The purpose of this study is to compare the accuracy of capsulotomy assisted by intraocular caliper and Verion navigation system in phacoemulsification surgery and postoperative visual quality of patients.
Continuous curvilinear capsulorhexis (CCC) is an important and technically challenging procedure in cataract surgery. However, most surgeons perform CCC by experience or referring to the pupil size, which is highly subjective and the accuracy of capsulorhexis is unsatisfactory. The investigators developed a new modified intraocular caliper with standard calibration on the rinse needle, which can measure and locate the position of capsulorhexis. The surgeon can gently use the blunt needle in the front of the intraocular caliper to further make corresponding markers on the lens anterior capsule, and then carry out capsulotomy according to the marks. We found intraocular caliper-assisted capsulotomy can obviously improve the accuracy of capsulorhexis. The purpose of this study is to compare the accuracy and postoperative visual quality of intraocular caliper-assisted capsulotomy with those of digital image projection device VERION navigation system.
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Intraocular caliper-assisted capsulotomy group | Experimental | In intraocular caliper-assisted capsulotomy group, a modified intraocular caliper with standard calibration on the rinse needle was used to measure and locate the position of capsulorhexis with 5.3 mm diameter and the principle of pupillary margin in concentric circles. The surgeon could gently use the blunt needle in the front of the intraocular caliper to further make corresponding markers on the lens anterior capsule, and then carry out capsulotomy according to the marks. |
|
| Verion navigation system-assisted capsulotomy group | Active Comparator | In the Verion-assisted capsulotomy group, Verion navigation system was applied to project a 5.3 mm capsulorhexis centered on the corneal vertex. According to the projected capsulorhexis, capsulotomy was performed. |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| Intraocular caliper-assisted capsulotomy | Procedure | Intraocular caliper is a modified intraocular caliper with standard calibration on the rinse needle. |
|
| Measure | Description | Time Frame |
|---|---|---|
| Deviation of horizontal and vertical diameter of capsulorhexis from the target diameter (5.3 mm) | Images of the edges of IOL and capsulorhexis are taken at the end of the operation. Image J image analysis software is used to measure the horizontal diameter and vertical diameter of capsulorhexis with the diameter of IOL optical surface (6 mm) as the reference by two independently professional graders at the meantime, and further calculate the deviation of horizontal and vertical diameter of capsulorhexis from the target diameter (5.3 mm) . | During the surgery |
| Measure | Description | Time Frame |
|---|---|---|
| Uncorrected and corrected visual acuity | Visual acuity is evaluated with ETDRS visual acuity chart. | One month after surgery |
| Ideal ratio of capsulorhexis | Ideal capsulorhexis is defined as a centered and round capsulorhexis, which can cover the edge of IOL optical surface continuously for 360 degrees with a diameter between 5.0 to 5.5 mm. |
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Not provided
| Name | Role | Phone | Extension | |
|---|---|---|---|---|
| Xiaoyun Chen, PhD | Contact | +86 13632343642 | chenxiaoyun@gzzoc.com |
Not provided
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| Zhongshan Ophthalmic Center,Sun Yat-Sen University | Guangzhou | 510060 | China |
| PubMed Identifier | Type | Citation | Retractions |
|---|---|---|---|
| 41231447 | Derived | Chen X, Liu Z, Jin L, Zheng Y, Luo L, Liu Y. Manual Marking Guidance vs Digital Guidance System-Assisted Capsulorhexis in Phacoemulsification: A Noninferiority Randomized Clinical Trial. JAMA Ophthalmol. 2026 Jan 1;144(1):15-22. doi: 10.1001/jamaophthalmol.2025.4448. |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Verion navigation system-assisted capsulotomy | Procedure | Verion Image Guided System (Alcon Laboratories, Inc.) is a digital image guidance system consisting of a diagnostic reference unit and a surgery pilot for intraoperative surgical assistance. The device can be used to align toric intraocular lens during surgery as well as support manual creation of a capsulorhexis with a predefined target diameter. |
|
| During the surgery |
| Grades of capsulorhexis-IOL overlap | The capsulorhexis-IOL overlap of each patient is graded as 360-degree continuous overlap, 270~360-degree overlap, 180~270-degree overlap, 90~180-degree overlap, and < 90-degree overlap. | During the surgery |
| Distance between pupil center and capsulorhexis center | Measured by Image J image analysis software. | During the surgery |
| Horizontal and vertical diameter of capsulorhexis | Measured by Image J image analysis software. | During the surgery |
| Intraocular lens tilt | Measured by anterior segment OCT (CASIA2). | One month after surgery |
| Intraocular lens decentration | Measured by anterior segment OCT (CASIA2). | One month after surgery |
| Intraocular aberrations | Measured by OPD-SCAN III. | One month after surgery |
| Deviation of horizontal and vertical diameter of capsulorhexis from the target diameter (5.3 mm) | One month after surgery |