Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Class |
|---|---|
| University Hospital, Geneva | OTHER |
| B.P. Koirala Institute of Health Sciences | OTHER |
Not provided
Not provided
Not provided
Not provided
With the rise of cardiovascular diseases (CVD) and diabetes, the global disease burden is shifting towards non-communicable diseases (NCDs). An increasing number of low- and middle-income countries (LMICs) are currently experiencing the double burden of infectious and non-communicable diseases. In order to facilitate a patient-centred approach to healthcare, there is an urgent need to ensure that primary healthcare (PHC) facilities in LMICs are capable of addressing diagnosis and monitoring of non-communicable diseases at the point-of-care (POC). Important minimum parameters for PHC POC diagnosis and monitoring of cardiometabolic diseases are lipids/lipoproteins, glucose, glycated haemoglobin (HbA1c) and serum creatinine, to address cardiovascular disease, diabetes and chronic kidney disease.
While several technologies of multi-parameter POC devices capable of supporting diagnosis and monitoring of cardiometabolic diseases exist, their quantitative accuracy is often not well evaluated outside of the manufacturer's laboratories and published independent evaluations can be rare, particularly in the settings of intended use. These settings are PHC facilities in varying climatic environments and with staff without specialist laboratory training. Our study aims to evaluate the quantitative accuracy of 2 cardiometabolic POC devices in a setting of intended use and performed by the intended user. (Evaluating the quantitative measurements of glucose, HbA1c, total cholesterol and creatinine as measured in a healthcare setting with point-of-care multiparameter devices compared to a laboratory reference method).
Independent performance evaluation of POC devices have shown that many devices, regardless of their intended use, do not perform as well as the manufacturers' claims may suggest. There are a variety of reasons for this, including difference in skills levels of personnel who were involved in generating the data for the manufacturer's claims and who actually perform the test in a clinic, limited evaluation of samples from different geographical locations and thus potential interfering substances or less controlled environmental conditions (e.g. dust, heat and humidity).
Independent performance evaluations to assess clinical accuracy, as well as system usability are important to drive adoption of any technology, and even more so, if the technology is intended to move testing outside of the traditional setting, i.e. away from the central laboratory to the point-of-care. Many devices appear ideally suited for certain settings, however when it comes to actual implementation, the users often discover that the device and workflow do not meet their needs. The choice of a suboptimal device means badly invested resources and may lead to inappropriate use, resulting in fewer reliable tests for patients in the absence of alternatives. Data from real-world evaluations in settings of the intended use can support decision makers to select the right device.
The setting for this study will be a primary healthcare facility and a health post in Nepal, where study participants attend the facility in the context of the "Early detection and management of Chronic Kidney Diseases, Hypertension, Diabetes and Cardiovascular disease in Community in Nepal (KHDC-Nepal) program", conducted by the University Hospital of Geneva and the BP Koirala Institute of Health Science in Nepal.
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Diagnostic | Other | Fingerstick collection and venous whole blood collection. |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| Determine the accuracy of quantitative measurements of GLU, HbA1c,CHOL and CRE with POC devices compared to a laboratory reference method (Roche) | Device | Fingerstick capillary blood will be collected according to each manufacturer's Instructions for use. The fingerstick capillary blood samples will be used immediately on the point-of-care assays. The venous blood will be collected in specific tubes for plasma (GLU), for whole blood (HbA1c) and for serum (CREP, CHOL). Plasma will be separated by centrifugation at the PHC facility and health post and aliquots will be prepared. Serum will be left to clot prior to centrifugation and serum aliquots will be prepared at the PHC facility and health post. The samples will be stored until transfer to the BPKIHS reference lab. |
| Measure | Description | Time Frame |
|---|---|---|
| Correlation estimation of quantitative measurements of cardiometabolic blood parameters measured on a point-of-care device versus a laboratory reference method | Estimates of correlation between quantitative measurements of blood glucose, HbA1c, total cholesterol and creatinine collected from capillary blood and tested with point-of-care multiparameter devices versus quantitative measurements of blood glucose, HbA1c, total cholesterol and creatinine collected from venous blood and tested with a laboratory reference method. | 4 months |
| Limits of agreements estimation of quantitative measurements of cardiometabolic blood parameters measured on a point-of-care device versus a laboratory reference method | Estimates of limits of agreement for quantitative measurements of blood glucose, HbA1c, total cholesterol and creatinine collected from capillary blood and tested with point-of-care multiparameter devices versus quantitative measurements of blood glucose, HbA1c, total cholesterol and creatinine collected from venous blood and tested with a laboratory reference method. | 4 months |
| Measure | Description | Time Frame |
|---|---|---|
| Operational characteristics of study device Tascom | Rate of invalid test results and error types for Tascom (POC device) | 4 months |
| Operational characteristics of study device JanaCare | Rate of invalid test results and error types for JanaCare(POC device) |
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Sanjib Sharma, Prof.Dr | head of the B.P Koirala Institute of Health Sciences | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| B.P Koirala Institute of Health Sciences | Dharān | Sunsari | 56700 | Nepal |
| PubMed Identifier | Type | Citation | Retractions |
|---|---|---|---|
| 39361597 | Derived | Giachino M, Vetter B, Perone SA, Correia JC, Erkosar B, Heller O, Khanal VK, Lab B, Pataky Z, Poudel S, Rai M, Sharma SK. Performance and usability of cardiometabolic point of care devices in Nepal: A prospective, quantitative, accuracy study. PLOS Glob Public Health. 2024 Oct 3;4(10):e0003760. doi: 10.1371/journal.pgph.0003760. eCollection 2024. |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Type | Includes Protocol | Includes SAP | Includes ICF | Document Label | Document Date | Document Uploaded Date | Document File Name |
|---|---|---|---|---|---|---|---|
| Prot | Yes | No | No | Study Protocol | Mar 30, 2022 | Mar 18, 2024 | Prot_000.pdf |
Not provided
| ID | Term |
|---|---|
| D003920 | Diabetes Mellitus |
| D051436 | Renal Insufficiency, Chronic |
| D006949 | Hyperlipidemias |
| ID | Term |
|---|---|
| D044882 | Glucose Metabolism Disorders |
| D008659 | Metabolic Diseases |
| D009750 | Nutritional and Metabolic Diseases |
| D004700 | Endocrine System Diseases |
Not provided
Not provided
Prospective quantitative accuracy study
Not provided
Not provided
Not provided
Not provided
|
| 4 months |
| Usability of study devices Tascom | System usability score for Tascom (POC device) SUS Score (>80.3, Grade A, Excellent) SUS Score (68-80.3, Grade B, Good) SUS Score (68, Grade C, Okay) SUS Score (51-68, Grade D, Poor) SUS Score (<51, Grade F, Awful) | 4 months |
| Usability of study devices JanaCare | System usability score for JanaCare(POC device) SUS Score (>80.3, Grade A, Excellent) SUS Score (68-80.3, Grade B, Good) SUS Score (68, Grade C, Okay) SUS Score (51-68, Grade D, Poor) SUS Score (<51, Grade F, Awful) | 4 months |
| D051437 | Renal Insufficiency |
| D007674 | Kidney Diseases |
| D014570 | Urologic Diseases |
| D052776 | Female Urogenital Diseases |
| D005261 | Female Urogenital Diseases and Pregnancy Complications |
| D000091642 | Urogenital Diseases |
| D052801 | Male Urogenital Diseases |
| D002908 | Chronic Disease |
| D020969 | Disease Attributes |
| D010335 | Pathologic Processes |
| D013568 | Pathological Conditions, Signs and Symptoms |
| D050171 | Dyslipidemias |
| D052439 | Lipid Metabolism Disorders |