To verify and evaluate the efficacy and safety of i3 CGM system manufactured by Changsha Sinocare Inc. for continuous or regular monitoring of interstitial fluid glucose level in pediatric patients with diabetes.
This clinical investigation is of a multi-center and self-controlled design.
This clinical investigation is planned to be conducted in 3 study sites, with a total of 78 subjects to be enrolled, of which no fewer than 15 cases are aged 2-5 years old, and no fewer than 60 cases are aged 6-17 years old. This clinical investigation is to verify the safety and efficacy of the investigational device. Referring to the number of pediatric patients with diabetes in the 3 study sites, it is expected that about 20 subjects will be enrolled in each site every month, and it will take about 2 months to close the site. This clinical investigation is expected to be completed in about 6 months from the first participant recruited.
The real-time glucose values measured by CGM system on the left and right abdomen are compared with the venous blood glucose value (EKF value) measured by EKF analyzer, respectively, and the consistency rate is analyzed.
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| 2-5 year old participants | Wear the investigational device continuously for 15 days. |
| |
| 6-17 year old participants | Wear the investigational device continuously for 15 days. |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| i3 CGM(continuous glucose monitor) | Device | The i3 CGM consists of a Sensor Pack (including probe, iCGM-S3), a Transmitter Pack (iCGM-t3), and an application (iCGM-APP (released version: V02)).application (i3 CGM App). The user can view their glucose data on the i3 CGM App running on a compatible mobile device. |
| Measure | Description | Time Frame |
|---|---|---|
| Accuracy | percentage of i3 CGM values in the measuring range less than 4.4 mmol/L(80 mg/dL) within ± 1.1 mmol/L(20 mg/dL) of the EKF control. | 0-15 days |
Not provided
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
2-17 year old patients with diabetes mellitus.
Not provided
Not provided
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| Hunan Children's Hospital | Changsha | Hunan | China | |||
| Children's Hospital of Soochow University |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| ID | Term |
|---|---|
| D003920 | Diabetes Mellitus |
| ID | Term |
|---|---|
| D044882 | Glucose Metabolism Disorders |
| D008659 | Metabolic Diseases |
| D009750 | Nutritional and Metabolic Diseases |
| D004700 | Endocrine System Diseases |
Not provided
Not provided
Not provided
Not provided
Not provided
|
| Suzhou |
| Jiangsu |
| China |
| Women and Children's Hospital of Ningbo University | Ningbo | Zhejiang | China |