This study is a prospective, multicenter, randomized, blind, parallel and superiority test study. It is planned to select 206 cases of subjects with small coronary artery vessel disease who meet the inclusion/exclusion criteria. They are randomly divide them into Biolimus release coronary balloon catheter treatment group and plain old balloon angioplasty catheter (Powerline) treatment group according to the ratio of 1:1. All subjects accept clinical follow-up after operation, at 30 days, 6 months, 9 months and 12 months after operation. Follow-up with angiography is conducted at 9 months. The late lumen loss in diseased segment at 9 months after operation is the primary endpoint.
The study was a prospective, multicenter, randomized, blind, parallel positive control, and superiority test study, 206 subjects with small coronary artery disease who met the inclusion/exclusion criteria were enrolled, subjects were randomly assigned to the treatment group of Biolimus and Powerline at 1:1. All subjects underwent clinical follow-up at 30 days, 6 months, 9 months, and 12 months after surgery, and angiographic follow-up at 9 months. The primary endpoint was late lumen loss at 9 months.
The study will enroll 206 subjects. The sample size determination process is as follows:
Sample size is calculated according to the primary endpoint, late lumen loss (LLL) in diseased segment at 9 months after operation. Through literature review, a Meta analysis combined 11 randomized and control study shows that the late lumen loss level of control group (POBA) in the treatment of small vessel disease is 0.54mm. Another Meta analysis shows that LLL after PTCA operation is 0.57±0.57mm. Clinical investigators and statisticians, according to literature and clinical practice, suppose that the LLL level of test group (DCB) in this study can be lowered to 0.32mm and the combined standard deviation of LLL of the two groups is estimated to be 0.50mm conservatively.
The calculation formula of sample size for parallel design and superiority test is used. In case of taking 0.025 at singles side for inspection level, taking 80% for power (1-β), and taking 0 for superiority cut-off, conduct random grouping according to the ratio of 1:1, the require sample size of each group is calculated to be 82 cases. Considering the maximum failure rate 20% in angiography follow-up (as well as the early withdrawal and random damage caused by other reasons), the planned total sample size of both groups is 206 cases finally, including 103 cases for test group (DCB) and 103 cases for control group.
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Biolimus | Experimental | BA9 Drug-eluting Coronary Artery Balloon Catheter |
|
| Powerline | Active Comparator | Balloon dilated catheter |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| Biolimus | Device | 103 subjects with small coronary artery disease who met the inclusion/exclusion criteria were enrolled, subjects were assigned to the treatment group of Biolimus |
| Measure | Description | Time Frame |
|---|---|---|
| The late lumen loss | The late lumen loss in the lesion segment 9 months after surgery | 9 months |
Not provided
Not provided
General inclusion criteria
Radiography inclusion criteria
Exclusion Criteria:
General exclusion criteria
Radiography exclusion criteria
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Yaling Han | The General Hospital of Northern Theater Command | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| General Hospital of Northern Theater Command | Shenyang | Liaoning | 110016 | China |
| PubMed Identifier | Type | Citation | Retractions |
|---|---|---|---|
| 35738744 | Derived | Xu K, Fu G, Tong Q, Liu B, Han X, Zhang J, Ma G, Yang Q, Li H, Zhou Y, Jing Q, Li Y, Han Y. Biolimus-Coated Balloon in Small-Vessel Coronary Artery Disease: The BIO-RISE CHINA Study. JACC Cardiovasc Interv. 2022 Jun 27;15(12):1219-1226. doi: 10.1016/j.jcin.2022.03.024. Epub 2022 May 25. |
Not provided
Not provided
Not provided
| ID | Term |
|---|---|
| D003324 | Coronary Artery Disease |
| ID | Term |
|---|---|
| D003327 | Coronary Disease |
| D017202 | Myocardial Ischemia |
| D006331 | Heart Diseases |
| D002318 | Cardiovascular Diseases |
Not provided
Not provided
Diameter: 2.0mm, 2.25mm, 2.5mm, 2.75mm Length: 10mm, 15mm, 20mm, 25mm, 30mm
Not provided
Not provided
Not provided
| Powerline | Device | 103 subjects with small coronary artery disease who met the inclusion/exclusion criteria were enrolled, subjects were assigned to the treatment group of Powerline |
|
| D001161 |
| Arteriosclerosis |
| D001157 | Arterial Occlusive Diseases |
| D014652 | Vascular Diseases |