The purpose of this clinical study is to validate the SpO2 accuracy of the Stryker Sustainability Solutions pulse oximetry sensors during motion conditions over the range of 70 -100% SaO2 as compared to arterial blood samples assessed by CO-Oximetry during conditions in which the subject is moving. The end goal is to provide supporting documentation for the SpO2 accuracy validation of the reprocessed sensors with motion indications.
In this study, the level of oxygen within the blood will be reduced in a controlled manner by reducing the concentration of oxygen the study volunteer breathes. The accuracy of a noninvasive pulse oximeter sensor will be assessed by comparison to the oxygen saturation measurements from a laboratory blood gas analyzer.
A machine will be used to induce motions of 20 mm during testing at a sine rate of 2Hz, 3Hz, and 4Hz. A third motion moves the arm a random distance of 0 to 30mm at a random speed up to 5Hz.
It is required that the Accuracy Root Mean Square (ARMS) performance of the Stryker pulse oximetry sensors will meet a specification of +/-3% with a target of +/-3% or better in motion conditions for the range of 70 - 100% SaO2 (typically, saturation is determined once with air breathing and then at three or six levels, e.g. 94%, 90%, 85%, 80%, 75% and 70% or 95%, 85% and 75% saturation for about 30-60 seconds or 60-90 seconds at each level), thereby demonstrating an acceptable SpO2 accuracy performance specification. This study should utilize a three level structure (95%, 85% and 75%).
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Experimental Arms | Experimental | Experimental: Healthy adult participants All participants are enrolled in the test group and receive the noninvasive adhesive reprocessed pulse oximeter sensors |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| Experimental Masimo RD SET SpO2 Adhesive Sensors | Device | Masimo RD SET SpO2 Adhesive Sensors, Adult (4000 Adt), Adult & Neonate (4003 Neo) |
|
| Measure | Description | Time Frame |
|---|---|---|
| Accuracy of Sensor by Comparing SpO2 Percent Blood Oxygen Saturation to SaO2 Percent Blood Oxygen Saturation Utilizing Arms Calculation Calculation [Time Frame: 1 - 5 Hours] | The outcome of this study is the accuracy of the SpO2 (percent oxygen saturation by pulse oximetry) measured by the Reprocessed Oximeter pulse oximetry sensors during motion conditions over the range of 70 - 100% saturation compared to SaO2 (percent oxygen saturation by co-oximetry from blood samples) from arterial blood samples. The accuracy is calculated using the Arms error between measured SpO2 and reference SaO2. The Arms must meet the 3% specification for each reprocessed pulse oximetry sensor style. Standard deviation of the differences is computed as the precision. Square root of the sum of the squares of bias and precision is computed as the Arms Error value. | 1 - 5 hours |
Not provided
Not provided
Inclusion Criteria:
Exclusion Criteria:
The subject is obese (BMI>30)
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Philip Bickler, MD, PhD | UCSF Hypoxia Research Laboratory | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| UCSF Hypoxia Research Laboratory | San Francisco | California | 94133 | United States |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| FG000 | Healthy Adult Participants | Experimental: Healthy adult participants All participants are enrolled in the test group and receive the noninvasive adhesive reprocessed pulse oximeter sensors Experimental Masimo RD SET SpO2 Adhesive Sensors: Masimo RD SET SpO2 Adhesive Sensors, Adult (4000 Adt), Adult & Neonate (4003 Neo) |
| Title | Milestones | Reasons Not Completed | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Overall Study |
|
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| BG000 | Healthy Adult Participants | Experimental: Healthy adult participants All participants are enrolled in the test group and receive the noninvasive adhesive reprocessed pulse oximeter sensors Experimental Masimo RD SET SpO2 Adhesive Sensors: Masimo RD SET SpO2 Adhesive Sensors, Adult (4000 Adt), Adult & Neonate (4003 Neo) |
| Units | Counts |
|---|---|
| Participants |
|
| Title | Description | Population Description | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Denominator Units Selected | Denominators | Classes |
|---|---|---|---|---|---|---|---|---|---|
| Age, Categorical | Count of Participants |
| Type | Title | Description | Population Description | Reporting Status | Anticipated Posting Date | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Time Frame | Units Analyzed | Denominator Units Selected | Arm/Group Information | Denominators | Classes | Analyses | |||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Primary | Accuracy of Sensor by Comparing SpO2 Percent Blood Oxygen Saturation to SaO2 Percent Blood Oxygen Saturation Utilizing Arms Calculation Calculation [Time Frame: 1 - 5 Hours] | The outcome of this study is the accuracy of the SpO2 (percent oxygen saturation by pulse oximetry) measured by the Reprocessed Oximeter pulse oximetry sensors during motion conditions over the range of 70 - 100% saturation compared to SaO2 (percent oxygen saturation by co-oximetry from blood samples) from arterial blood samples. The accuracy is calculated using the Arms error between measured SpO2 and reference SaO2. The Arms must meet the 3% specification for each reprocessed pulse oximetry sensor style. Standard deviation of the differences is computed as the precision. Square root of the sum of the squares of bias and precision is computed as the Arms Error value. | Posted | Number | Root Mean Square Error (%) | 1 - 5 hours |
|
2 Days
Not provided
Not provided
| ID | Title | Description | Deaths (Affected) | Deaths (At Risk) | Serious Events (Affected) | Serious Events (At Risk) | Other Events (Affected) | Other Events (At Risk) |
|---|---|---|---|---|---|---|---|---|
| EG000 | Healthy Adult Participants | Experimental: Healthy adult participants All participants are enrolled in the test group and receive the noninvasive adhesive reprocessed pulse oximeter sensors Experimental Masimo RD SET SpO2 Adhesive Sensors: Masimo RD SET SpO2 Adhesive Sensors, Adult (4000 Adt), Adult & Neonate (4003 Neo) |
Not provided
Not provided
Not provided
| Title | Organization | Phone | Extension | |
|---|---|---|---|---|
| Senior Principal Regulatory Affairs | Stryker Sustainability Solutions | 5602-410-0204 | moira.barton@stryker.com |
Not provided
| Type | Includes Protocol | Includes SAP | Includes ICF | Document Label | Document Date | Document Uploaded Date | Document File Name |
|---|---|---|---|---|---|---|---|
| Prot_SAP_ICF | Yes | Yes | Yes | Study Protocol, Statistical Analysis Plan, and Informed Consent Form | May 25, 2023 | May 27, 2024 | Prot_SAP_ICF_000.pdf |
Not provided
Not provided
Not provided
| ID | Term |
|---|---|
| D000860 | Hypoxia |
| ID | Term |
|---|---|
| D012818 | Signs and Symptoms, Respiratory |
| D012816 | Signs and Symptoms |
| D013568 | Pathological Conditions, Signs and Symptoms |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Participants |
|
| Sex: Female, Male | Count of Participants | Participants |
|
| Race/Ethnicity, Customized | Count of Participants | Participants |
|
| Region of Enrollment | Count of Participants | Participants |
|
Experimental: Healthy adult participants All participants are enrolled in the test group and receive the noninvasive adhesive reprocessed pulse oximeter sensors
|
|
| 0 |
| 11 |
| 0 |
| 11 |
| 0 |
| 11 |
Not provided
Not provided