Prospective, observational, pre-market, verification study to confirm accuracy of the INVOS 7100 system in compliance with ISO 80601-2-85:2021 standard
Verify that the INVOS PM7100 system in conjunction with the INVOS Adult Sensors meets the requirements for tissue saturation accuracy.
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Healthy Volunteers | Healthy, non-smoking volunteers ages ≥18 to <47 years who meet inclusion criteria |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| Observational | Device | The study objectives require placement of a jugular venous bulb catheter and radial arterial line placement with ultrasound guidance. Two INVOS™ sensors (PMSENS71-A) will be placed bilaterally on the subject's forehead and two on the calf. The INVOS™ device will be evaluated for a targeted range of 70-100% SpO2 levels. Hypoxic mixtures of gas are delivered, and data are collected at approximately (~) 5-minute intervals during steady periods of increasing and decreasing oxygen concentration. |
| Measure | Description | Time Frame |
|---|---|---|
| Accuracy in terms of the root mean square error | Performance assessed by comparing INVOS™ rSO2 to the calculated global field saturation (fSO2) value obtained from simultaneous arterial and jugular venous blood samples during normocapnia. | Performance evaluation is expected take approximately 2.5 hours |
| Measure | Description | Time Frame |
|---|---|---|
| Trend accuracy |
| Evaluation is expected take approximately 2.5 hours |
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Healthy, non-smoking volunteers ages of ≥18 to <47 years who meet inclusion criteria
Not provided
| Name | Affiliation | Role |
|---|---|---|
| David MacLeod, MDDS | Duke University | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| Duke University Medical Center | Durham | North Carolina | 27710 | United States |
Not provided
| ID | Term |
|---|---|
| D019370 | Observation |
| ID | Term |
|---|---|
| D008722 | Methods |
| D008919 | Investigative Techniques |
Not provided
Not provided
Not provided
Not provided
Not provided
|
| Trend accuracy in a hypercapneic state | Device performance will be evaluated during hypercapnia | Evaluation is expected take approximately 2.5 hours |