The purpose of this study it to evaluate the accuracy of heart rate, heart rate variability, breathing rate and blood pressure measurement with the use of Shen.AI software.
The primary objective of this clinical study is to assess the accuracy of heart rate (HR), heart rate variability (HRV), breathing rate (BR), and blood pressure (BP) measurements obtained using Shen.AI software as the investigational device under its intended conditions of use, enabling users to obtain reliable information about their physiological state. These conditions include measurements taken at rest, in a home-like environment, with participants independently operating a mobile device or laptop equipped with a camera.
Physiological parameters, including HR, HRV, BR, and BP, will be measured and estimated using remote photoplethysmography (rPPG) based on a 1-minute facial video analysis processed by Shen.AI software.
Throughout all measurements, participants will be connected to a reference device that records ECG while simultaneously measuring chest impedance (impedance pneumography). Additionally, each participant will wear a respiratory belt transducer. Depending on the measurement type, the participant's blood pressure will be assessed using either an auscultatory or an automatic blood pressure monitor. The study will be conducted on a diverse group of participants, varying in blood pressure, gender, age, skin phototype, and BMI.
For each measured parameter, results obtained from both methods will be compared using Bland-Altman analysis or non-parametric limits of agreement.
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| HR, HRV, BR, and BP measured using Shen.AI and reference methods | Experimental |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| Shen.AI | Device | Contactless measurement of heart rate, heart rate variability, breathing rate, and blood pressure using Shen.AI software and reference methods. |
|
| Measure | Description | Time Frame |
|---|---|---|
| Accuracy of the average heart rate (HR) measurement using the device under test compared to the reference measurement. | Average HR will be estimated using Shen.AI software and compared with simultaneously recorded ECG . | One appointment (day 1) |
| Accuracy of the measurement of the heart rate variability index - Standard Deviation of Normal-to-Normal intervals (SDNN) using the device under test compared to the reference measurement. | The heart rate variability index (SDNN) will be measured using Shen.AI software and compared with the simultaneously recorded ECG. | One appointment (day 1) |
| Accuracy of the breathing rate (BR) measurement using the device under test compared to the reference measurement. | Average BR will be estimated using Shen.AI software and compared with the recorded impedance pneumography signal in the absence of read-out capability supported by the respiratory belt transducer. | One appointment (day 1) |
| Accuracy of systolic and diastolic blood pressure (BP) measurements using the device under test in relation to the reference measurement. | Average BP will be estimated using Shen.AI software and compared with auscultatory blood pressure measurements. | One appointment (day 1) |
| Accuracy of systolic and diastolic blood pressure (BP) measurement using the device under test, taking into account the calibration of the BP measurement function, in relation to the reference measurement. | Average BP will be estimated using Shen.AI software and compared with auscultatory blood pressure measurements. Measurements will be performed during two visits, with the second visit occurring 14 ± 3 days apart. | Two appointments (14 ± 3 days apart) |
| Measure | Description | Time Frame |
|---|---|---|
| Accuracy of the measurement of other heart rate variability metrics - Root Mean Square of Successive Differences (RMSSD) using the device under test. | The heart rate variability index (RMSSD) will be measured using Shen.AI software and compared with the simultaneously recorded ECG. | One appointment (day 1) |
| Measure | Description | Time Frame |
|---|---|---|
| Utility of the software under test for self-measurement of physiological parameters, as assessed by user and observer questionnaires. | At the end of the appointment, two questionnaires will be used to assess the usability of the Shen.AI software. User Usability Questionnaire - completed by participants to evaluate the clarity of instructions, ease of use, calibration process, interface intuitiveness, perceived safety, and overall experience. Observer Usability Questionnaire - completed by study personnel to assess patient behavior, difficulties encountered during measurements, and potential usability issues. Each questionnaire consists of multiple-choice questions (e.g., "Very intuitive", "Intuitive", "Non-intuitive", "Very unintuitive") and open-ended questions. Higher-rated responses (e.g., "Very intuitive", "Intuitive") indicate better usability and user experience, while lower-rated responses suggest usability challenges or areas for improvement. Open-ended responses will be analyzed by a dedicated research team to identify key themes, usability issues, and potential improvements. |
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Justyna Chojdak-Ćukasiewicz, MD, PhD | Wroclaw Medical University | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| OVO MEDICAL Sp. z o.o. | Wroclaw | Lower Silesian Voivodeship | 50-414 | Poland | ||
| Przychodnia rehabilitacyjna R-MED |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Accuracy of blood oxygen saturation measurement using the device under test |
The blood oxygen saturation will be measured using Shen.AI software and compared with the simultaneously obtained blood oxygen measurements. |
| One appointment (day 1) |
| Accuracy of Body Mass Index (BMI) estimation using the device under test. | The BMI will be measured using Shen.AI software and compared with the values obtained from a weight scale and stadiometer. | One appointment (day 1) |
| One appointment (day 1) |
| Wroclaw |
| Lower Silesian Voivodeship |
| 53-033 |
| Poland |