Undetected bowel perforation is a rare but dangerous complication of laparoscopic surgery. If the injury is not detected and treated at the time of the surgical procedure, the patient can suffer severe complications, including septic shock and eventually death. The investigator's goal is to test a novel device that can detect bowel gas leakage from perforation and alert the surgeon during the operation by evaluating the gases present in the insufflated abdomen during surgery. This study will determine the ability of the device to be attached to a standard trocar during the operation and periodically draw small samples or aliquots of gas from the abdomen to evaluate the gas and accurately detect gaseous content from the bowel. Before the device can be used to detect bowel perforations, the investigators must first ensure that it can accurately detect bowel gas in an insufflated abdomen.
The goal of this study is to test a novel device that can detect bowel gas leakage from perforation and alert the surgeon during the operation by evaluating the gases present in the insufflated abdomen during surgery. During laparoscopic surgery, CO2 is inserted into the abdominal cavity in order to perform the operation. This is a dynamic process as insufflation is constant during the entire procedure to maintain constant pressure and compensate for any small leaks due to the insertion and retrieval of instruments.
This study will determine the ability of a novel device to be attached to a standard trocar during the operation and periodically draw a small amount of gas from the abdomen to evaluate the gas and accurately detect gaseous content from the bowel. Before the device can be used to detect bowel perforations, first the investigators must ensure that it can accurately detect bowel gas in an insufflated abdomen.
The Investigational device, Perf-AlertTM, consists of a 15" x 13" x 7" unit containing gas sensors and valves, a small box containing a one way, ultra-low pressure pump, user controls (buttons), and software algorithms, which control the operation of the device ( Image1). It is used in conjunction with a single-use, disposable kit consisting of sterile tubing and filters used for sample collection and transport (Image 2). The sensing unit will be connected via a sterile tube/filter set to any trocar port in use during the procedure. At specific time points during the procedure, a one-way valve will be opened and the unit's pump turned on pulling small aliquots of gas from the abdominal cavity to the sensing unit. The system's architecture is configured such that sample collection, transport, analysis, and feedback occur in a single step such that sample collection and sensor feedback occur in real-time. A standard laptop running an analytical software program is connected to the unit to record and log sensor readings. No changes to the device are anticipated during the course of the study.
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Cohort 1: Predetermined points measurement | Other | Cohort 1: Predetermined points measurement: 10 subjects with measurement of bowel gas at 8 predetermined time points during right laparoscopic colectomy as follow:
|
|
| Cohort 2: Continuous monitoring | Other | Cohort 2: Continuous monitoring: 10 subjects with continuous monitoring of bowel gases through the surgery. The level of H2 and CH4 gases will be noted at the 8 predetermined time points during the continuous monitoring as well. |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| Obtaining bowel gas samples | Device | The Perf-AlertTM prototype consists of a 15" x 13" x 7" unit containing gas sensors and valves, a small box containing a one-way, ultra-low pressure pump, user controls (buttons), and software algorithms, which control the operation of the device. It is used in conjunction with a single-use, disposable kit of sterile tubing and filters used for sample collection. The sensing unit will be connected via a sterile tube/filter set to any trocar port in use during the procedure. At specific time points during surgery, a one-way valve will be opened, and the unit's pump turned on and withdrawing a small amount of gas from the abdominal cavity. The system is configured that sample collection, transport, analysis, and feedback occur in a single step such that sample collection and sensor feedback occur in real-time. A standard laptop running an analytical software program is connected to the unit to record and log sensor readings. |
| Measure | Description | Time Frame |
|---|---|---|
| Methane (CH4) Gas | Methane (CH4) gas measured in ppm or mg/L (parts per million or milligrams per liter (mg/L) at 8-time points | during surgery(from start of laparoscopy to the end of surgical procedure) |
| Hydrogen (H2) Gas | Hydrogen (H2) gas measured in ppm or mg/L (parts per million or milligrams per liter (mg/L) at 8-time points | during surgery(from start of laparoscopy to the end of surgical procedure) |
| Measure | Description | Time Frame |
|---|---|---|
| Response Time of the Perf-AlertTM Device. | The response time of the Perf-AlertTM device measured in seconds from obtaining the sample (from opening the valve to obtain sample) to first detecting a gas presence at the device electronic sensors | during surgery(from start of laparoscopy to the end of surgical procedure) |
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Mehraneh Jafari, MD | Weill Medical College of Cornell University | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| Weill Cornell Medicine | New York | New York | 10065 | United States |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| FG000 | Cohort 1: Predetermined Points Measurement | Cohort 1: Predetermined points measurement: 10 subjects with measurement of bowel gas at 8 predetermined time points during right laparoscopic colectomy as follow:
Obtaining bowel gas samples: The Perf-AlertTM prototype consists of a 15" x 13" x 7" unit containing gas sensors and valves, a small box containing a one-way, ultra-low pressure pump, user controls (buttons), and software algorithms, which control the operation of the device. It is used in conjunction with a single-use, disposable kit of sterile tubing and filters used for sample collection. The sensing unit will be connected via a sterile tube/filter set to any trocar port in use during the procedure. At specific time points during surgery, a one-way valve will be opened, and the unit's pump turned on and withdrawing a small amount of gas from the abdominal cavity. The system is configured that sample collection, transport, analysis, and feedback occur in a single step such that sample collection and sensor feedback occur in real-time. A standard laptop running an analytical software program is connected to the unit to record and log sensor readings. |
| Title | Milestones | Reasons Not Completed | |||||
|---|---|---|---|---|---|---|---|
| Overall Study |
|
Not provided
| Type | Includes Protocol | Includes SAP | Includes ICF | Document Label | Document Date | Document Uploaded Date | Document File Name |
|---|---|---|---|---|---|---|---|
| Prot_SAP | Yes | Yes | No | Study Protocol and Statistical Analysis Plan | Jun 3, 2024 |
Not provided
Prospective pilot, single-center, single-arm, open-label, phase 1 investigator-initiated study to determine device ability to detect bowel gas during laparoscopic right colectomy with intracorporeal anastomosis using Perf-AlertTM gas detection system on 20 subjects into 2 Cohorts.
Cohort 1: 10 subjects with measurement of bowel gas at 8 predetermined time points during right laparoscopic colectomy as follow:
Cohort 2: 10 subjects with continuous monitoring of bowel gases through the surgery. The level of H2 and CH4 gases will be noted at the 8 predetermined time points during the continuous monitoring as well.
Not provided
Not provided
Not provided
Not provided
|
| FG001 | Cohort 2: Continuous Monitoring | Cohort 2: Continuous monitoring: 10 subjects with continuous monitoring of bowel gases through the surgery. The level of H2 and CH4 gases will be noted at the 8 predetermined time points during the continuous monitoring as well. Obtaining bowel gas samples: The Perf-AlertTM prototype consists of a 15" x 13" x 7" unit containing gas sensors and valves, a small box containing a one-way, ultra-low pressure pump, user controls (buttons), and software algorithms, which control the operation of the device. It is used in conjunction with a single-use, disposable kit of sterile tubing and filters used for sample collection. The sensing unit will be connected via a sterile tube/filter set to any trocar port in use during the procedure. At specific time points during surgery, a one-way valve will be opened, and the unit's pump turned on and withdrawing a small amount of gas from the abdominal cavity. The system is configured that sample collection, transport, analysis, and feedback occur in a single step such that sample collection and sensor feedback occur in real-time. A standard laptop running an analytical software program is connected to the unit to record and log sensor readings. |
| COMPLETED |
|
| NOT COMPLETED |
|
Study enrollment closed prior to cohort 2 enrollment. Zero participants were enrolled in Cohort 2
Not provided
| ID | Title | Description |
|---|---|---|
| BG000 | Cohort 1: Predetermined Points Measurement | Cohort 1: Predetermined points measurement: 10 subjects with measurement of bowel gas at 8 predetermined time points during right laparoscopic colectomy as follow:
Obtaining bowel gas samples: The Perf-AlertTM prototype consists of a 15" x 13" x 7" unit containing gas sensors and valves, a small box containing a one-way, ultra-low pressure pump, user controls (buttons), and software algorithms, which control the operation of the device. It is used in conjunction with a single-use, disposable kit of sterile tubing and filters used for sample collection. The sensing unit will be connected via a sterile tube/filter set to any trocar port in use during the procedure. At specific time points during surgery, a one-way valve will be opened, and the unit's pump turned on and withdrawing a small amount of gas from the abdominal cavity. The system is configured that sample collection, transport, analysis, and feedback occur in a single step such that sample collection and sensor feedback occur in real-time. A standard laptop running an analytical software program is connected to the unit to record and log sensor readings. |
| BG001 | Cohort 2: Continuous Monitoring | Cohort 2: Continuous monitoring: 10 subjects with continuous monitoring of bowel gases through the surgery. The level of H2 and CH4 gases will be noted at the 8 predetermined time points during the continuous monitoring as well. Obtaining bowel gas samples: The Perf-AlertTM prototype consists of a 15" x 13" x 7" unit containing gas sensors and valves, a small box containing a one-way, ultra-low pressure pump, user controls (buttons), and software algorithms, which control the operation of the device. It is used in conjunction with a single-use, disposable kit of sterile tubing and filters used for sample collection. The sensing unit will be connected via a sterile tube/filter set to any trocar port in use during the procedure. At specific time points during surgery, a one-way valve will be opened, and the unit's pump turned on and withdrawing a small amount of gas from the abdominal cavity. The system is configured that sample collection, transport, analysis, and feedback occur in a single step such that sample collection and sensor feedback occur in real-time. A standard laptop running an analytical software program is connected to the unit to record and log sensor readings. |
| BG002 | Total | Total of all reporting groups |
| Units | Counts |
|---|---|
| Participants |
|
| Title | Description | Population Description | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Denominator Units Selected | Denominators | Classes | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Age, Categorical | Count of Participants | Participants |
| ||||||||||||||||||
| Age, Continuous | Mean | Standard Deviation | years |
| |||||||||||||||||
| Sex: Female, Male | Count of Participants | Participants |
| ||||||||||||||||||
| Ethnicity (NIH/OMB) | Count of Participants | Participants |
| ||||||||||||||||||
| Race (NIH/OMB) | Count of Participants | Participants |
| ||||||||||||||||||
| Region of Enrollment | Number | participants |
|
| Type | Title | Description | Population Description | Reporting Status | Anticipated Posting Date | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Time Frame | Units Analyzed | Denominator Units Selected | Arm/Group Information | Denominators | Classes | Analyses | |||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Primary | Methane (CH4) Gas | Methane (CH4) gas measured in ppm or mg/L (parts per million or milligrams per liter (mg/L) at 8-time points | Zero participants were enrolled in Cohort 2. | Posted | Mean | Standard Deviation | ppm | during surgery(from start of laparoscopy to the end of surgical procedure) |
|
|
| ||||||||||||||||||||||||||||
| Primary | Hydrogen (H2) Gas | Hydrogen (H2) gas measured in ppm or mg/L (parts per million or milligrams per liter (mg/L) at 8-time points | Zero participants were enrolled in Cohort 2. | Posted | Mean | Standard Deviation | ppm | during surgery(from start of laparoscopy to the end of surgical procedure) |
| ||||||||||||||||||||||||||||||
| Secondary | Response Time of the Perf-AlertTM Device. | The response time of the Perf-AlertTM device measured in seconds from obtaining the sample (from opening the valve to obtain sample) to first detecting a gas presence at the device electronic sensors | Zero participants were enrolled in Cohort 2. | Posted | Mean | Standard Deviation | seconds | during surgery(from start of laparoscopy to the end of surgical procedure) |
|
90 days
Zero participants were enrolled in Cohort 2.
Not provided
| ID | Title | Description | Deaths (Affected) | Deaths (At Risk) | Serious Events (Affected) | Serious Events (At Risk) | Other Events (Affected) | Other Events (At Risk) |
|---|---|---|---|---|---|---|---|---|
| EG000 | Cohort 1: Predetermined Points Measurement | Cohort 1: Predetermined points measurement: 10 subjects with measurement of bowel gas at 8 predetermined time points during right laparoscopic colectomy as follow:
Obtaining bowel gas samples: The Perf-AlertTM prototype consists of a 15" x 13" x 7" unit containing gas sensors and valves, a small box containing a one-way, ultra-low pressure pump, user controls (buttons), and software algorithms, which control the operation of the device. It is used in conjunction with a single-use, disposable kit of sterile tubing and filters used for sample collection. The sensing unit will be connected via a sterile tube/filter set to any trocar port in use during the procedure. At specific time points during surgery, a one-way valve will be opened, and the unit's pump turned on and withdrawing a small amount of gas from the abdominal cavity. The system is configured that sample collection, transport, analysis, and feedback occur in a single step such that sample collection and sensor feedback occur in real-time. A standard laptop running an analytical software program is connected to the unit to record and log sensor readings. | 0 | 3 | 0 | 3 | 0 | 3 |
| EG001 | Cohort 2: Continuous Monitoring | Cohort 2: Continuous monitoring: 10 subjects with continuous monitoring of bowel gases through the surgery. The level of H2 and CH4 gases will be noted at the 8 predetermined time points during the continuous monitoring as well. Obtaining bowel gas samples: The Perf-AlertTM prototype consists of a 15" x 13" x 7" unit containing gas sensors and valves, a small box containing a one-way, ultra-low pressure pump, user controls (buttons), and software algorithms, which control the operation of the device. It is used in conjunction with a single-use, disposable kit of sterile tubing and filters used for sample collection. The sensing unit will be connected via a sterile tube/filter set to any trocar port in use during the procedure. At specific time points during surgery, a one-way valve will be opened, and the unit's pump turned on and withdrawing a small amount of gas from the abdominal cavity. The system is configured that sample collection, transport, analysis, and feedback occur in a single step such that sample collection and sensor feedback occur in real-time. A standard laptop running an analytical software program is connected to the unit to record and log sensor readings. | 0 | 0 | 0 | 0 | 0 | 0 |
Not provided
Not provided
This study was terminated leading to small numbers of subjects enrolled and analyzed.
Not provided
Not provided
Not provided
| Title | Organization | Phone | Extension | |
|---|---|---|---|---|
| Dr. Mehraneh Jafari | Weill Cornell Medical Center | +1 646 962 2270 | mdj9003@med.cornell.edu |
| May 14, 2025 |
| Prot_SAP_000.pdf |
| ID | Term |
|---|---|
| D003424 | Crohn Disease |
| D003110 | Colonic Neoplasms |
| ID | Term |
|---|---|
| D015212 | Inflammatory Bowel Diseases |
| D005759 | Gastroenteritis |
| D005767 | Gastrointestinal Diseases |
| D004066 | Digestive System Diseases |
| D007410 | Intestinal Diseases |
| D015179 | Colorectal Neoplasms |
| D007414 | Intestinal Neoplasms |
| D005770 | Gastrointestinal Neoplasms |
| D004067 | Digestive System Neoplasms |
| D009371 | Neoplasms by Site |
| D009369 | Neoplasms |
| D003108 | Colonic Diseases |
Not provided
Not provided
| >=65 years |
|
| Unknown or Not Reported |
|
| Native Hawaiian or Other Pacific Islander |
|
| Black or African American |
|
| White |
|
| More than one race |
|
| Unknown or Not Reported |
|
| OG001 | Cohort 2: Continuous Monitoring | Cohort 2: Continuous monitoring: 10 subjects with continuous monitoring of bowel gases through the surgery. The level of H2 and CH4 gases will be noted at the 8 predetermined time points during the continuous monitoring as well. Obtaining bowel gas samples: The Perf-AlertTM prototype consists of a 15" x 13" x 7" unit containing gas sensors and valves, a small box containing a one-way, ultra-low pressure pump, user controls (buttons), and software algorithms, which control the operation of the device. It is used in conjunction with a single-use, disposable kit of sterile tubing and filters used for sample collection. The sensing unit will be connected via a sterile tube/filter set to any trocar port in use during the procedure. At specific time points during surgery, a one-way valve will be opened, and the unit's pump turned on and withdrawing a small amount of gas from the abdominal cavity. The system is configured that sample collection, transport, analysis, and feedback occur in a single step such that sample collection and sensor feedback occur in real-time. A standard laptop running an analytical software program is connected to the unit to record and log sensor readings. |
|
|
| OG001 | Cohort 2: Continuous Monitoring | Cohort 2: Continuous monitoring: 10 subjects with continuous monitoring of bowel gases through the surgery. The level of H2 and CH4 gases will be noted at the 8 predetermined time points during the continuous monitoring as well. Obtaining bowel gas samples: The Perf-AlertTM prototype consists of a 15" x 13" x 7" unit containing gas sensors and valves, a small box containing a one-way, ultra-low pressure pump, user controls (buttons), and software algorithms, which control the operation of the device. It is used in conjunction with a single-use, disposable kit of sterile tubing and filters used for sample collection. The sensing unit will be connected via a sterile tube/filter set to any trocar port in use during the procedure. At specific time points during surgery, a one-way valve will be opened, and the unit's pump turned on and withdrawing a small amount of gas from the abdominal cavity. The system is configured that sample collection, transport, analysis, and feedback occur in a single step such that sample collection and sensor feedback occur in real-time. A standard laptop running an analytical software program is connected to the unit to record and log sensor readings. |
|
|