The goal of the study was to compare the current gold standard blood pressure instrument (a mercury-filled sphygmomanometer with cuff and stethoscope) with the new experimental algorithm developed by HeartBeat Technologies Ltd. To that end, participants were recruited from Markham, Ontario and participated in a series of alternating blood pressure measurements over a period of 45 minutes. Comparisons were made between manual measurements and device measurements to determine the quality of the device readings. Measurements were taken by trained nurses and staff were consistent across the study duration.
The gold standard instrument for blood pressure measurement is a mercury-filled sphygmomanometer with cuff and stethoscope (Canadian Agency for Drugs and Technologies in Health; Canadian Electronic Library, 2012). However, such a device is not the most convenient for client self-monitoring or virtual (remote) care. Thus, there is a desire for an accurate, reliable alternative for the traditional mercury-containing sphygmomanometer.
HeartBeat Technologies Ltd. ("Heartbeat") has developed a novel approach to measuring blood pressure after an initial blood pressure reading using the conventional measurement method, supplemented by specific characteristics of a person (age, gender, height, weight, and heart rate), has established a "baseline measurement" for the person. The "novel approach" uses a finger pulse oximeter, the Contec CMS50EW device (Figure 1) (manufactured by Contec Medical Systems Co., Ltd., Shanghai, People's Republic of China). The finger pulse oximeter detects the changes in blood volume directly below the person's skin and indirectly measures oxygen saturation in the blood. The measurement is in the form of a photoplethysmogram (PPG) which is captured by the Contec oximeter and then, by Bluetooth technology, transmitted to a smartphone or tablet. A HeartBeat application, called MediBeat, on the smartphone or tablet then transmits the PPG to a server where a proprietary algorithm analyzes the baseline measurement for the person and the PPG to calculate the person's current blood pressure.
Heartbeat's intent is that the current blood pressure reading would then be made available to both the client and the health provider through separate web-based interfaces. Providers would be able to log onto the HeartBeat web-based application from anywhere to review and monitor the blood pressure of multiple clients, while clients can log onto the application to track their own measurements.
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Low Blood Pressure | Other | Low blood pressure (systolic: 0-129, diastolic: 0-79) |
|
| Medium Blood Pressure | Other | Medium blood pressure (systolic: 130-160, diastolic: 80-100) |
|
| High Blood Pressure | Other | High blood pressure (systolic: 161 or higher, diastolic: 101 or higher) |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| Contec CMS50EW | Device | Participants' blood pressure will be measured 3 times with the wearable finger pulse oximeter at their study visit. |
|
| Measure | Description | Time Frame |
|---|---|---|
| Difference Between Systolic Manual Sphygmomanometer and Finger Pulse Oximeter Readings | Each device reading was compared to the manual blood pressure reading taken before (R1) and after it (R2). The degree of alignment for systolic blood pressure was calculated by taking the difference between the two scores (manual and device). The reading with the best alignment was used. | 45 - 60 minutes |
| Difference Between Diastolic Manual and Device Blood Pressure Readings | Each device reading was compared to the manual reading before (R1) and after (R2) it, to calculate the degree of alignment. The difference of the device and R1, and the device and R2 were taken. The smaller number was used for results reporting. | 45-60 minutes |
Not provided
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Paul Holyoke, PhD | SE Health | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| SE Health | Markham | Ontario | L3R 6H3 | Canada |
| PubMed Identifier | Type | Citation | Retractions |
|---|---|---|---|
| 33890856 | Derived | Holyoke P, Yogaratnam K, Kalles E. Web-Based Smartphone Algorithm for Calculating Blood Pressure From Photoplethysmography Remotely in a General Adult Population: Validation Study. J Med Internet Res. 2021 Apr 23;23(4):e19187. doi: 10.2196/19187. |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Healthy participants who were 25 years old and worked in Markham were recruited. Only healthy subjects were included; those who were under 25, had heart arrhythmias, atrial fibrillations or atrial flutters, an inaudible Korotkoff sound, wounds of the upper arms/wrists, missing fingers, and/or an arm circumference of more than 55 cm were excluded.
Not provided
| ID | Title | Description |
|---|---|---|
| FG000 | Low Blood Pressure | Low blood pressure (systolic: 0-129, diastolic: 0-79) Contec CMS50EW: Participants' blood pressure will be measured 3 times with the wearable finger pulse oximeter at their study visit. Sphygmomanometer: Participants' blood pressure will be measured 4 times manually with a blood pressure cuff at their study visit. |
| FG001 | Medium Blood Pressure | Medium blood pressure (systolic: 130-160, diastolic: 80-100) Contec CMS50EW: Participants' blood pressure will be measured 3 times with the wearable finger pulse oximeter at their study visit. Sphygmomanometer: Participants' blood pressure will be measured 4 times manually with a blood pressure cuff at their study visit. |
| FG002 | High Blood Pressure | High blood pressure (systolic: 161 or higher, diastolic: 101 or higher) Contec CMS50EW: Participants' blood pressure will be measured 3 times with the wearable finger pulse oximeter at their study visit. Sphygmomanometer: Participants' blood pressure will be measured 4 times manually with a blood pressure cuff at their study visit. |
| Title | Milestones | Reasons Not Completed | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Overall Study |
|
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| BG000 | Low Blood Pressure | Low blood pressure (systolic: 0-129, diastolic: 0-79) Contec CMS50EW: Participants' blood pressure will be measured 3 times with the wearable finger pulse oximeter at their study visit. Sphygmomanometer: Participants' blood pressure will be measured 4 times manually with a blood pressure cuff at their study visit. |
| Units | Counts |
|---|---|
| Participants |
|
| Title | Description | Population Description | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Denominator Units Selected | Denominators | Classes |
|---|---|---|---|---|---|---|---|---|---|
| Age, Continuous | Mean |
| Type | Title | Description | Population Description | Reporting Status | Anticipated Posting Date | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Time Frame | Units Analyzed | Denominator Units Selected | Arm/Group Information | Denominators | Classes | Analyses | |||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Primary | Difference Between Systolic Manual Sphygmomanometer and Finger Pulse Oximeter Readings | Each device reading was compared to the manual blood pressure reading taken before (R1) and after it (R2). The degree of alignment for systolic blood pressure was calculated by taking the difference between the two scores (manual and device). The reading with the best alignment was used. | The overall participants analyzed indicate the number of individuals who matched the ranges listed in the description for low, medium or high blood pressure categories, where they had blood pressure ranges falling between the respective values indicated for the systolic pressure measured. | Posted | Mean | Standard Deviation | mmHg | 45 - 60 minutes |
|
For each participant, adverse event phenomena were monitored from the initial informed consent process and up to the remuneration process in the data gathering visit. This took approximately 45 minutes for each participant, during which they were continuously in the presence of trained medical professionals. Following this active assessment period, there was no follow-up conducted with participants, but participants were able to contact the research team with any concerns that may have arisen.
Adverse events or risks (for all-cause mortality, serious, or non-serious) were monitored throughout study duration, however, none were observed across the 94 study participants.
Not provided
| ID | Title | Description | Deaths (Affected) | Deaths (At Risk) | Serious Events (Affected) | Serious Events (At Risk) | Other Events (Affected) | Other Events (At Risk) |
|---|---|---|---|---|---|---|---|---|
| EG000 | Low Blood Pressure | Low blood pressure (systolic: 0-129, diastolic: 0-79) Contec CMS50EW: Participants' blood pressure will be measured 3 times with the wearable finger pulse oximeter at their study visit. Sphygmomanometer: Participants' blood pressure will be measured 4 times manually with a blood pressure cuff at their study visit. |
Not provided
Not provided
Not provided
| Title | Organization | Phone | Extension | |
|---|---|---|---|---|
| SE Research Centre | SE Health | N/A | research@sehc.com |
Not provided
| Type | Includes Protocol | Includes SAP | Includes ICF | Document Label | Document Date | Document Uploaded Date | Document File Name |
|---|---|---|---|---|---|---|---|
| Prot_SAP | Yes | Yes | No | Study Protocol and Statistical Analysis Plan | Nov 4, 2019 | Jan 6, 2020 | Prot_SAP_000.pdf |
Not provided
| ID | Term |
|---|---|
| D007022 | Hypotension |
| D006973 | Hypertension |
| ID | Term |
|---|---|
| D014652 | Vascular Diseases |
| D002318 | Cardiovascular Diseases |
Not provided
Not provided
| ID | Term |
|---|---|
| D019722 | Sphygmomanometers |
| ID | Term |
|---|---|
| D019719 | Diagnostic Equipment |
| D004864 | Equipment and Supplies |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Sphygmomanometer | Device | Participants' blood pressure will be measured 4 times manually with a blood pressure cuff at their study visit. |
|
| BG001 |
| Medium Blood Pressure |
Medium blood pressure (systolic: 130-160, diastolic: 80-100) Contec CMS50EW: Participants' blood pressure will be measured 3 times with the wearable finger pulse oximeter at their study visit. Sphygmomanometer: Participants' blood pressure will be measured 4 times manually with a blood pressure cuff at their study visit. |
| BG002 | High Blood Pressure | High blood pressure (systolic: 161 or higher, diastolic: 101 or higher) Contec CMS50EW: Participants' blood pressure will be measured 3 times with the wearable finger pulse oximeter at their study visit. Sphygmomanometer: Participants' blood pressure will be measured 4 times manually with a blood pressure cuff at their study visit. |
| BG003 | Total | Total of all reporting groups |
| years |
|
| Sex: Female, Male | Count of Participants | Participants |
|
| Race and Ethnicity Not Collected | Race and Ethnicity were not collected from any participant. | Count of Participants | Participants |
|
| OG001 |
| Medium Blood Pressure |
Systolic score between 130 - 160mmHg. |
| OG002 | High Blood Pressure | Systolic score greater than 161mmHg. |
|
|
| Primary | Difference Between Diastolic Manual and Device Blood Pressure Readings | Each device reading was compared to the manual reading before (R1) and after (R2) it, to calculate the degree of alignment. The difference of the device and R1, and the device and R2 were taken. The smaller number was used for results reporting. | The overall participants analyzed indicate the number of individuals who matched the ranges listed in the description for low, medium or high blood pressure categories, where they had blood pressure ranges falling between the respective values indicated for the diastolic pressure measured. | Posted | Mean | Standard Deviation | mmHg | 45-60 minutes |
|
|
|
| 0 |
| 94 |
| 0 |
| 94 |
| 0 |
| 94 |
| EG001 | Medium Blood Pressure | Medium blood pressure (systolic: 130-160, diastolic: 80-100) Contec CMS50EW: Participants' blood pressure will be measured 3 times with the wearable finger pulse oximeter at their study visit. Sphygmomanometer: Participants' blood pressure will be measured 4 times manually with a blood pressure cuff at their study visit. | 0 | 94 | 0 | 94 | 0 | 94 |
| EG002 | High Blood Pressure | High blood pressure (systolic: 161 or higher, diastolic: 101 or higher) Contec CMS50EW: Participants' blood pressure will be measured 3 times with the wearable finger pulse oximeter at their study visit. Sphygmomanometer: Participants' blood pressure will be measured 4 times manually with a blood pressure cuff at their study visit. | 0 | 94 | 0 | 94 | 0 | 94 |