Evaluate the Safety and Efficacy of Sinocare CGM System Regarding Real Time Glucose
Evaluate the Safety and Efficacy of Sinocare CGM System Regarding Real Time Glucose
From Jan 2022 to June 2022, 120 subjects (22-75 years of age, mean ±standard deviation [SD] of 53.20 ±14.49 years) with type 1 or type 2 diabetes participated in the study in P.R. China. Subjects wore two sensors in the abdomen which were paired with smart devices running a glucose monitoring mobile application (i3/H3 CGM APP). Subjects were asked to undergo one in-clinic visit (on one of the study days 2, 7-9, or 15) for 7-hour long YSI frequent sampling test (FST) periods, where intravenous (IV) blood samples were drawn every 15 min and analyzed using the YSI 2300 STAT Plus Glucose & Lactate Analyzer (Yellow Springs Instrument, YSI Life Sciences, Yellow Springs, OH).
Inclusion Criteria:
Exclusion Criteria: