Oxygen Savings With Administered Oxygen and High Flow Ambient Air At Rest
Oxygen Savings With Administered Oxygen and High Flow Ambient Air At Rest
This study is meant to compare the amount of oxygen required for hypoxemia relief between current standard of care (oxygen only) and oxygen with the addition of high flow air for Chronic Obstructive Pulmonary Disease (COPD), Interstitial Lung Disease (ILD), and Pulmonary Hypertension (PH) patients during rest. Subjects will be titrated from 0 L/min until they maintain 95% SpO2 for each of the following delivery methods:
Each participant will require (1) visit, which will last approximately 4-5 hours.
At start of visit:
Pulsed Oxygen Control at Rest:
Participant's oxygen will be turned off for 10 minutes as a washout period.
Using the oxygen tank with the pulse regulator, the participant will be titrated to a pulse rate that maintains their SpO2 at an average of 95%.
a. Increasing volumes of oxygen in 1 L(liter)/min increments every thirty seconds until an average of 95% SpO2 is maintained for 30 seconds.
Once the participant is titrated to the correct amount of oxygen, they will remain on the oxygen for an additional 3 minutes
Final oxygen flow rate, SpO2, Heart Rate, and Borg Dyspnea Score will be recorded in the last minute of the test.
Mixed Continuous Air/Oxygen Efficacy at Rest:
Participant's oxygen will be turned off for 10 minutes as a washout period.
The participant will be delivered pulsed oxygen and high flow ambient air using the oxygen tank with the pulse regulator and Vapotherm setup, the participant will be titrated to a pulse rate that maintains their SpO2 at an average of 95%.
Once the participant is titrated the correct amount of oxygen, they will remain on the oxygen for an additional 3 minutes.
Final oxygen flow rate, SpO2, Heart Rate, and Borg Dyspnea Score will be recorded in the last minute of the test.
Steps 5-8 are then repeated two more times for pulsed oxygen plus continuous flow ambient air at flow rates of 20 and 25 L/min.
Mixed Pulsed Air/Oxygen Efficacy at Rest:
Participant's oxygen will be turned off for 10 minutes as a washout period
The participant will be delivered the out-of-phase pulsed oxygen and high flow ambient air using the oxygen tank with the pulse regulator and Vapotherm setup, the participant will be titrated to a pulse rate that maintains their SpO2 at an average of 95%
Once the participant is titrated the correct amount of oxygen, they will remain on the oxygen for an additional 3 minutes
Final oxygen flow rate, SpO2, Heart Rate, and Borg Dyspnea Score will be recorded in the last minute of the test.
Steps 10-13 are then repeated two more times for out-of phase pulsed oxygen plus pulsed high flow ambient air at flow rates of 20 and 25 L/min.
Inclusion Criteria:
Exclusion Criteria: