The purpose of this study is to determine whether the ControlRad system device when installed in Fluoroscopic (x-ray) machine reduces radiation dose to operators (Doctors and nurses) and to the patient who is undergoing cardiac Catheterization and electrophysiology (EP) implant procedures.
This is a Prospective, randomized (like a "flip of a coin") single center study evaluating the ControlRad system device when installed - retrofitted on existing C-arms for fluoroscopy (x-ray) will reduce radiation exposure without compromising image quality in the region of interest and overall workflow during cardiac catheterization and electrophysiology (EP) implant procedures. Enrollment will be limited to 40 cardiac catheterization, and 30 pacemaker and/or ICD procedures. All patients presenting to Beaumont Hospital Royal Oak who are scheduled for cardiac catheterization with coronary angiography and electrophysiology implant in Cath Lab 5 where device is installed will be considered for inclusion in the study. Physicians and staff (20 fellows; 35 attending physicians; and 56 Cath-lab staff,including Nurse/Anesthesia team) will be invited to participate in the study. For each procedure, 3 staff members will be study subjects: a) the primary operator, b) the secondary operator, c) the nurse/tech and or Anesthetist who circulates in the room. Different operators and staff may participate in each case depending on cath-lab scheduling. The Primary operator is the physician who usually stands nearest to the C-arm. In most cases in Beaumont this is a Cardiology Fellow in cardiac catheterization cases and Attending Physician for the EP cases. The Secondary operator is the physician who typically stands to the right of the primary operator and in most cases is the attending cardiologist or scrub assistant (nurse/technician). The Nurse or Anesthesia team typically circulates in the room and is responsible for monitoring the patient, administering medications and obtaining equipment for the procedure.
Two methods will be used to measure radiation dose during the catheterization procedure. The Real-time Dosimeter Badge (RaySafe) and the Landauer Luxel Aluminium Oxide Dosimeter Badges. Cardiac Catheterization and electrophysiology implant procedures will be performed according to standard guidelines.
Participants will be randomized 1:1 to either cardiac catheterization without ControlRad or cardiac catheterization with the ControlRad system. The same 1:1 randomization process will be applied to the EP patient population. Participants will be blinded to the randomization.
Once procedure is done,participation also ended and the badges will be removed by the Radiation Safety Officer or research coordinator.
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| ControlRad System | Experimental | Participant will undergo Cardiac catheterization or electrophysiology implant procedures with the ControlRad system installed in Cath lab room 5 |
|
| Without ControlRad System | Active Comparator | Participant will undergo the cardiac catheterization or electrophysiology implant procedures per standard of care. Meaning, without the ControlRad system installed in Cath lab room 5. |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| ControlRad System | Radiation | Cardiac Catheterization or electrophysiology implant procedures with Control Rad system device installed in the cath lab room #5. |
|
| Measure | Description | Time Frame |
|---|---|---|
| Radiation Dose to Primary Operator (RaySafe), Cardiac Catheterization | Median effective radiation dose in microsievert (μSv) to the primary operator at the thyroid. This was measured using a RaySafe dosimeter to assess the radiation dose in each case. | During each cardiac catheterization, an average of 90 minutes |
| Radiation Dose to Primary Operator (Luxel Dosimeter), Cardiac Catheterization | Mean effective radiation dose in microsievert (μSv) to the primary operator at the thyroid. This was measured using a single Luxel dosimeter over all cases to assess the cumulative dose in the study arm. | During each cardiac catheterization, an average of 90 minutes |
| Radiation Dose to Primary Operator, EP Procedure (Luxel) | Mean effective radiation dose in microsievert (μSv) to the primary operator at the end of each EP implant procedure. A single Luxel dosimeter was and a cumulative dose was recorded. | During each EP implant procedure, an average of 90 minutes |
| Radiation Dose to Primary Operator, EP Procedure (RaySafe) | Median effective radiation dose in microsievert (μSv) to the primary operator at the end of each EP implant procedure using a RaySafe dosimeter. This is measured after each case. | During each EP implant procedure, an average of 90 minutes duration |
| Measure | Description | Time Frame |
|---|---|---|
| Radiation Dose to Secondary Operator, Cardiac Catheterization (RaySafe) | Median effective radiation dose (μSv) to the secondary operator at the thyroid. This was measured using a RaySafe dosimeter to assess the radiation dose in each case. | During each cardiac catheterization, an average of 90 minutes |
| Radiation Dose to Secondary Operator (Luxel Dosimeter), Cardiac Catheterization |
Not provided
Inclusion Criteria:
- Patient
Physicians and Staff o Able to provide consent for the study
Exclusion Criteria:
- Patient
Physicians and Staff
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Simon Dixon, MD | Corewell Health East | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| Beaumont Hospital - Royal Oak | Royal Oak | Michigan | 48073 | United States |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| FG000 | ControlRad System | Participant will undergo Cardiac catheterization or electrophysiology implant procedures with the ControlRad system installed in Cath lab room 5 ControlRad System: Cardiac Catheterization or electrophysiology implant procedures with Control Rad system device installed in the cath lab room #5. |
| FG001 | Without ControlRad System | Participant will undergo the cardiac catheterization or electrophysiology implant procedures per standard of care. Meaning, without the ControlRad system installed in Cath lab room 5. Without ControlRad System: Cardiac Catheterization or electrophysiology implant procedures in cath lab room 5 without the ControlRad system device. |
| Title | Milestones | Reasons Not Completed | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Overall Study |
|
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| BG000 | ControlRad System | Participant will undergo Cardiac catheterization or electrophysiology implant procedures with the ControlRad system installed in Cath lab room 5 ControlRad System: Cardiac Catheterization or electrophysiology implant procedures with Control Rad system device installed in the cath lab room #5. |
| BG001 |
| Units | Counts |
|---|---|
| Participants |
|
| Title | Description | Population Description | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Denominator Units Selected | Denominators | Classes |
|---|---|---|---|---|---|---|---|---|---|
| Age, Continuous | Mean |
| Type | Title | Description | Population Description | Reporting Status | Anticipated Posting Date | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Time Frame | Units Analyzed | Denominator Units Selected | Arm/Group Information | Denominators | Classes | Analyses | |||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Primary | Radiation Dose to Primary Operator (RaySafe), Cardiac Catheterization | Median effective radiation dose in microsievert (μSv) to the primary operator at the thyroid. This was measured using a RaySafe dosimeter to assess the radiation dose in each case. | This outcome measured only the cardiac catheterization procedures, 42 participants in the ControlRad group and 38 in the group without ControlRad | Posted | Median | Inter-Quartile Range | microsievert | During each cardiac catheterization, an average of 90 minutes |
|
During each cardiac catheterization/EP procedure, an average of 90 minutes
Not provided
Not provided
| ID | Title | Description | Deaths (Affected) | Deaths (At Risk) | Serious Events (Affected) | Serious Events (At Risk) | Other Events (Affected) | Other Events (At Risk) |
|---|---|---|---|---|---|---|---|---|
| EG000 | ControlRad System | Participant will undergo Cardiac catheterization or electrophysiology implant procedures with the ControlRad system installed in Cath lab room 5 ControlRad System: Cardiac Catheterization or electrophysiology implant procedures with Control Rad system device installed in the cath lab room #5. |
| Term | Organ System | Source Vocabulary | Assessment Type | Notes | Statistical Information |
|---|---|---|---|---|---|
| Cardiac Perforation | Injury, poisoning and procedural complications | Systematic Assessment |
Not provided
Not provided
| Title | Organization | Phone | Extension | |
|---|---|---|---|---|
| Dr. Simon Dixon, MD | William Beaumont Hospitals | 248-898-4198 | simon.dixon@beaumont.edu |
Not provided
| Type | Includes Protocol | Includes SAP | Includes ICF | Document Label | Document Date | Document Uploaded Date | Document File Name |
|---|---|---|---|---|---|---|---|
| Prot_SAP | Yes | Yes | No | Study Protocol and Statistical Analysis Plan | Jul 14, 2021 | Oct 19, 2022 | Prot_SAP_000.pdf |
Not provided
Not provided
Not provided
Randomized 1:1 like a flip of a coin, to receive either cardiac catheterization and or electrophysiology implant procedures without ControlRad system or cardiac catheterization and or electrophysiology (EP) implant procedures with the ControlRad system.
Not provided
Not provided
Participant will be blinded to the arm assigned to; but the investigators, cath lab operators and staff will be informed of the randomized assignment.
| Without ControlRad System | Radiation | Cardiac Catheterization or electrophysiology implant procedures in cath lab room 5 without the ControlRad system device. |
|
Mean effective radiation dose in microsievert (μSv) to the secondary operator at the thyroid. This was measured using a single Luxel dosimeter over all cases to assess the cumulative dose in the study arm. |
| During each cardiac catheterization, an average of 90 minutes |
| Radiation Dose to Nurse or Anesthetist, Cardiac Catheterization (RaySafe) | Median effective radiation dose in microsievert (μSv) to the Nurse 0r Anesthetist at the thyroid. This was measured using a RaySafe dosimeter to assess the radiation dose in each case. | During each cardiac catheterization, an average of 90 minutes |
| Radiation Dose to Nurse or Anesthetist, Cardiac Catheterization (Luxel) | Mean effective radiation dose in microsievert (μSv) to the Nurse or Anesthetist at the thyroid. This was measured using a single Luxel dosimeter over all cases to assess the cumulative dose in the study arm. | During each cardiac catheterization, an average of 90 minutes |
| Radiation Dose to Patient, Cardiac Catheterization | Median radiation dose-area product (DAP) to the patient at the end of each cardiac catheterization case using data from a fluoroscopy machine. DAP is defined as the absorbed dose multiplied by the area irradiated micro-gray μGy*m^2. | During each cardiac catheterization, an average of 90 minutes |
| Radiation Dose in Cath Lab Room #5, Cardiac Catheterization | Mean radiation dose microsieverts (μSv) at the end of each cardiac catheterization case at a standard location on the wall of the catheterization laboratory. This was measured using a single Luxel dosimeter over all cases to assess the cumulative dose in the study arm. | During each cardiac catheterization, an average of 90 minutes |
| Radiation Dose to Secondary Operator, EP Procedure (RaySafe) | Mean effective radiation dose (μSv) to the secondary operator at the at the thyroid. This was measured using a RaySafe dosimeter to assess the radiation dose in each case. | During each EP implant procedure, an average of 90 minutes |
| Radiation Dose to Secondary Operator, EP Procedure (Luxel) | Mean effective radiation dose (μSv) to the secondary operator at the at the thyroid. This was measured using a single Luxel dosimeter over all cases to assess the cumulative dose in the study arm. | During each EP implant procedure, an average of 90 minutes |
| Radiation Dose to Nurse or Anesthetist, EP Procedure (RaySafe) | Median effective radiation dose in microsievert (μSv) to the primary operator at the thyroid. This was measured using a RaySafe dosimeter to assess the radiation dose in each case. | During each EP implant procedure, an average of 90 minutes |
| Radiation Dose to Nurse or Anesthetist, EP Procedure (Luxel) | Mean effective radiation dose in microsievert (μSv) to the primary operator at the thyroid. This was measured using a single Luxel dosimeter over all cases to assess the cumulative dose in the study arm. | During each EP implant procedure, an average of 90 minutes |
| Radiation Dose to Patient, EP Procedure | Mean radiation dose-area product (DAP) to the patient at the end of each EP implant procedure using data from a fluoroscopy machine. DAP is defined as the absorbed dose multiplied by the area irradiated (μGy*m^2) | During each EP implant procedure, an average of 30-45 minutes of active radiation |
| Radiation Dose in Cath Lab Room #5, EP Procedure | Mean radiation dose (μSv) at the end of each EP implant procedure at a standard location on the wall of the catheterization laboratory. This was measured using a single Luxel dosimeter over all cases to assess the cumulative dose in the study arm. | During each EP implant procedure, an average of 90 minutes |
| Without ControlRad System |
Participant will undergo the cardiac catheterization or electrophysiology implant procedures per standard of care. Meaning, without the ControlRad system installed in Cath lab room 5. Without ControlRad System: Cardiac Catheterization or electrophysiology implant procedures in cath lab room 5 without the ControlRad system device. |
| BG002 | Total | Total of all reporting groups |
| years |
|
| Sex: Female, Male | Count of Participants | Participants |
|
| Race and Ethnicity Not Collected | Race and Ethnicity were not collected from any participant. | Count of Participants | Participants |
|
| Region of Enrollment | Number | participants |
|
| OG001 | Without ControlRad System | Participant will undergo the cardiac catheterization or electrophysiology implant procedures per standard of care. Meaning, without the ControlRad system installed in Cath lab room 5. Without ControlRad System: Cardiac Catheterization or electrophysiology implant procedures in cath lab room 5 without the ControlRad system device. |
|
|
| Primary | Radiation Dose to Primary Operator (Luxel Dosimeter), Cardiac Catheterization | Mean effective radiation dose in microsievert (μSv) to the primary operator at the thyroid. This was measured using a single Luxel dosimeter over all cases to assess the cumulative dose in the study arm. | This outcome measured only the cardiac catheterization procedures, 42 participants in the ControlRad group and 38 in the group without ControlRad. A single badge was used for all cases, the result is provided as a number per arm. The number is cumulative of the arm with no measure of dispersion. | Posted | Number | microsievert | During each cardiac catheterization, an average of 90 minutes |
|
|
|
| Primary | Radiation Dose to Primary Operator, EP Procedure (Luxel) | Mean effective radiation dose in microsievert (μSv) to the primary operator at the end of each EP implant procedure. A single Luxel dosimeter was and a cumulative dose was recorded. | This outcome measured only the EP procedures, 43 in the ControlRad group and 47 in the Without ControlRad group. A single badge was used for all cases, the result is provided as a number per arm. The number is cumulative of the arm with no measure of dispersion. | Posted | Number | microsieverts | During each EP implant procedure, an average of 90 minutes |
|
|
|
| Primary | Radiation Dose to Primary Operator, EP Procedure (RaySafe) | Median effective radiation dose in microsievert (μSv) to the primary operator at the end of each EP implant procedure using a RaySafe dosimeter. This is measured after each case. | This outcome measured only the EP procedures, 43 in the ControlRad group and 47 in the Without ControlRad group. | Posted | Median | Inter-Quartile Range | microsieverts | During each EP implant procedure, an average of 90 minutes duration |
|
|
|
| Secondary | Radiation Dose to Secondary Operator, Cardiac Catheterization (RaySafe) | Median effective radiation dose (μSv) to the secondary operator at the thyroid. This was measured using a RaySafe dosimeter to assess the radiation dose in each case. | This outcome measured only the cardiac catheterization procedures, 42 participants in the ControlRad group and 38 in the group without ControlRad | Posted | Median | Inter-Quartile Range | microsievert | During each cardiac catheterization, an average of 90 minutes |
|
|
|
| Secondary | Radiation Dose to Secondary Operator (Luxel Dosimeter), Cardiac Catheterization | Mean effective radiation dose in microsievert (μSv) to the secondary operator at the thyroid. This was measured using a single Luxel dosimeter over all cases to assess the cumulative dose in the study arm. | This outcome measured only the cardiac catheterization procedures, 42 participants in the ControlRad group and 38 in the group without ControlRad. A single badge was used for all cases, the result is provided as a number per arm. The number is cumulative of the arm with no measure of dispersion. | Posted | Number | microsievert | During each cardiac catheterization, an average of 90 minutes |
|
|
|
| Secondary | Radiation Dose to Nurse or Anesthetist, Cardiac Catheterization (RaySafe) | Median effective radiation dose in microsievert (μSv) to the Nurse 0r Anesthetist at the thyroid. This was measured using a RaySafe dosimeter to assess the radiation dose in each case. | This outcome measured only the cardiac catheterization procedures, 42 participants in the ControlRad group and 38 in the group without ControlRad | Posted | Median | Inter-Quartile Range | microsievert | During each cardiac catheterization, an average of 90 minutes |
|
|
|
| Secondary | Radiation Dose to Nurse or Anesthetist, Cardiac Catheterization (Luxel) | Mean effective radiation dose in microsievert (μSv) to the Nurse or Anesthetist at the thyroid. This was measured using a single Luxel dosimeter over all cases to assess the cumulative dose in the study arm. | This outcome measured only the cardiac catheterization procedures, 42 participants in the ControlRad group and 38 in the group without ControlRad | Posted | Number | microsievert | During each cardiac catheterization, an average of 90 minutes |
|
|
|
| Secondary | Radiation Dose to Patient, Cardiac Catheterization | Median radiation dose-area product (DAP) to the patient at the end of each cardiac catheterization case using data from a fluoroscopy machine. DAP is defined as the absorbed dose multiplied by the area irradiated micro-gray μGy*m^2. | This outcome measured only the cardiac catheterization procedures, 42 participants in the ControlRad group and 38 in the group without ControlRad | Posted | Median | Inter-Quartile Range | μGy*m^2 | During each cardiac catheterization, an average of 90 minutes |
|
|
|
| Secondary | Radiation Dose in Cath Lab Room #5, Cardiac Catheterization | Mean radiation dose microsieverts (μSv) at the end of each cardiac catheterization case at a standard location on the wall of the catheterization laboratory. This was measured using a single Luxel dosimeter over all cases to assess the cumulative dose in the study arm. | This outcome measured only the cardiac catheterization procedures, 42 participants in the ControlRad group and 38 in the group without ControlRad. A single badge was used for all cases, the result is provided as a number per arm. The number is cumulative of the arm with no measure of dispersion. | Posted | Number | microsieverts | During each cardiac catheterization, an average of 90 minutes |
|
|
|
| Secondary | Radiation Dose to Secondary Operator, EP Procedure (RaySafe) | Mean effective radiation dose (μSv) to the secondary operator at the at the thyroid. This was measured using a RaySafe dosimeter to assess the radiation dose in each case. | This outcome measured only the EP procedures, 43 in the ControlRad group and 47 in the Without ControlRad group. | Posted | Median | Inter-Quartile Range | microsieverts | During each EP implant procedure, an average of 90 minutes |
|
|
|
| Secondary | Radiation Dose to Secondary Operator, EP Procedure (Luxel) | Mean effective radiation dose (μSv) to the secondary operator at the at the thyroid. This was measured using a single Luxel dosimeter over all cases to assess the cumulative dose in the study arm. | This outcome measured only the EP procedures, 43 in the ControlRad group and 47 in the Without ControlRad group. A single badge was used for all cases, the result is provided as a number per arm. The number is cumulative of the arm with no measure of dispersion. | Posted | Number | microsieverts | During each EP implant procedure, an average of 90 minutes |
|
|
|
| Secondary | Radiation Dose to Nurse or Anesthetist, EP Procedure (RaySafe) | Median effective radiation dose in microsievert (μSv) to the primary operator at the thyroid. This was measured using a RaySafe dosimeter to assess the radiation dose in each case. | This outcome measured only the EP procedures, 43 in the ControlRad group and 47 in the Without ControlRad group. | Posted | Median | Inter-Quartile Range | microsieverts | During each EP implant procedure, an average of 90 minutes |
|
|
|
| Secondary | Radiation Dose to Nurse or Anesthetist, EP Procedure (Luxel) | Mean effective radiation dose in microsievert (μSv) to the primary operator at the thyroid. This was measured using a single Luxel dosimeter over all cases to assess the cumulative dose in the study arm. | This outcome measured only the EP procedures, 43 in the ControlRad group and 47 in the Without ControlRad group. A single badge was used for all cases, the result is provided as a number per arm. The number is cumulative of the arm with no measure of dispersion. | Posted | Number | microsieverts | During each EP implant procedure, an average of 90 minutes |
|
|
|
| Secondary | Radiation Dose to Patient, EP Procedure | Mean radiation dose-area product (DAP) to the patient at the end of each EP implant procedure using data from a fluoroscopy machine. DAP is defined as the absorbed dose multiplied by the area irradiated (μGy*m^2) | This outcome measured only the EP procedures, 43 in the ControlRad group and 47 in the Without ControlRad group. | Posted | Median | Inter-Quartile Range | μGy*m^2 | During each EP implant procedure, an average of 30-45 minutes of active radiation |
|
|
|
| Secondary | Radiation Dose in Cath Lab Room #5, EP Procedure | Mean radiation dose (μSv) at the end of each EP implant procedure at a standard location on the wall of the catheterization laboratory. This was measured using a single Luxel dosimeter over all cases to assess the cumulative dose in the study arm. | This outcome measured only the EP procedures, 43 in the ControlRad group and 47 in the Without ControlRad group. A single badge was used for all cases, the result is provided as a number per arm. The number is cumulative of the arm with no measure of dispersion. | Posted | Number | microsieverts | During each EP implant procedure, an average of 90 minutes |
|
|
|
| 0 |
| 85 |
| 0 |
| 85 |
| 0 |
| 85 |
| EG001 | Without ControlRad System | Participant will undergo the cardiac catheterization or electrophysiology implant procedures per standard of care. Meaning, without the ControlRad system installed in Cath lab room 5. Without ControlRad System: Cardiac Catheterization or electrophysiology implant procedures in cath lab room 5 without the ControlRad system device. | 0 | 85 | 1 | 85 | 0 | 85 |