Fluid therapy guided by stroke volume response to repeated fluid challenges is used for hemodynamic optimization during general anesthesia. Two types of devices that measure stroke volume non-invasively are primarily available for intermediate-risk surgery under general anesthesia. They are based on two different techniques, (1) pulse contour analysis, and (2) transthoracic bioimpedance. They have never been compared with each other. Therefore, it is not known (1) whether they are interchangeable for the measurement of stroke volume changes (usually assessed as the "concordance" of two techniques), and (2) whether one is better than the other.
The aim is to study the concordance of these two non-invasive hemodynamic devices. it will use both of them in patients undergoing intermediate risk surgery. They will be tested during modifications of cardiac preload induced by Trendelenburg and anti-Trendelenburg positioning, as well as during intraoperative fluid challenges, vasopressor boluses and alveolar recruitment maneuvers.
The main objective is to assess the concordance of two non-invasive devices using the technology described above : the Clearsight (Edwards Lifesciences, pulse contour analysis), and the Niccomo (Medis, bioimpedance). Both will be used in each patient included in the study. They will be tested during modifications of cardiac preload induced by sequences of position changes (see below) including Trendelenburg and anti-Trendelenburg positioning, performed systematically before and after the procedure, as well as during fluid challenges, vasopressor boluses and alveolar recruitment maneuvers (performed during the procedure according to the needs of the anesthesiologist in charge of the patient). Both monitors will also be compared with an invasive monitor of stroke volume (Vigileo/Flotrac, Edwards Lifesciences), and the respiratory pulse pressure variation will also be measured during the sequences of position changes and will be used as a gold standard for fluid responsiveness.
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Sequence A before sequence B | Experimental | Patients eligible for intermediate risk surgery requiring the placement of an invasive continuous blood pressure measurement device and stroke volume monitoring for fluid management. Sequence A of positioning (before the procedure), and sequence B after the procedure. |
|
| Sequence B before sequence A | Experimental | Patients eligible for intermediate risk surgery requiring the placement of an invasive continuous blood pressure measurement device and stroke volume monitoring for fluid management. Sequence B of positioning before the procedure, and sequence A after the procedure. |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| CLEARSIGHT and NICCOMO measurements in sequence A | Device | Starting from 0° position, shift to position +20°, then tilt to -20°, and finally back to the 0° position. |
|
| Measure | Description | Time Frame |
|---|---|---|
| concordance of stroke volume changes measured with each monitor during the standardized maneuvers of sequences A and B (total: 6 observations per individual). | 4-quadrant plot | during each changes of positioning (i.e., for 1-2 minutes in each condition, for a total of 10-15 minutes following initial measurements in basal conditions), during sequences A and B |
| Measure | Description | Time Frame |
|---|---|---|
| Fluid responsiveness defined by areas under the Receiver Operating Characteristic curve (sensitivity and specificity) | Comparison of areas under the Receiver Operating Characteristic curve (sensitivity and specificity). Each maneuver will be classified as "responder" or "nonresponder" according to the value of Pulse Pressure Variation (PPV) recorded before the maneuver and to the changes in PPV during the maneuver (as previously published) |
Not provided
Inclusion Criteria:
There is no definitive definition of intermediate risk. By reference to the literature, surgeries longer than 2 hours (excluding "low risk") and less than 3% postoperative mortality (excluding high-risk surgery). It includes the majority of non-serious visceral (abdominal and gynecological) surgery, plastic or cervical-facial surgery for neoplastic removal or long-term reconstruction, or long-term spinal surgery or neurosurgery of the base of the skull.
- For which the anesthesiologist in charge has planned an invasive monitoring (radial catheter) of arterial blood pressure.
Exclusion Criteria:
Urgent surgery
Intracranial hypertension
Orthostatic hypotension
Extreme anthropometry: weight >150 kg and 30 kg, size 120 cm and >230 cm
Related to NICCOMO monitoring:
Related to CLEARSIGHT monitoring
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Benoit Tavernier, MD,PhD | University Hospital Lille, France | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| Hôpital Roger Salengro, CHU Lille | Lille | 59037 | France |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
The primary outcome will be based on measurements performed during a sequence of position changes.
Sequence A will consist of starting from 0° position, shift to position +20°, then tilt to -20°, and finally back to the 0° position.
Sequence B will consist of starting from 0° position, shift to position -20°, then tilt to +20°, and finally back to the 0° position.
Patients will be randomized to take sequence A before the procedure, and then the sequence B at the end of the procedure, or vice-versa (sequence B before sequence A).
The randomization is not intended to compare the two groups with each other, but only to compensate for a possible"order-effect" in the execution of the maneuvers.
Not provided
Not provided
Not provided
|
| CLEARSIGHT and NICCOMO measurements in sequence B | Device | Starting from 0° position, shift to position -20°, then tilt to +20°, and finally back to the 0° position and collect the last group of figures. |
|
|
| during each changes of positioning (i.e., for 1-2 minutes in each condition, for a total of 10-15 minutes following initial measurements in basal conditions), during sequences A and B |
| concordance of stroke volume changes measured with each monitor during the standardized maneuvers mimicking preload decrease. | 4-quadrant plot | during each changes of positioning where cardiac preload is supposed to decrease (i.e., for 1-2 minutes in one out of two condition, for a total of 10-15 minutes following initial measurements in basal conditions), during sequences A and B |
| concordance of stroke volume changes measured with each monitor during the standardized maneuver mimicking preload increase. | 4-quadrant plot | during each changes of positioning where cardiac preload is supposed to increase (i.e., for 1-2 minutes in one out of two condition, for a total of 10-15 minutes following initial measurements in basal conditions), during sequences A and B |
| concordance of stroke volume changes measured with each monitor during the surgical procedure (fluid challenges, vasopressor boluses, alveolar recruitment maneuver). | 4-quadrant plot | throughout surgery (mean length expected between 3 and 5 hours). Measurements performed during sequences of 5-20 minutes that can occur several times during surgery |
| interchangeability of stroke volume measurements with each monitor during sequences A and B | Bland and Altman plot for repeated measures | before and after each changes of positioning (i.e., for 1-2 minutes in each condition, for a total of 10-15 minutes following initial measurements in basal conditions), during sequences A and B |
| interchangeability of stroke volume measured with each monitor during the surgical procedure (fluid challenges, vasopressor boluses, alveolar recruitment maneuver). | Bland and Altman plot for repeated measures | throughout surgery (mean length expected between 3 and 5 hours). Measurements performed before and after sequences of 5-20 minutes that can occur several times during surgery |
| concordance of stroke volume changes measured with each monitor with that measured by a Vigileo/Flotrac monitor during the standardized maneuvers of sequences A and B (total: 6 observations per individual). | 4-quadrant plot | during each changes of positioning (i.e., for 1-2 minutes in each condition, for a total of 10-15 minutes following initial measurements in basal conditions), during sequences A and B |