During hemodialysis (HD) the blood of the patient pass an extracorporeal circuit that contains a dialyzer for rinsing and a venous chamber (air trap) to prevent from air embolism through the return blood into the patient. However, air traps in clinical use have limited capacity to prevent from microemboli of air to enter the return bloodline and deposit as emboli in the body such as lungs, heart and brain. The Investigator developed the air trap Emboless that was patented. In vitro studies showed significantly better reduction of microemboli contaminations than air traps compared to that in clinical use.
The present randomized clinical trial compares two different air traps used by the same patients in a cross-over design (as pairs) using the Emboless compared with the Fresenius 4008/5008 (F5008).
Chronic HD patients are randomized to perform the first HD with either their standard air trap (F5008) in the venous bloodline tubing or using the Emboless bloodline and vice versa. Each patient was included to make two paired series. A safety committee evaluates if significantly worse outcome appears especially with the Emboless, to stop the study.
During HD the microbubbles are counted by a GAMPT ultrasound device using two probes. One probe is set at the inlet side of the air trap and the second at the outlet side. The outlet side represents data of microbubbles in the blood that are entering into the patient. Comparative non-parametric paired analyses are performed between the air traps.
Monitoring of the study is performed.
During hemodialysis (HD) the blood of the patient pass an extracorporeal circuit that contains a dialyzer for rinsing and a venous chamber (air trap) to prevent from air embolism through the return blood into the patient. However, air traps in clinical use have limited capacity to prevent from microemboli of air to enter the return bloodline and deposit as emboli in the body such as lungs, heart and brain. The Investigator developed the air trap Emboless that was patented in Europe, USA and India. In vitro studies showed significantly better reduction of microemboli contaminations than air traps compared to that in clinical use.
The present randomized clinical trial compares two different air traps used by the same patients in a cross-over design (as pairs) using the Emboless compared with the Fresenius 4008/5008 (F5008).
Chronic HD patients are randomized to perform the first HD with either their standard air trap (F5008) in the venous bloodline tubing or using the Emboless bloodline and vice versa. Each patient is included to make two paired series. A maximum of 30 patients are planned. Each with 2 series of each two different sets of air traps that would give a total of 120 dialyses. A safety committee evaluates if significantly worse outcome appears especially with the Emboless.
During HD the microbubbles are counted by a GAMPT ultrasound device using two probes. One probe is set at the inlet side of the air trap and the second at the outlet side. The outlet side represents data of microbubbles in the blood that are entering into the patient. Comparative non-parametric paired analyses are performed between the air traps.
Monitoring of the study is performed.
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Standard | Active Comparator | A regular standard dialysis is performed and held as active comparator for analysis |
|
| Emboless | Experimental | Emboless tubing set is part of the extracorporeal venous bloodline instead of the venous chamber by Fresenius (see standard above). For more information see reference Jonsson et al 2023- mentioned in references. |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| Emboless | Other | Emboless has the US patent number 8894749 is compared to standard dialysis See also reference Jonsson et al 2023 referred to in the main text. |
|
| Measure | Description | Time Frame |
|---|---|---|
| Comparison of change in air micro bubbles during hemodialysis | Numbers of air micro bubbles (diameter sizes from 20-500µm diameter) are counted at inlet and outlet of the air trap with parallel probes using ultrasound GAMPT 200 device during 30minutes of HD. A calculation clarifies change in percentage of outlet versus inlet amount/size of microbubble air return into the return bloodline that is connected to the return needle placed in the vein of the patient. The size in diameter and numbers within the dialysis time/30 minutes of HD are given. | 1 year |
| Measure | Description | Time Frame |
|---|---|---|
| Comparison of change in air micro bubbles for hemodialysis versus hemodiafiltration | The Investigator aims to compare the microbubble reduction of the air traps also between the hemodialysis versus hemodiafiltration if the specific patient performs this dialysis under normal circumstances. | 1 year |
| adverse events |
Not provided
Inclusion Criteria:
Accepted are all patients performing chronic HD due to End Stage renal disease of any reason:
Exclusion Criteria:
Patients that are expected not to fulfil a whole series of two dialysis within the study such as those:
Home-haemodialysis treatments and self-care dialysis treatments are not included in the study. If such patient accepts to participate in the study, he/she has to adapt to study criteria (assisted treatment).
consecutively based on the approval of patients. Excluding those with active infections.
Not provided
Not provided
Not provided
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Ulf Forsberg, MD, PhD | Region Vasterbotten Skelleftea Hospital, Sweden | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| Region Vasterbotten, Skelleftea Sjukhus, Dialysen | Skellefteå | Sweden |
| PubMed Identifier | Type | Citation | Retractions |
|---|---|---|---|
| 27528100 | Background | Stegmayr BG. Sources of Mortality on Dialysis with an Emphasis on Microemboli. Semin Dial. 2016 Nov;29(6):442-446. doi: 10.1111/sdi.12527. Epub 2016 Aug 16. | |
| 27576675 | Background | Stegmayr B. Air contamination during hemodialysis should be minimized. Hemodial Int. 2017 Apr;21(2):168-172. doi: 10.1111/hdi.12474. Epub 2016 Aug 30. |
Not provided
Not provided
We can share data upon reasonable request.
Expected time of completing data is 1 year from now.
Reasonable request to share information such as safety analyses, repeating studies or control authorities.
Not provided
Not provided
| ID | Term |
|---|---|
| D004618 | Embolism, Air |
| ID | Term |
|---|---|
| D004617 | Embolism |
| D016769 | Embolism and Thrombosis |
| D014652 | Vascular Diseases |
| D002318 | Cardiovascular Diseases |
Not provided
Not provided
Each patient is randomized to start with either standard or intervention vice versa. Paired analyses are done.
Not provided
Not provided
Measuring data of microemboli counts is masked for the patient.
Not provided
| Standard | Other | Standard dialysis set is used as comparative |
|
If a patient experience adverse events these will be reported |
| 1 year |
| 30417399 | Background | Matsuda K, Fissell R, Ash S, Stegmayr B. Long-Term Survival for Hemodialysis Patients Differ in Japan Versus Europe and the USA. What Might the Reasons Be? Artif Organs. 2018 Dec;42(12):1112-1118. doi: 10.1111/aor.13363. Epub 2018 Nov 11. No abstract available. |
| 30973284 | Result | Forsberg U, Jonsson P, Stegmayr B. Air contamination during medical treatment results in deposits of microemboli in the lungs: An autopsy study. Int J Artif Organs. 2019 Sep;42(9):477-481. doi: 10.1177/0391398819840363. Epub 2019 Apr 11. |
| 36726427 | Result | Forsberg U, Jonsson P, Stegmayr B. Microemboli induced by air bubbles may be deposited in organs as a consequence of contamination during medical care. Clin Kidney J. 2022 Oct 6;16(1):159-166. doi: 10.1093/ckj/sfac217. eCollection 2023 Jan. |
| 36594759 | Result | Jonsson P, Stegmayr C, Stegmayr B, Forsberg U. Venous chambers in clinical use for hemodialysis have limited capacity to eliminate microbubbles from entering the return bloodline: An in vitro study. Artif Organs. 2023 Jun;47(6):961-970. doi: 10.1111/aor.14495. Epub 2023 Jan 10. |