Based Rate Adaption (Stratos LV) | Patients are implanted with a Stratos LV device, with the atrial port plugged and the left ventricular port for the left ventricular lead and the right ventricular port for the right ventricular lead to provide biV pacing. The device is programmed to right ventricular pacing only. The acceleromer is activated for rate adaptation. |
| | | Title | Denominators | Categories |
|---|
| | | Title | Measurements |
|---|
| - OG00040.1± 5.7
- OG00142.5± 8.9
- OG00231.5± 9.0
|
|
| | Group IDs | Group Description | Statistical Method | Statistical Comment | P-Value | P-Value Comment | Parameter Type | Parameter Value | Dispersion Type | Dispersion Value | Confidence Interval Sides | Confidence Interval % | CI Lower Limit | CI Upper Limit | CI Lower Limit Comment | CI Upper Limit Comment | Estimate Comment | Tested Non-Inferiority | Non-Inferiority Type | Non-Inferiority Comment | Other Analysis Description |
|---|
| | t-test, 2 sided | | 0.437 | | | | | | | | | | | | | | Superiority | | | | | t-test, 2 sided | | 0.405 | | |
|
| Primary | Number of Participants Free of System-related Complications (Related to the Device, Leads, or Implant Procedure) at 6 Months Post-procedure | Complication-free rate was evaluated in an equivalence (non-inferiority) format compared to a target of 85% minus delta (10%), where delta is the clinically significant difference for establishing equivalence. This endpoint evaluated system-related complications. | | Posted | | Count of Participants | | Participants | | At six months post-procedure | | | | ID | Title | Description |
|---|
| OG000 | Protos DR/CLS System | Protos DR/CLS (biV pacing with CLS rate adaptation) |
| |
| Primary | Number of Participants Free of System-related Complications (Related to the Device, Leads, or Implant Procedure) at 6 Months Post-procedure | Complication-free rate was evaluated in an equivalence (non-inferiority) format compared to a target of 85% minus delta (10%), where delta is the clinically significant difference for establishing equivalence. This endpoint evaluated system-related complications. All Stratos systems (in both the biV and RV pacing arms) were evaluated together as pre-specified in the protocol. | | Posted | | Count of Participants | | Participants | | At six months post-procedure | | | | ID | Title | Description |
|---|
| OG000 | Stratos LV System | Stratos LV (biV or RV pacing groups with accelerometer based rate adaptation) |
| |
| Secondary | Change in Six-minute Walk Test | | Subjects with paired baseline and 6-month six-minute walk test data. | Posted | | Mean | Standard Error | Meters | | Change from baseline to 6 months post-procedure | | | | ID | Title | Description |
|---|
| OG000 | biV-pacing With CLS Rate Adaption (Protos CLS) | Biventricular pacing group with Closed Loop Stimulation rate adaptation (Protos CLS device) | | OG001 | biV Pacing With Accelerometer Based Rate Adaption (Stratos LV) | Biventricular pacing group with accelerometer based rate adaption (Stratos LV device) | | OG002 | RV Pacing With Accelerometer Based Rate Adaption (Stratos LV) | Right Ventricular pacing group with accelerometer based rate adaption (Stratos LV device) |
| |
| Secondary | Change in Quality Of Life (QOL) Score Over 6 Months Calculated as QOL Score at Baseline - QOL Score at 6 Months | Quality of Life was evaluated using the Minnesota Living with Heart Failure (MLHF) Quality of Life (QOL) Questionnaire.The questionnaire consists of 21 questions to measure the subjects' perception of how their HF and its treatment affected their ability to live as they wanted during the last month. The questions describe different ways in which some people are affected (i.e. physical, socioeconomic, and psychological impairments). If a question does not apply to a subject or is not related to their HF, then they can answer with a 0. If it does apply to them, then they can rate (from 1 to 5) how much it has affected them. From the 21 questions, the lowest possible total score is 0, and the highest possible total score is 105. A lower score is desirable. This outcome was calculated as QOL score at baseline minus QOL score at 6 months. Therefore, a positive change in QOL score represents an improvement in quality of life, while a negative change in QOL score represents a worsening. | Subjects with paired baseline and 6-month QOL data. A positive score represents improvement from baseline to 6-months. | Posted | | Mean | Standard Error | units on a scale | | Change from baseline to six months post-procedure | | | | ID | Title | Description |
|---|
| OG000 | biV-pacing With CLS Rate Adaption (Protos CLS) | Biventricular pacing group with Closed Loop Stimulation rate adaptation (Protos CLS device) | | OG001 | biV Pacing With Accelerometer Based Rate Adaption (Stratos LV) |
|
| Secondary | Cardiac Remodeling Assessments by Echocardiography - Change in Left Ventricular Ejection Fraction | | Subjects with paired baseline and 6-month echocardiography data. | Posted | | Mean | Standard Error | percent | | Change from baseline to six months post-procedure | | | | ID | Title | Description |
|---|
| OG000 | biV-pacing With CLS Rate Adaption (Protos CLS) | Biventricular pacing group with Closed Loop Stimulation rate adaptation (Protos CLS device) | | OG001 | biV Pacing With Accelerometer Based Rate Adaption (Stratos LV) | Biventricular pacing group with accelerometer based rate adaption (Stratos LV device) | | OG002 | RV Pacing With Accelerometer Based Rate Adaption (Stratos LV) | Right Ventricular pacing group with accelerometer based rate adaption (Stratos LV device) |
| |
| Secondary | Changes in New York Heart Association (NYHA) Classification | The purpose is to evaluate the change in the participant's NYHA classification. There are four NYHA classes: Class I: Patients with cardiac disease, but without resulting limitation of physical activity. Ordinary physical activity does not cause undue fatigue, palpitation, dyspnea, or anginal pain. Class II: Patients with cardiac disease resulting in slight limitation of physical activity. They are comfortable at rest. Ordinary physical activity results in fatigue, palpitation, dyspnea, or anginal pain. Class III: Patients with cardiac disease resulting in marked limitation of physical activity. They are comfortable at rest. Less than ordinary activity causes fatigue, palpitation, dyspnea, or anginal pain. Class IV: Patients with cardiac disease resulting in inability to carry on any physical activity without discomfort. Symptoms of cardiac insufficiency or of anginal syndrome may be present even at rest. If any physical activity is undertaken, discomfort is increased. | Subjects with paired baseline and 6-month NYHA classification data. The outcome of "Improved at least 1 Class" represents a numerically higher NYHA class at baseline than at 6 months (e.g. NYHA class III at baseline changed to NYHA class II at 6 months). | Posted | | Count of Participants | | Participants | | Change from baseline to six months post-procedure | | | | ID | Title | Description |
|---|
| OG000 | biV-pacing With CLS Rate Adaption (Protos CLS) | Biventricular pacing group with Closed Loop Stimulation rate adaptation (Protos CLS device) | | OG001 |
|
| Secondary | Percentage of Patients With Congestive Heart Failure (CHF) Related Hospitalizations | | Subjects successfully implanted with investigational system. | Posted | | Number | | percentage of participants analyzed | | At six months post-procedure | | | | ID | Title | Description |
|---|
| OG000 | biV-pacing With CLS Rate Adaption (Protos CLS) | Biventricular pacing group with Closed Loop Stimulation rate adaptation (Protos CLS device) | | OG001 | biV Pacing With Accelerometer Based Rate Adaption (Stratos LV) | Biventricular pacing group with accelerometer based rate adaption (Stratos LV device) | | OG002 | RV Pacing With Accelerometer Based Rate Adaption (Stratos LV) | Right Ventricular pacing group with accelerometer based rate adaption (Stratos LV device) |
| |
| Secondary | Mortality Rate | | | Posted | | Number | | percentage of participants analyzed | | At six months post-procedure | | | | ID | Title | Description |
|---|
| OG000 | biV-pacing With CLS Rate Adaption (Protos CLS) | Biventricular pacing group with CLS rate adaptation (Protos CLS device) | | OG001 | biV Pacing With Accelerometer Based Rate Adaption (Stratos LV) | Biventricular pacing group with accelerometer based rate adaption (Stratos LV device) | | OG002 | RV Pacing With Accelerometer Based Rate Adaption (Stratos LV) | Right Ventricular pacing group with accelerometer based rate adaption (Stratos LV device) |
| |
| Secondary | Cardiac Remodeling Assessments by Echocardiography - Change in Left Atrial Volume | | Subjects with paired baseline and 6-month echocardiography data. | Posted | | Mean | Standard Error | mL | | Change from baseline to six months post-procedure | | | | ID | Title | Description |
|---|
| OG000 | biV-pacing With CLS Rate Adaption (Protos CLS) | Biventricular pacing group with Closed Loop Stimulation rate adaptation (Protos CLS device) | | OG001 | biV Pacing With Accelerometer Based Rate Adaption (Stratos LV) | Biventricular pacing group with accelerometer based rate adaption (Stratos LV device) | | OG002 | RV Pacing With Accelerometer Based Rate Adaption (Stratos LV) | Right Ventricular pacing group with accelerometer based rate adaption (Stratos LV device) |
| |
| Secondary | Cardiac Remodeling Assessments by Echocardiography - Change in Left Ventricular Mass | | Subjects with paired baseline and 6-month echocardiography data. | Posted | | Mean | Standard Error | grams | | Change from baseline to six months post-procedure | | | | ID | Title | Description |
|---|
| OG000 | biV-pacing With CLS Rate Adaption (Protos CLS) | Biventricular pacing group with Closed Loop Stimulation rate adaptation (Protos CLS device) | | OG001 | biV Pacing With Accelerometer Based Rate Adaption (Stratos LV) | Biventricular pacing group with accelerometer based rate adaption (Stratos LV device) | | OG002 | RV Pacing With Accelerometer Based Rate Adaption (Stratos LV) | Right Ventricular pacing group with accelerometer based rate adaption (Stratos LV device) |
| |
| Secondary | Cardiac Remodeling Assessments by Echocardiography - Change in Left Ventricular End Systolic Volume | | Subjects with paired baseline and 6-month echocardiography data. | Posted | | Mean | Standard Error | mL | | Change from baseline to six months post-procedure | | | | ID | Title | Description |
|---|
| OG000 | biV-pacing With CLS Rate Adaption (Protos CLS) | Biventricular pacing group with Closed Loop Stimulation rate adaptation (Protos CLS device) | | OG001 | biV Pacing With Accelerometer Based Rate Adaption (Stratos LV) | Biventricular pacing group with accelerometer based rate adaption (Stratos LV device) | | OG002 | RV Pacing With Accelerometer Based Rate Adaption (Stratos LV) | Right Ventricular pacing group with accelerometer based rate adaption (Stratos LV device) |
| |
| Secondary | Cardiac Remodeling Assessments by Echocardiography - Change in Left Ventricular End Diastolic Volume | | Subjects with paired baseline and 6-month echocardiography data. | Posted | | Mean | Standard Error | mL | | Change from baseline to six months post-procedure | | | | ID | Title | Description |
|---|
| OG000 | biV-pacing With CLS Rate Adaption (Protos CLS) | Biventricular pacing group with Closed Loop Stimulation rate adaptation (Protos CLS device) | | OG001 | biV Pacing With Accelerometer Based Rate Adaption (Stratos LV) | Biventricular pacing group with accelerometer based rate adaption (Stratos LV device) | | OG002 | RV Pacing With Accelerometer Based Rate Adaption (Stratos LV) | Right Ventricular pacing group with accelerometer based rate adaption (Stratos LV device) |
| |