| Primary | Efficacy Measured as Percentage of Time Spent in Optimal Range | Objective of this study is to compare the efficacy of the investigational automated ventilation (INTELLiVENT-ASV) with a control ventilation group (conventional non-automated ventilation) in ICU subjects. The hypothesis is that INTELLiVENT-ASV is better than conventional non-automated ventilation in terms of efficacy. Efficacy endpoints are composite endpoints & are based on six variables that will be continuously recorded during the study period: These 6 variables: tidal volume, respiratory rate, end tidal CO2, SpO2, maximum inspiratory pressure, and maximum inspiratory pressure minus positive end-expiratory pressure. Each breath was analyzed and the breath was considered as optimal when all six variables were within predefined optimal ranges | | Posted | | Median | Inter-Quartile Range | percentage of time in optimal ranges | | up to 7 days after enrollment | | | | ID | Title | Description |
|---|
| OG000 | INTELLiVENT-ASV | INTELLiVENT-ASV is a software accessory that automatically adjusts ventilation and oxygenation variables to keep the patient within clinician-set target ranges, from intubation until extubation. INTELLiVENT-ASV: INTELLiVENT-ASV is software accessory that automatically adjust ventilation and oxygenation variables to keep the patient within clinician-set target ranges. | | OG001 | Non-Automated Ventilation | In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. Non-automated ventilation: In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. |
| | | Title | Denominators | Categories |
|---|
| | | Title | Measurements |
|---|
| - OG00028(19 to 37)
- OG00129(8 to 39)
|
|
| |
| Primary | Safety Measured as Percentage of Time Spent in Sub-optimal Range | Objective of this study is to compare the safety of the investigational automated ventilation (INTELLiVENT-ASV) with a control ventilation group (conventional non-automated ventilation) in ICU subjects. The hypothesis is that INTELLiVENT-ASV is better than conventional non-automated ventilation in terms of safety. Safety endpoints are composite endpoints & are based on six variables that will be continuously recorded during the study period: sub-optimal range: tidal volume, respiratory rate, end tidal CO2, SpO2, maximum inspiratory pressure, and maximum inspiratory pressure minus positive end-expiratory pressure. Each breath was analyzed and the breath was considered as sub-optimal when at least one of the six variables was within predefined sub-optimal ranges. | | Posted | | Median | Inter-Quartile Range | percentage of time in sub-optimal range | | 7 days after enrollment. | | | | ID | Title | Description |
|---|
| OG000 | INTELLiVENT-ASV | INTELLiVENT-ASV is a software accessory that automatically adjusts ventilation and oxygenation variables to keep the patient within clinician-set target ranges, from intubation until extubation. INTELLiVENT-ASV: INTELLiVENT-ASV is software accessory that automatically adjust ventilation and oxygenation variables to keep the patient within clinician-set target ranges. | | OG001 | Non-Automated Ventilation | In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. Non-automated ventilation: In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. |
|
| Secondary | Fraction of Inspiration Oxygen (FiO2) (%) | FiO2 was measured breath-by-breath for all the study periods. | | Posted | | Median | Inter-Quartile Range | percentage of inspired oxygen | | up to 28 days after enrollment | | | | ID | Title | Description |
|---|
| OG000 | INTELLiVENT-ASV | INTELLiVENT-ASV is a software accessory that automatically adjusts ventilation and oxygenation variables to keep the patient within clinician-set target ranges, from intubation until extubation. INTELLiVENT-ASV: INTELLiVENT-ASV is software accessory that automatically adjust ventilation and oxygenation variables to keep the patient within clinician-set target ranges. | | OG001 | Non-Automated Ventilation | In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. Non-automated ventilation: In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. |
| |
| Secondary | Expiratory Time Constant (s) | Expiratory time constant was calculated breath-by-breath as the ratio of volume divided by flow at 75% of the expired volume. | | Posted | | Median | Inter-Quartile Range | second (s) | | up to 28 days after enrollment | | | | ID | Title | Description |
|---|
| OG000 | INTELLiVENT-ASV | INTELLiVENT-ASV is a software accessory that automatically adjusts ventilation and oxygenation variables to keep the patient within clinician-set target ranges, from intubation until extubation. INTELLiVENT-ASV: INTELLiVENT-ASV is software accessory that automatically adjust ventilation and oxygenation variables to keep the patient within clinician-set target ranges. | | OG001 | Non-Automated Ventilation | In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. Non-automated ventilation: In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. |
| |
| Secondary | Positive End-Expiratory Pressure (PEEP) (cm H2O) | PEEP was measured breath-by-breath for all the study period. | | Posted | | Median | Inter-Quartile Range | Cm H2O | | up to 28 days after enrollment | | | | ID | Title | Description |
|---|
| OG000 | INTELLiVENT-ASV | INTELLiVENT-ASV is a software accessory that automatically adjusts ventilation and oxygenation variables to keep the patient within clinician-set target ranges, from intubation until extubation. INTELLiVENT-ASV: INTELLiVENT-ASV is software accessory that automatically adjust ventilation and oxygenation variables to keep the patient within clinician-set target ranges. | | OG001 | Non-Automated Ventilation | In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. Non-automated ventilation: In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. |
| |
| Secondary | Tidal Volume (ml) | Tidal volume was measured breath-by-breath for all the study period. | | Posted | | Median | Inter-Quartile Range | milliliter (ml) | | up to 28 days after enrollment | | | | ID | Title | Description |
|---|
| OG000 | INTELLiVENT-ASV | INTELLiVENT-ASV is a software accessory that automatically adjusts ventilation and oxygenation variables to keep the patient within clinician-set target ranges, from intubation until extubation. INTELLiVENT-ASV: INTELLiVENT-ASV is software accessory that automatically adjust ventilation and oxygenation variables to keep the patient within clinician-set target ranges. | | OG001 | Non-Automated Ventilation | In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. Non-automated ventilation: In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. |
| |
| Secondary | Tidal Volume mL/kg Ideal Body Weight (IBW) | Tidal volume mL/kg ideal body weight (IBW) will be assessed automatically breath-by-breath | | Posted | | Median | Inter-Quartile Range | ml/kg of ideal body weight (IBW) | | up to 28 days after enrollment | | | | ID | Title | Description |
|---|
| OG000 | INTELLiVENT-ASV | INTELLiVENT-ASV is a software accessory that automatically adjusts ventilation and oxygenation variables to keep the patient within clinician-set target ranges, from intubation until extubation. INTELLiVENT-ASV: INTELLiVENT-ASV is software accessory that automatically adjust ventilation and oxygenation variables to keep the patient within clinician-set target ranges. | | OG001 | Non-Automated Ventilation | In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. Non-automated ventilation: In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. |
| |
| Secondary | Ventilation - Respiratory Rate | Respiratory rate was measured breath-by-breath for all the study period. | | Posted | | Median | Inter-Quartile Range | breath/min | | up to 28 days after enrollment | | | | ID | Title | Description |
|---|
| OG000 | INTELLiVENT-ASV | INTELLiVENT-ASV is a software accessory that automatically adjusts ventilation and oxygenation variables to keep the patient within clinician-set target ranges, from intubation until extubation. INTELLiVENT-ASV: INTELLiVENT-ASV is software accessory that automatically adjust ventilation and oxygenation variables to keep the patient within clinician-set target ranges. | | OG001 | Non-Automated Ventilation | In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. Non-automated ventilation: In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. |
| |
| Secondary | Numbers of Manual Settings Changes Per Day | The total number of automatic setting changes was recorded during the study period. Automatic settings included inspiratory pressure, pressure support, respiratory rate, PEEP, and inspired oxygen fraction. | | Posted | | Median | Inter-Quartile Range | manual setting changes per day | | up to 28 days after enrollment | | | | ID | Title | Description |
|---|
| OG000 | INTELLiVENT-ASV | INTELLiVENT-ASV is a software accessory that automatically adjusts ventilation and oxygenation variables to keep the patient within clinician-set target ranges, from intubation until extubation. INTELLiVENT-ASV: INTELLiVENT-ASV is software accessory that automatically adjust ventilation and oxygenation variables to keep the patient within clinician-set target ranges. | | OG001 | Non-Automated Ventilation | In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. Non-automated ventilation: In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. |
| |
| Secondary | Automatic Settings Changes Per Day | The total number of automatic setting changes was recorded during the study period. Automatic settings included inspiratory pressure, pressure support, respiratory rate, PEEP, and inspired oxygen fraction. | | Posted | | Median | Inter-Quartile Range | automatic setting changes per day | | up to 28 days after enrollment | | | | ID | Title | Description |
|---|
| OG000 | INTELLiVENT-ASV | INTELLiVENT-ASV is a software accessory that automatically adjusts ventilation and oxygenation variables to keep the patient within clinician-set target ranges, from intubation until extubation. INTELLiVENT-ASV: INTELLiVENT-ASV is software accessory that automatically adjust ventilation and oxygenation variables to keep the patient within clinician-set target ranges. | | OG001 | Non-Automated Ventilation | In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. Non-automated ventilation: In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. |
| |
| Other Pre-specified | Mortality at 28 Days (n (%)) | | | Posted | | Count of Participants | | Participants | | up to 28 days after enrollment | | | | ID | Title | Description |
|---|
| OG000 | INTELLiVENT-ASV | INTELLiVENT-ASV is a software accessory that automatically adjusts ventilation and oxygenation variables to keep the patient within clinician-set target ranges, from intubation until extubation. INTELLiVENT-ASV: INTELLiVENT-ASV is software accessory that automatically adjust ventilation and oxygenation variables to keep the patient within clinician-set target ranges. | | OG001 | Non-Automated Ventilation | In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. Non-automated ventilation: In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. |
| |
| Other Pre-specified | Duration Invasive Mechanical Ventilation (Days) | Mechanical ventilation duration was recorded from the date of intubation to the date of last extubation or death if the patient died while still intubated. | | Posted | | Median | Inter-Quartile Range | days | | up to 28 days from enrollment | | | | ID | Title | Description |
|---|
| OG000 | INTELLiVENT-ASV | INTELLiVENT-ASV is a software accessory that automatically adjusts ventilation and oxygenation variables to keep the patient within clinician-set target ranges, from intubation until extubation. INTELLiVENT-ASV: INTELLiVENT-ASV is software accessory that automatically adjust ventilation and oxygenation variables to keep the patient within clinician-set target ranges. | | OG001 | Non-Automated Ventilation | In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. Non-automated ventilation: In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. |
| |
| Other Pre-specified | Passive Ventilation Duration (h) | Passive ventilation is defined when the percentage of subject's triggered breath is lower than 25% of total respiratory rate. | | Posted | | Median | Inter-Quartile Range | hours (h) | | up to 28 days from enrollment | | | | ID | Title | Description |
|---|
| OG000 | INTELLiVENT-ASV | INTELLiVENT-ASV is a software accessory that automatically adjusts ventilation and oxygenation variables to keep the patient within clinician-set target ranges, from intubation until extubation. INTELLiVENT-ASV: INTELLiVENT-ASV is software accessory that automatically adjust ventilation and oxygenation variables to keep the patient within clinician-set target ranges. | | OG001 | Non-Automated Ventilation | In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. Non-automated ventilation: In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. |
| |
| Other Pre-specified | Ventilator Free Days at Day 28 (d) | | | Posted | | Median | Inter-Quartile Range | days | | up to 28 days from enrollment | | | | ID | Title | Description |
|---|
| OG000 | INTELLiVENT-ASV | INTELLiVENT-ASV is a software accessory that automatically adjusts ventilation and oxygenation variables to keep the patient within clinician-set target ranges, from intubation until extubation. INTELLiVENT-ASV: INTELLiVENT-ASV is software accessory that automatically adjust ventilation and oxygenation variables to keep the patient within clinician-set target ranges. | | OG001 | Non-Automated Ventilation | In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. Non-automated ventilation: In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. |
| |
| Other Pre-specified | Percentage of Time in Allocated Mode (%) | The percentage of time in the allocated mode was calculated breath-by-breath according to the ventilation mode used. | | Posted | | Median | Inter-Quartile Range | percentage | | up to 28 days after enrollment | | | | ID | Title | Description |
|---|
| OG000 | INTELLiVENT-ASV | INTELLiVENT-ASV is a software accessory that automatically adjusts ventilation and oxygenation variables to keep the patient within clinician-set target ranges, from intubation until extubation. INTELLiVENT-ASV: INTELLiVENT-ASV is software accessory that automatically adjust ventilation and oxygenation variables to keep the patient within clinician-set target ranges. | | OG001 | Non-Automated Ventilation | In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. Non-automated ventilation: In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. |
| |
| Other Pre-specified | Efficacy Measured as Percentage of Time Spent in Optimal Range | Percentage of time spent in optimal range was calculated breath-by-breath during all the study period. Breath were classified as optimal when all six variables were in predefined optimal range: tidal volume, respiratory rate, end tidal CO2, SpO2, maximum inspiratory pressure, and maximum inspiratory pressure minus positive end-expiratory pressure. | | Posted | | Median | Inter-Quartile Range | percentage of time in optimal ranges | | up to 28 days from enrollment | | | | ID | Title | Description |
|---|
| OG000 | INTELLiVENT-ASV | INTELLiVENT-ASV is a software accessory that automatically adjusts ventilation and oxygenation variables to keep the patient within clinician-set target ranges, from intubation until extubation. INTELLiVENT-ASV: INTELLiVENT-ASV is software accessory that automatically adjust ventilation and oxygenation variables to keep the patient within clinician-set target ranges. | | OG001 | Non-Automated Ventilation | In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. Non-automated ventilation: In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. |
| |
| Other Pre-specified | Safety Measured as Percentage of Time Spent in Sub-optimal Range | Percentage of time spent in sub-optimal range was calculated breath-by-breath during all the study period. Breath were classified as sub-optimal when at least one variable was in predefined sub-optimal range: tidal volume, respiratory rate, end tidal CO2, SpO2, maximum inspiratory pressure, and maximum inspiratory pressure minus positive end-expiratory pressure. | | Posted | | Median | Inter-Quartile Range | percentage of time in sub-optimal ranges | | up to 28 days from enrollment | | | | ID | Title | Description |
|---|
| OG000 | INTELLiVENT-ASV | INTELLiVENT-ASV is a software accessory that automatically adjusts ventilation and oxygenation variables to keep the patient within clinician-set target ranges, from intubation until extubation. INTELLiVENT-ASV: INTELLiVENT-ASV is software accessory that automatically adjust ventilation and oxygenation variables to keep the patient within clinician-set target ranges. | | OG001 | Non-Automated Ventilation | In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. Non-automated ventilation: In conventional ventilation modes, clinicians adjust and modify the oxygenation and ventilation parameters manually. |
| |