To compare the clinical performance of AuraGain including insertion time and success rate at first attempt, fiberoptic bronchoscopic view, oropharyngeal leak pressure, the degree of ease of gastric tube insertion, and peri-operative adverse effects in young children in supine, moderate lateral and extreme lateral position.
Ambu® AuraGain, a relatively novel supraglottic airway device, has been introduced recently. AuraGain has an inflatable cuff and a curved body. In addition, because the airway tube of AuraGain is wide, it can be used as a conduit for tracheal intubation. However, only a limited number of studies have been conducted on the performance of AuraGain in pediatric patients.
Changes in the position of the head and neck alter the shape of the pharynx leading to changes in the oropharyngeal leak pressure and the quality of ventilation through the LMA in both the adult and pediatric population. Having a large occiput and a relatively cephalically placed glottis, the ventilation in children is more compromised due to poor alignment of the pharyngeal-laryngeal axes during flexion of the head and neck. The hanging epiglottis covers the laryngeal inlet during flexion and the cuff of the LMA compresses the narrow laryngeal inlet both decreasing the delivered tidal volume. Different changes occur during extension and lateral rotation
Although the clinical performance of the AuraGain would be expected to be high, or at least non-inferior, to that of the widely used currently available SGAs, clinical trials are required to verify its performance in head and neck positions other than the neutral position (i.e flexion, extension and lateral rotation).
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Moderate lateral Group | Patients will be anesthetized using ® AuraGain airway that will be inserted lubricated with partially deflated cuff. After insertion, the cuff will be inflated with the recommended volume of air and operation will undergo while patient's neck in moderate lateral position. |
| |
| Extreme lateral Group | Patients will be anesthetized using ® AuraGain airway that will be inserted lubricated with partially deflated cuff. After insertion, the cuff will be inflated with the recommended volume of air and operation will undergo while patient's neck in extreme lateral position. |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| AuraGain | Device | Ambu® AuraGain, a relatively novel supraglottic airway device, has been introduced recently. AuraGain has an inflatable cuff and a curved body. |
|
| Measure | Description | Time Frame |
|---|---|---|
| Oropharyngeal leak pressure | Airway leak pressure will be determined by closing the adjustable expiratory pressure-limiting (APL) valve and setting the fresh gas flow rate to 3 L/min. The airway pressure at which an audible leak in the mouth is heard is recorded as the "leak Pressure". | Intraoperative from 1 minute after insertion of the device as a baseline to 60 minutes after insertion of the device |
| Measure | Description | Time Frame |
|---|---|---|
| Ventilation score (grade 0, 1, 2 or 3), | the ventilation score which is scored from 0 to 3 based on three criteria: no leakage with an airway pressure of 15 cm H2O, bilateral chest excursions with a peak inspiratory pressure of 20 cm of H2O, and a square wave capnogram, with each item scoring 0 or 1 point . Thus, if all three criteria are satisfied, the ventilation score is 3. | Intraoperative from 1 minute after insertion of the device as baseline to 60 minutes after insertion of the device |
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
- children scheduled for elective outpatient surgery in which airway management with a® AuraGain would be appropriate.
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Hala S Abdel-Ghaffar, MD | Assiut University | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| Hala Abdel-Ghaffar | Asyut | Asyut Governorate | 71515 | Egypt |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Fiberoptic glottic view Score | The fiberoptic images will be graded with a score from 1 to 5, which has been defined and proposed previously (20, 21); (grade 1-only larynx seen; grade 2-larynx and epiglottis posterior surface seen; grade 3-larynx and epiglottis tip of anterior surface seen, <50%visual obstruction of epiglottis to larynx; grade 4-epiglottis down folded and its anterior surface seen, >50%visual obstruction of epiglottis to larynx; and grade 5-epiglottis down folded and larynx cannot be seen directly). The images will be taken in the neutral position and in mid and extreme laterally rotated neck positions. | Intraoperative from 1 minute after insertion of the device to 60 minutes after insertion of the device. |