A well-functioning vascular access is essential for providing adequate life-sustaining treatment in patients with end stage renal disease on maintenance hemodialysis. The preferred long-term vascular access is the arteriovenous fistula (AVF), which is created using the vessels of the patient by surgically connecting an artery with a superficial vein to increase the blood flow (Qa) in the venous system, which will dilate allowing the insertion of two needles, one to carry the blood to the dialyzer, and the other to return the cleansed blood to the body with the aid of a dialysis machine. Unfortunately, the high prevalence of vascular disease of the hemodialysis patients make difficult to create an adequate AVF in as many as 20 to 60% of the patients.In these persons, a valid alternative is the arteriovenous graft: in graft method an artery is surgically connected to a vein with a short piece of synthetic soft tube which is implanted under the skin. Needles are inserted in the graft during the dialysis treatment. Compared to an AV, however, graft is at higher risk of complications. The most frequent complication is thrombosis (i.e. the formation of blood clot inside the graft). Usually, thrombosis is the consequence of an underlying significant stenosis (i.e. a greater than 50% narrowing of the vessel or graft lumen by comparison with the lumen of a normal adjacent vessel or graft) and its hemodynamic consequences of decreasing the access blood flow (Qa) and/or increasing pressure within the graft. Therefore, all vascular access guidelines recommend regular noninvasive screening programs of grafts for timely identification of a stenosis associated with some type of functional or hemodynamic impairment, because its repair may prevent thrombosis and lengthen the useful life of the access. Screening methods include clinical monitoring and surveillance, which uses special equipment either to assess the hemodynamic consequences of stenosis by measuring Qa and static venous intra-access pressure ratio (VAPR) or to visualize the stenosis by means of duplex ultrasound (DU). Guidelines also state that there is insufficient evidence to prefer one method to another due to the lack of adequate comparative studies. The purpose of our study is to identify an optimal screening program for stenosis detection and elective repair by comparing the diagnostic performance for stenosis and incipient thrombosis of all the available screening tools in the same graft population
A well-functioning vascular access (a system that allows the blood of the patient to leave the body, be brought to the artificial kidney or dialyzer to be cleaned and return the cleansed blood to the body) is essential for providing adequate life-sustaining treatment in patients with end stage renal disease on maintenance haemodialysis.
The preferred long-term vascular access is the arteriovenous fistula (AVF), which is created using the vessels of the patient by surgically connecting an artery with a superficial vein to increase the blood flow (Qa) in the venous system, which after few weeks will dilate (maturation) allowing the insertion of two needles, one to carry the blood to the dialyzer (arterial needle), and the other to return the cleansed blood to the body (venous needle) with the aid of a dialysis machine, equipped with a blood pump and pressure sensors at the level of the "arterial" and "venous" needle to monitor for needles dislodgment.
Unfortunately, the high prevalence of vascular disease of the haemodialysis patients makes difficult to create an adequate AVF in as many as 20 to 60% of the patients.
In these persons, a valid alternative is the arteriovenous graft: in graft method an artery is surgically connected to a vein with a short piece of synthetic soft tube which is implanted under the skin. Needles are inserted in the graft during the dialysis treatment.
Compared to an AV, however, graft is at higher risk of complications. The most frequent and feared complication is thrombosis (i.e. the formation of blood clot inside the graft).
Usually, thrombosis is the consequence of an underlying stenosis (i.e. a greater than 50% narrowing of the vessel or graft lumen by comparison with the lumen of a normal adjacent vessel or graft due to the thickening of the vessels and/or the graft wall) and its hemodynamic consequences of decreasing the access blood flow (Qa) and/or increasing pressure within the graft. Less frequently thrombosis occurs in the absence of significant stenosis and may be triggered by a drop in the patient's arterial blood pressure (hypotension).
All vascular access guidelines recommend regular non-invasive screening programs of grafts for timely identification of a stenosis associated with some type of functional or hemodynamic impairment, because its repair may prevent thrombosis and lengthen the useful life of the access.
Non-invasive screening methods include clinical monitoring (i.e. noting signs of access dysfunction during the haemodialysis session) and surveillance, which uses special equipment either to assess the hemodynamic consequences of stenosis by measuring Qa and static venous intra-access pressure ratio (VAPR) or to visualize the stenosis by means of duplex ultrasound (DU).
In graft, all guidelines have opted for surveillance. However, they also state that there is insufficient evidence to prefer one technique to another because of the lack of adequate comparative studies.
Indeed, the great majority of the studies have evaluated the ability of detecting stenosis and predicting incipient thrombosis (within 1 to 3 months) by one surveillance method, and very few have compared two-to-three techniques at the best (1).
The purpose of the investigators was to compare in the same graft population the diagnostic performance of all the currently available screening tools in detecting stenosis and predicting incipient thrombosis (to identify the best criteria for elective stenosis repair).
The primary aim of the study was to identify an optimal screening program to reduce the risk of thrombosis in graft by assessing the diagnostic accuracy (i.e. sensitivity, false positive rate, positive and negative predictive value, and area under the receiver operator curve) of all the available screening tools and the occurrence of acute hypotension and identify the best threshold/s for continuous variables in a) detecting stenosis, b) predicting incipient thrombosis (within a 4-month period following graft assessment for stenosis).
The following screening methods will be tested:
Subsequently all grafts underwent the biplanar Digital Subtraction Angiography (DSA), our gold standard for stenosis detection. The access was visualized in its entirety from the feeding artery to the right atrium seeking for the presence of stenosis. Each stenosis was graded with regard to location and degree as outlined for DU and was considered significant (StA) when the lumen reduction was > 50%. To ensure blinding the investigator performing DSA was unaware of the results of the other screening strategies.
Secondary aims of the study were to assess:
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Graft stenosis and thrombosis | Diagnostic performance for stenosis at angiography of non invasive screening tools (duplex ultrasound, access blood flow (Qa), dynamic and static dialysis machine venous pressures, dynamic dialysis machine arterial pressure, and monitoring) and incipient thrombosis (within 4-month period) of the presence and degree of stenosis at angiography, non invasive screening techniques and acute hypotensive episode/s during the follow-up |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| Graft stenosis and thrombosis | Diagnostic Test |
|
| Measure | Description | Time Frame |
|---|---|---|
| Presence of a significant stenosis at DSA (yes/not) | presence of a greater than 50% reduction in the vessel or graft lumen diameter by comparison with the lumen diameter of a normal adjacent vessel or graft in mm/mm. | 1 hour |
| Presence of significant stenosis at DU (yes/not) | presence of a greater than 50% reduction in the vessel or graft lumen diameter by comparison with the lumen diameter of a normal adjacent vessel or graft in mm/mm and/or a peak systolic velocity > 400 cm/sec at the stenotic site | 30 minutes |
| Abnormal clinical monitoring (yes/not) | signs of graft dysfunction noted during dialysis: difficult cannulation, aspiration of clots, inability to achieve the prescribed dialysis pump blood flow (Qb), excessive post-dialysis bleeding or a >0.3 drop in single pool dialysis dose | 15 minutes |
| Qa measured by ultrasound dilution: QaU (ml/min) | Qa was measured by the Ultrasound dilution method during dialysis using the Transonic HD03 device, in the same dialysis session in which pressures were measured: each value is the mean of triplicate measurement | 15 minutes |
| Qa measured by DU: QaD (ml/min) | Measurement of Qa is made in a straight portion of the brachial artery in the mid-third of the upper arm. The diameter of the blood flow was measured directly on the vessel thanks to b-flow color technology. Sampling volume was placed in the centre of the lumen and in the longitudinal plane. Typically, measurements were obtained over a sequence of 3 to 5 cardiac cycles (to allow for time-averaged mean velocities, TAV). TAV was calculated directly by the device from a doppler spectral waveform by the duplex scanner system. The Qa (in ml/min) is calculated by the device as the product of the artery diameter and the TAV. The mean value of at least 3 separate measurements was reported. |
| Measure | Description | Time Frame |
|---|---|---|
| correlation coefficient (r) between QaU and QaD measurements (ml/min / ml/mn) | the correlation between the Qa values measured by ultrasound dilution and DU (ml of blood flowing within the access per minute, ml/min) | 30 minutes |
| Concordance for the presence of significant stenosis between two radiologists (yes/not) |
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Haemodialysis patients with an AV graft as vascular access
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Nicola Tessitore, MD | AO Verona | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| Azienda Ospedaliera Integrata di Verona - Policlinico Borgo Roma | Verona | 37135 | Italy |
| PubMed Identifier | Type | Citation | Retractions |
|---|---|---|---|
| 16813989 | Background | Vascular Access 2006 Work Group. Clinical practice guidelines for vascular access. Am J Kidney Dis. 2006 Jul;48 Suppl 1:S176-247. doi: 10.1053/j.ajkd.2006.04.029. No abstract available. | |
| 17507428 | Background | Tordoir J, Canaud B, Haage P, Konner K, Basci A, Fouque D, Kooman J, Martin-Malo A, Pedrini L, Pizzarelli F, Tattersall J, Vennegoor M, Wanner C, ter Wee P, Vanholder R. EBPG on Vascular Access. Nephrol Dial Transplant. 2007 May;22 Suppl 2:ii88-117. doi: 10.1093/ndt/gfm021. No abstract available. |
Not provided
Not provided
The following data will be available to other researchers from professor Albino Poli on request: patient age, gender, comorbidities and time on dialysis; graft age, site, configuration and previous interventions; results of angiography and screening tools (duplex ultrasound, dialysis dynamic and static venous pressures, dialysis arterial pressure/blood pump flow, ultrasound dilution access blood flow, monitoring) and outcomes during the follow-up (symptomatic hypotension, graft potency, thrombosis).
Not provided
one year from february 1st 2019
Not provided
Not provided
Not provided
| ID | Term |
|---|---|
| D001777 | Blood Coagulation |
| ID | Term |
|---|---|
| D006487 | Hemostasis |
| D001790 | Blood Physiological Phenomena |
| D002943 | Circulatory and Respiratory Physiological Phenomena |
Not provided
Not provided
Not provided
Not provided
Not provided
| 10 minutes |
| Dynamic arterial pressure / dialysis pump blood flow: dAP/Qb (mmHg/ml/min) | Dynamic arterial pressure (dAP) was measured in the initial 5 minutes of dialysis and detected by the dialysis machine using the pressure sensor connected with the "arterial" needle and expressed as the ratio with dialysis blood pump flow Qb. | 5 minutes |
| dynamic venous pressure: dVP (mmHg) | Dynamic venous pressure (dVP) was measured in the initial 5 minutes of dialysis and detected by the dialysis machine using the pressure sensor connected with the "venous" needle | 5 minutes |
| Derived static venous pressure ratio: VAPR (mmHg/mmHg) | obtained in the initial 5 minutes of dialysis by the dVP, Qb, haematocrit and systemic systolic and diastolic blood pressure values, according to literature in mmHg/mmHg | 10 minutes |
| occurrence of symptomatic acute hypotension during the follow up (yes/not) | during the follow up an episode of acute symptomatic hypotension in the intra- and inter-dialytic interval was recorded. Hypotension was defined as a sudden fall of systemic blood pressure associated with one or more of fainting palpitation, nausea, blurred vision, feeling weak or cold | 4 months |
the concordance of radiologist 1 and radiologist 2 in detecting the presence of significant stenosis at DSA. |
| 1 hour |
| intra-assay coefficient of variation of QaU and QaD (%) | the coefficient of variation was obtained by dividing the standard deviation by the mean value of multiple measurements | 15 minutes |
| inter-assay coefficient of variation of QaU, dAP/Qb,dVP and VAPR (%) | the coefficient of variation was obtained by dividing the standard deviation by the mean value of multiple measurements obtained over 1 week period for dAP/Qb, dVP and VAPR and within 1 month period for QaU | 1 month |
| 18713118 | Background | Polkinghorne K; Caring for Australians with Renal Impairment (CARI). The CARI guidelines. Vascular access surveillance. Nephrology (Carlton). 2008 Jul;13 Suppl 2:S1-11. doi: 10.1111/j.1440-1797.2008.00992.x. No abstract available. |
| 24817450 | Background | Tessitore N, Bedogna V, Verlato G, Poli A. Clinical access assessment. J Vasc Access. 2014;15 Suppl 7:S20-7. doi: 10.5301/jva.5000242. Epub 2014 Apr 12. |
| 12324911 | Background | Frinak S, Zasuwa G, Dunfee T, Besarab A, Yee J. Dynamic venous access pressure ratio test for hemodialysis access monitoring. Am J Kidney Dis. 2002 Oct;40(4):760-8. doi: 10.1053/ajkd.2002.35687. |
| 7564085 | Background | Krivitski NM. Theory and validation of access flow measurement by dilution technique during hemodialysis. Kidney Int. 1995 Jul;48(1):244-50. doi: 10.1038/ki.1995.290. |
| 24494667 | Background | Tessitore N, Bedogna V, Verlato G, Poli A. The rise and fall of access blood flow surveillance in arteriovenous fistulas. Semin Dial. 2014 Mar;27(2):108-18. doi: 10.1111/sdi.12187. Epub 2014 Feb 5. |
| 21454718 | Background | Tessitore N, Bedogna V, Melilli E, Millardi D, Mansueto G, Lipari G, Mantovani W, Baggio E, Poli A, Lupo A. In search of an optimal bedside screening program for arteriovenous fistula stenosis. Clin J Am Soc Nephrol. 2011 Apr;6(4):819-26. doi: 10.2215/CJN.06220710. Epub 2011 Mar 31. |