The aim of this preliminary study was to evaluate the new CompuFlo instrument which allows pulsatile waveform recordings.
30 epidural catheters previously successfully used for obstetric anesthesia or analgesia and about to be removed were tested. All patients were given 5 mL 2% lidocaine to test the catheter before its removal. After priming with 5 mL saline, the catheter was connected to CompuFlo to record the occurrence of pulsatile waveforms and/or their disappearance during its removal. The epidural catheter was marked at the skin level to record the distance between the skin at the time of measurements. The power analysis required a sample of 30 observations to set 80% test power and 95% significance level.
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| CompuFlo | Device | After priming with 5 mL saline, the catheter was connected to CompuFlo to record the occurrence of pulsatile waveforms and/or their disappearance during its removal |
| Measure | Description | Time Frame |
|---|---|---|
| pulsatile waveforms observation (presence or absence) | The epidural catheter was considered to be in the epidural space if the appearance of pulsatile waveforms were observed, and confirmed by the occurrence of L2-3 sensory block after the test dose. The disappearance of the pulsatile waveforms when the catheter was extracted from the epidural space was also noted. | 30 minutes |
Not provided
Not provided
Inclusion Criteria:
Exclusion criteria
• Non-obstetric patients
pregnant patients
Obstetric patients who received an epidural block and have an epidural catheter
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Giorgio Capogna, MD | European e-Learning School in Obstetric Anesthesia | Study Director |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| EESOA Maternal Neonatal Simulation Centre | Roma | 00151 | Italy | |||
| Città di Roma Hospital |
| PubMed Identifier | Type | Citation | Retractions |
|---|---|---|---|
| 30046305 | Background | Capogna G, Coccoluto A, Capogna E, Del Vecchio A. Objective Evaluation of a New Epidural Simulator by the CompuFlo(R) Epidural Instrument. Anesthesiol Res Pract. 2018 Jun 26;2018:4710263. doi: 10.1155/2018/4710263. eCollection 2018. |