After the implantation of a left ventricular assist device (LVAD), many patients continue to experience exercise intolerance. LVAD echo guided optimization (EO) determines a more favorable hemodynamic profile and could provide an improvement on functional capacity (FC). VAFRACT is a prospective randomized trial to evaluate the additional benefit of an EO approach on FC, measured by cardiopulmonary exercise test (CPET), in LVAD optimization free population.
The use of Left Ventricular Assist Devices (LVAD) is a viable therapeutic option to improve survival and quality of life of patients with advanced and refractory heart failure (HF). However, after the implantation of an LVAD, patients' functional capacity (FC) is still reduced with maximum oxygen uptake (VO2 peak) values calculated at the cardiopulmonary exercise test (CPET) ranging from 11 to 20 ml/kg/min. LVAD echo guided optimization (EO) determines a more favorable hemodynamic profile, in particular for the right ventricle, and could provide an improvement on FC. VAFRACT is a prospective randomized trial to evaluate the additional benefit of an EO approach on FC, measured by CPET in LVAD optimization free population.
The optimal device speed is defined as the one that allows an intermittent aortic valve opening and a neutral position of the interventricular septum without increasing aortic or tricuspid regurgitation associated or not to a dilatation of the right ventricle (RV). In this condition it is reasonable to expect a more favorable hemodynamic profile for the RV and a better response in terms of functional capacity.
Subjects studied are HF patients supported with a continuous-flow LVAD: HeartMate II (Thoratec Inc., Pleasanton, CA) and HeartMate 3 (HM 3, Abbott, North Chicago, IL). Patients are recruited from our Day Hospital of the Cardiac Surgery Department at the University Hospital of Verona.
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| EO GROUP | Experimental | Patients underwent LVAD echo-optimization; the optimal device speed is confirmed at the end of procedure. |
|
| CONTROL GROUP | No Intervention | Patients underwent LVAD echo-optimization, but the optimal device speed is not confirmed at the end of procedure. |
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| echo-optimization | Device | LVAD echo-optimization consists of routine comprehensive transthoracic echocardiography at the baseline speed setting, followed by stepwise incremental adjustments to the LVAD speed (revolutions per minute: rpm), with collection of prespecified echocardiographic parameters at each new speed (eg, left ventricle end-diastolic diameter, interventricular septal position, aortic valve opening frequency/duration, tricuspid and/or mitral regurgitation severity). The optimal velocity is defined as the one that allows an intermittent aortic valve opening and a neutral position of the interventricular septum without increasing aortic or tricuspid regurgitation, associated or not to a dilatation of the right ventricle. The recommended speed range varies according to the indications given in the data sheet for each specific device. |
| Measure | Description | Time Frame |
|---|---|---|
| VO2 Peak | peak oxygen uptake measured by cardiopulmonary exercise test | three months |
| Measure | Description | Time Frame |
|---|---|---|
| Number of Participants With LVAD-related Hospitalizations | Rate of hospitalizations for: device thrombosis, hemorrhagic events, infections, right heart failure, arrhythmias. | Three months |
| QoL: The EQ-5D Questionnaire |
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Affiliation | Role |
|---|---|---|
| MARZIA LILLIU, MD | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| Department of Cardiac Surgery, University of Verona | Verona | 37126 | Italy |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| FG000 | EO GROUP | Patients underwent LVAD echo-optimization; the optimal device speed is confirmed at the end of procedure. echo-optimization: LVAD echo-optimization consists of routine comprehensive transthoracic echocardiography at the baseline speed setting, followed by stepwise incremental adjustments to the LVAD speed (revolutions per minute: rpm), with collection of prespecified echocardiographic parameters at each new speed (eg, left ventricle end-diastolic diameter, interventricular septal position, aortic valve opening frequency/duration, tricuspid and/or mitral regurgitation severity). The optimal velocity is defined as the one that allows an intermittent aortic valve opening and a neutral position of the interventricular septum without increasing aortic or tricuspid regurgitation, associated or not to a dilatation of the right ventricle. The recommended speed range varies according to the indications given in the data sheet for each specific device. |
| FG001 | CONTROL GROUP | Patients underwent LVAD echo-optimization, but the optimal device speed is not confirmed at the end of procedure. |
| Title | Milestones | Reasons Not Completed | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Overall Study |
|
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| BG000 | EO GROUP | Patients underwent LVAD echo-optimization; the optimal device speed is confirmed at the end of procedure. echo-optimization: LVAD echo-optimization consists of routine comprehensive transthoracic echocardiography at the baseline speed setting, followed by stepwise incremental adjustments to the LVAD speed (revolutions per minute: rpm), with collection of prespecified echocardiographic parameters at each new speed (eg, left ventricle end-diastolic diameter, interventricular septal position, aortic valve opening frequency/duration, tricuspid and/or mitral regurgitation severity). The optimal velocity is defined as the one that allows an intermittent aortic valve opening and a neutral position of the interventricular septum without increasing aortic or tricuspid regurgitation, associated or not to a dilatation of the right ventricle. The recommended speed range varies according to the indications given in the data sheet for each specific device. |
| Units | Counts |
|---|---|
| Participants |
|
| Title | Description | Population Description | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Denominator Units Selected | Denominators | Classes |
|---|---|---|---|---|---|---|---|---|---|
| Age, Categorical | Count of Participants |
| Type | Title | Description | Population Description | Reporting Status | Anticipated Posting Date | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Time Frame | Units Analyzed | Denominator Units Selected | Arm/Group Information | Denominators | Classes | Analyses | |||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Primary | VO2 Peak | peak oxygen uptake measured by cardiopulmonary exercise test | Posted | Mean | Standard Deviation | mL/Kg/min | three months |
|
3 months
Not provided
Not provided
| ID | Title | Description | Deaths (Affected) | Deaths (At Risk) | Serious Events (Affected) | Serious Events (At Risk) | Other Events (Affected) | Other Events (At Risk) |
|---|---|---|---|---|---|---|---|---|
| EG000 | EO GROUP | Patients underwent LVAD echo-optimization; the optimal device speed is confirmed at the end of procedure. echo-optimization: LVAD echo-optimization consists of routine comprehensive transthoracic echocardiography at the baseline speed setting, followed by stepwise incremental adjustments to the LVAD speed (revolutions per minute: rpm), with collection of prespecified echocardiographic parameters at each new speed (eg, left ventricle end-diastolic diameter, interventricular septal position, aortic valve opening frequency/duration, tricuspid and/or mitral regurgitation severity). The optimal velocity is defined as the one that allows an intermittent aortic valve opening and a neutral position of the interventricular septum without increasing aortic or tricuspid regurgitation, associated or not to a dilatation of the right ventricle. The recommended speed range varies according to the indications given in the data sheet for each specific device. |
Not provided
Not provided
Not provided
| Title | Organization | Phone | Extension | |
|---|---|---|---|---|
| Marzia Lilliu, MD, PhD | Division of Cardiac Surgery, Department of Surgery - University of Verona | +39 3281383539 | marzialilliu@gmail.com |
Not provided
| Type | Includes Protocol | Includes SAP | Includes ICF | Document Label | Document Date | Document Uploaded Date | Document File Name |
|---|---|---|---|---|---|---|---|
| Prot_SAP | Yes | Yes | No | Study Protocol and Statistical Analysis Plan | Jul 19, 2017 | Oct 9, 2021 | Prot_SAP_000.pdf |
Not provided
| ID | Term |
|---|---|
| D006333 | Heart Failure |
| ID | Term |
|---|---|
| D006331 | Heart Diseases |
| D002318 | Cardiovascular Diseases |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
|
Changes in quality of life perceived by EuroQoL (quality of life) Scale. It is an instrument which evaluates the quality of life measuring with one question: mobility, self-care, usual activities, pain/discomfort, and anxiety/depression. The answers given to these question (1 = no problems; 2 = some problems; 3 = extreme problems) are converted into EQ-5D index, an utility scores anchored at 0 for death and 1 for perfect health. The EQ-5D questionnaire also includes a Visual Analog Scale (VAS), by which respondents can report their perceived health status with a grade ranging from 0 (the worst possible health status) to 100 (the best possible health status).
| Three months |
| Nt-proBNP | Nt-proBNP levels | Three months |
| Right Ventricular Function | Fractional area change of right ventricle assessed by echocardiography | Three months |
| CPET Exercise Time | Cardiopulmonary exercise time | Three months |
| BG001 | CONTROL GROUP | Patients underwent LVAD echo-optimization, but the optimal device speed is not confirmed at the end of procedure. |
| BG002 | Total | Total of all reporting groups |
| Participants |
|
| Age, Continuous | Mean | Standard Deviation | years |
|
| Sex: Female, Male | Count of Participants | Participants |
|
| Race (NIH/OMB) | Count of Participants | Participants |
|
| Region of Enrollment | Count of Participants | Participants |
|
| VO2 peak | Mean | Standard Deviation | mL/kg/min |
|
| OG001 | CONTROL GROUP | Patients underwent LVAD echo-optimization, but the optimal device speed is not confirmed at the end of procedure. |
|
|
| Secondary | Number of Participants With LVAD-related Hospitalizations | Rate of hospitalizations for: device thrombosis, hemorrhagic events, infections, right heart failure, arrhythmias. | Posted | Count of Participants | Participants | Three months |
|
|
|
| Secondary | QoL: The EQ-5D Questionnaire | Changes in quality of life perceived by EuroQoL (quality of life) Scale. It is an instrument which evaluates the quality of life measuring with one question: mobility, self-care, usual activities, pain/discomfort, and anxiety/depression. The answers given to these question (1 = no problems; 2 = some problems; 3 = extreme problems) are converted into EQ-5D index, an utility scores anchored at 0 for death and 1 for perfect health. The EQ-5D questionnaire also includes a Visual Analog Scale (VAS), by which respondents can report their perceived health status with a grade ranging from 0 (the worst possible health status) to 100 (the best possible health status). | Posted | Mean | Standard Deviation | score on a scale | Three months |
|
|
|
| Secondary | Nt-proBNP | Nt-proBNP levels | Posted | Mean | Standard Deviation | ng/L | Three months |
|
|
|
| Secondary | Right Ventricular Function | Fractional area change of right ventricle assessed by echocardiography | Posted | Mean | Standard Deviation | percentage of fractional area change | Three months |
|
|
|
| Secondary | CPET Exercise Time | Cardiopulmonary exercise time | Posted | Mean | Standard Deviation | seconds | Three months |
|
|
|
| 0 |
| 13 |
| 0 |
| 13 |
| 0 |
| 13 |
| EG001 | CONTROL GROUP | Patients underwent LVAD echo-optimization, but the optimal device speed is not confirmed at the end of procedure. | 0 | 14 | 0 | 14 | 0 | 14 |