The purpose of the study is to verify that BulbiCAM®, a video-oculography medical device, gives similar results regardless of the Operating System (OS) used. The study compares three platforms: Windows, Linux Ubuntu, and Linux NixOS.
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Ubuntu-NixOS-Windows | Experimental | order of tested platforms |
|
| Ubuntu - Windows - NixOS | Experimental | order of tested platforms |
|
| Windows - Ubuntu - NixOS | Experimental | order of tested platforms |
|
| Windiws - NixOS - Ubuntu | Experimental | order of tested platforms |
|
| NixOS - Ubuntu - Windows | Experimental | order of tested platforms |
|
| NixOS - Windows - Ubuntu |
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| administration of 15 neuro-ophtalmological tests per platform by video-oculography | Device | List of the 15 tests per platform:
|
| Measure | Description | Time Frame |
|---|---|---|
| No statistical difference between acuracy of horizontal pro-saccades when comparing the three platforms | unit: % | single visit at enrollment |
| No statistical difference between peak velocity of horizontal pro-saccades when comparing the three platforms | unit: degree/s | Single visit at enrollment |
| No statistical difference between acuracy of multidirectional pro-saccades when comparing the three platforms | Unit: % | Single visit at enrollment |
| No statistical difference between peak velocity of multidirectional pro-saccades when comparing the three platforms | unit: degree/s | Single visit at enrollment |
| No statistical difference between contrast sensitivity results when comparing the three platforms | unit: Pelli-Robson | Single visit at enrollment |
| No statistical difference between variable strobe test results when comparing the three platforms | unit: Pelli-Robson | Single visit at enrollment |
| No statistical difference between strobe contrast sensitivity results when comparing the three platforms | unit: Pelli-Robson | Single visit at enrollment |
| Measure | Description | Time Frame |
|---|---|---|
| No statistical difference between pupil areas when comparing Windows versus Ubuntu using the corrective formulas for signal processing enhancement, modified stimulus and timing calibration | Ubuntu and NixOS platforms contain an enhanced algorithm incorporating Fourier transform analysis of pupil data. The light used as a stimulus for the pupil dynamics test has been modified (in intensity, wavelength, and spatial characteristics) under Ubuntu/NixOS platform compared to the original stimulus under Windows A timing calibration was implemented for Ubuntu and NixOS to improve accuracy results by accounting for the device delay. unit: mm*s |
Not provided
Inclusion Criteria:
Exclusion Criteria:
Ophthalmic:
Neurological/Medical:
Other:
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| Bulbitech | Trondheim | Norway |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Each subject will be tested on each of the three platform and serve as its own comparator.
Not provided
Not provided
Not provided
Not provided
order of tested platforms |
|
|
| No statistical difference between RAPD test result when comparing the three platforms |
no unit |
| Single visit at enrollment |
| No statistical difference between ptosis MRD1 and 2 when comparing the three platforms | unit mm | Single visit at enrollment |
| No statistical difference between duration of fixation when comparing the three platforms | Unit: ms | Single visit at enrollment |
| No statistical difference between the amplitude of intrusive saccades when comparing the three platforms | Unit: degree | Single visit at enrollment |
| No statistical difference between the peak velocity of intrusive saccades when comparing the three platforms | unit: degree/s | Single visit at enrollment |
| No statistical difference between the mean fixation time when comparing the three platforms | unit: ms | Single visit at enrollment |
| No statistical difference between mean pupil size when comparing the three platforms | unit: mm | Single visit at enrollment |
| No statistical difference between anti-saccade amplitudes when comparing the three platforms | unit: degree | Single visit at enrollment |
| No statistical difference between anti-saccades peak velocity when comparing the three platforms | unit: degree/s | Single visit at enrollment |
| No statistical difference between memory saccades accuracy when comparing the three platforms | unit: % | Single visit at enrollment |
| No statistical difference between number of errors in memory saccades task when comparing the three platforms | no unit | Single visit at enrollment |
| No statistical difference between memory saccades latencies when comparing Ubuntu and NixOS platforms | unit: ms | Single visit at enrollment |
| No statistical difference between anti saccades latencies when comparing Ubuntu and NixOS platforms | unit: ms | Single visit at enrollment |
| No statistical difference between saccadic reaction time in visual field test when comparing Ubuntu and NixOS platforms | unit: ms | Single visit at enrollment |
| No statistical difference between pupil area when comparing Ubuntu and NixOS platforms | unit: mm*s | Single visit at enrollment |
| No statistical difference between pupil diameters when comparing Ubuntu and NixOS platforms | unit: mm | Single visit at enrollment |
| No statistical difference between pupil contraction latencies when comparing Ubuntu and NixOS platforms | unit: ms | Single visit at enrollment |
| No statistical difference between peak velocity of pupil dilation when comparing Ubuntu and NixOS platforms | unit: mm/s | Single visit at enrollment |
| No statistical difference between multidirectional pro-saccade latencies when comparing Ubuntu and NixOS platforms | unit: ms | Single visit at enrollment |
| No statistical difference between horizontal pro-saccades latencies when comparing Ubuntu and NixOS platforms | unit: ms | Single visit at enrollment |
| No statistical difference between smooth pursuit gain when comparing Ubuntu and NixOS platforms | unit: % | Single visit at enrollment |
| No statistical difference between the calibration ratio when comparing the three platforms | no unit | Single visit at enrollment |
| No statistical difference between pursuit acuity results when comparing Windows versus NixOS platforms | unit: LogMAR | Single visit at enrollment |
| Single visit at enrollment |
| No statistical difference between pupil diameters when comparing Windows versus Ubuntu using the corrective formulas for signal processing enhancement, modified stimulus and timing calibration | Ubuntu and NixOS platforms contain an enhanced algorithm incorporating Fourier transform analysis of pupil data. The light used as a stimulus for the pupil dynamics test has been modified (in intensity, wavelength, and spatial characteristics) under Ubuntu/NixOS platform compared to the original stimulus under Windows A timing calibration was implemented for Ubuntu and NixOS to improve accuracy results by accounting for the device delay. unit: mm | Single visit at enrollment |
| No statistical difference between latency of pupil contraction when comparing Windows versus Ubuntu using the corrective formulas for signal processing enhancement, modified stimulus and timing calibration | Ubuntu and NixOS platforms contain an enhanced algorithm incorporating Fourier transform analysis of pupil data. The light used as a stimulus for the pupil dynamics test has been modified (in intensity, wavelength, and spatial characteristics) under Ubuntu/NixOS platform compared to the original stimulus under Windows A timing calibration was implemented for Ubuntu and NixOS to improve accuracy results by accounting for the device delay. unit: ms | Single visit at enrollment |
| No statistical difference between smooth pursuit gain when comparing Windows versus Ubuntu using the corrective formulas for timing calibration | A timing calibration was implemented for Ubuntu and NixOS to improve accuracy results by accounting for the device delay. unit: % | Single visit at enrollment |
| No statistical difference between horizontal pro-saccade latency when comparing Windows versus Ubuntu using the corrective formulas for timing calibration | A timing calibration was implemented for Ubuntu and NixOS to improve accuracy results by accounting for the device delay. Unit: ms | Single visit at enrollment |
| No statistical difference between multidirectional pro-saccade latencies when comparing Windows versus Ubuntu using the corrective formulas for timing calibration. | A timing calibration was implemented for Ubuntu and NixOS to improve accuracy results by accounting for the device delay. Unit: ms | Single visit at enrollment |
| No statistical difference between saccadic reaction times in visual field test when comparing Windows versus Ubuntu using the corrective formulas for timing calibration. | A timing calibration was implemented for Ubuntu and NixOS to improve accuracy results by accounting for the device delay. Unit: ms | Single visit at enrollment |
| No statistical difference between anti-saccade latencies when comparing Windows versus Ubuntu using the corrective formulas for timing calibration | A timing calibration was implemented for Ubuntu and NixOS to improve accuracy results by accounting for the device delay. unit: ms | Single visit at enrollment |
| No statistical difference between memory saccades latencies when comparing Windows versus Ubuntu using the corrective formulas for timing calibration | A timing calibration was implemented for Ubuntu and NixOS to improve accuracy results by accounting for the device delay. unit: ms | Single visit at enrollment |
| No statistical difference between peak velocity of pupil dilation when comparing Windows versus Ubuntu using the corrective formulas for signal processing enhancement and modified stimulus | Ubuntu and NixOS platforms contain an enhanced algorithm incorporating Fourier transform analysis of pupil data. The light used as a stimulus for the pupil dynamics test has been modified (in intensity, wavelength, and spatial characteristics) under Ubuntu/NixOS platform compared to the original stimulus under Windows. Unit: mm/s | Single visit at enrollment |
| No statistical difference between pupil area when comparing Windows versus NixOS using the corrective formulas for signal processing enhancement, modified stimulus and timing calibration | Ubuntu and NixOS platforms contain an enhanced algorithm incorporating Fourier transform analysis of pupil data. The light used as a stimulus for the pupil dynamics test has been modified (in intensity, wavelength, and spatial characteristics) under Ubuntu/NixOS platform compared to the original stimulus under Windows A timing calibration was implemented for Ubuntu and NixOS to improve accuracy results by accounting for the device delay. Unit: mm*s | Single visit at enrollment |
| No statistical difference between pupil diameters when comparing Windows versus NixOS using the corrective formulas for signal processing enhancement, modified stimulus and | Ubuntu and NixOS platforms contain an enhanced algorithm incorporating Fourier transform analysis of pupil data. The light used as a stimulus for the pupil dynamics test has been modified (in intensity, wavelength, and spatial characteristics) under Ubuntu/NixOS platform compared to the original stimulus under Windows A timing calibration was implemented for Ubuntu and NixOS to improve accuracy results by accounting for the device delay. unit: mm | Single visit at enrollment |
| No statistical difference between latencies of pupil contraction when comparing Windows versus NixOS using the corrective formulas for signal processing enhancement, modified stimulus and timing calibration | Ubuntu and NixOS platforms contain an enhanced algorithm incorporating Fourier transform analysis of pupil data. The light used as a stimulus for the pupil dynamics test has been modified (in intensity, wavelength, and spatial characteristics) under Ubuntu/NixOS platform compared to the original stimulus under Windows A timing calibration was implemented for Ubuntu and NixOS to improve accuracy results by accounting for the device delay. Unit: ms | Single visit at enrollment |
| No statistical difference between smooth pursuit gain when comparing Windows versus NixOS using the corrective formulas for timing calibration | A timing calibration was implemented for Ubuntu and NixOS to improve accuracy results by accounting for the device delay. . Unit: % | Single visit at enrollment |
| No statistical difference between horizontal pro-saccades latencies when comparing Windows versus NixOS using the corrective formulas for timing calibration | A timing calibration was implemented for Ubuntu and NixOS to improve accuracy results by accounting for the device delay. Unit: ms | Single visit at enrollment |
| No statistical difference between multidirectional pro-saccades latencies when comparing Windows versus NixOS using the corrective formulas for timing calibration | A timing calibration was implemented for Ubuntu and NixOS to improve accuracy results by accounting for the device delay. unit: ms | Single visit at enrollment |
| No statistical difference between saccadic reaction time fo visual field test when comparing Windows versus NixOS using the corrective formulas for timing calibration | A timing calibration was implemented for Ubuntu and NixOS to improve accuracy results by accounting for the device delay. unit: ms | Single visit at enrollment |
| No statistical difference between anti-saccades latencies when comparing Windows versus NixOS using the corrective formulas for timing calibration | A timing calibration was implemented for Ubuntu and NixOS to improve accuracy results by accounting for the device delay. Unit: ms | Single visit at enrollment |
| No statistical difference between memory saccades latencies when comparing Windows versus NixOS using the corrective formulas for timing calibration | A timing calibration was implemented for Ubuntu and NixOS to improve accuracy results by accounting for the device delay. Unit: ms | Single visit at enrollment |
| No statistical difference between peak velocity of the pupil when comparing Windows versus NixOS using the corrective formulas for signal processing enhancement and modified stimulus | Ubuntu and NixOS platforms contain an enhanced algorithm incorporating Fourier transform analysis of pupil data. The light used as a stimulus for the pupil dynamics test has been modified (in intensity, wavelength, and spatial characteristics) under Ubuntu/NixOS platform compared to the original stimulus under Windows Unit: mm/s | Single visit at enrollment |