A Prospective, Multicenter, Open-label, Self-controlled Clinical Trial Protocol to Evaluate the Efficacy and Safety of a Multi-parameter Non-Invasive Glucose Sensors for Measuring Blood Glucose.
A Prospective, Multicenter, Open-label, Self-controlled Clinical Trial Protocol to Evaluate the Efficacy and Safety of a Multi-parameter Non-Invasive Glucose Sensors for Measuring Blood Glucose.
This was a prospective, multicenter, open-label, self-controlled study to evaluate the efficacy and safety of multi-parameter non-invasive glucose meters for blood glucose measurement,planned enrollment of 200 patients with type 1 or type 2 diabetes.
Comparing glucose measurements from the Non-Invasive Glucose monitoring system to venous blood glucose or fingertip blood glucose to evaluate System accuracy capability.
At the same time, blood glucose was measured by non-invasive blood glucose meter, Accu-Chek Guide blood glucose meter and EKF Biosen C-Line. Eight groups of blood glucose values were collected from each subject during fasting, after breakfast and after lunch.After the end of blood glucose measurement, the data were recorded, including adverse events and device defects during the test. The subjects were required to stay in the examination room for 5-10 minutes, during which they completed the product use evaluation questionnaire.
Inclusion Criteria:
Exclusion Criteria: