This study assesses the investigational sensor's (INVSENSOR00027) performance for detection of subjects' position and posture in bed, along with heart rate and respiratory rate.
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| INVSENSOR00027 Test group | Experimental | The subjects will be enrolled into the test group and will receive the INVSENSOR00027 investigational sensor. |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| INVSENSOR00027 | Device | Investigational, noninvasive sensor that will be placed on the subject's chest and/or back. INVSENSOR00027 measures signals regarding subject's movement and posture, heart rate, and respiratory rate. |
| Measure | Description | Time Frame |
|---|---|---|
| Sensitivity of INVSENSOR00027's Posture Change Detection | For each subject's position detection, INVSENSOR00027 output will be compared against observer's notes at each turn of the subject. Correctly detected position changes will be marked as True Positive. Missed detections (if any) will be counted. The sensitivity of INVSENSOR00027's posture change detection will be computed as Sensitivity = True Positive/(True Positive+Missed Detection) | 1-5 hours |
| Sensitivity of INVSENSOR00027's Activity Change Detection | For each subject's activity (e.g., walking and upright) detection, INVSENSOR00027 output will be compared against observer's notes. Correctly detected activity changes will be marked as True Positive. Missed detections (if any) will be counted. The sensitivity of INVSENSOR00027's activity change detection will be computed as Sensitivity = True Positive/(True Positive+Missed Detection). | 1-5 hours |
| Sensitivity of INVSENSOR00027's Fall Event Detection | 1-5 hours |
| Measure | Description | Time Frame |
|---|---|---|
| Respiratory Rate Detection | Accuracy will be determined by comparing the respiratory rate measurement of the INVSENSOR00027 to the respiratory rate reference (RRef) and calculating the arithmetic root mean square (Arms) value. In order to obtain the Arms value, the RR measurement from the reference is subtracted from the INVSENSOR00027 RR measurement for a number of samples. The average of this difference is computed as the bias. The standard deviation of the differences is computed as the precision. The square root of the sum of the squares of bias and precision is computed as the Arms value. |
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Ahmed Alghazi | Masimo Corporation | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| Masimo Clinical Lab | Irvine | California | 92618 | United States |
Not provided
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| FG000 | INVSENSOR00027 Test Group | The subjects will be enrolled into the test group and will receive the INVSENSOR00027 investigational sensor. INVSENSOR00027: Investigational, noninvasive sensor that will be placed on the subject's chest and/or back. INVSENSOR00027 measures signals regarding subject's movement and posture, heart rate, and respiratory rate. |
| Title | Milestones | Reasons Not Completed | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Overall Study |
|
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| BG000 | INVSENSOR00027 Test Group | The subjects will be enrolled into the test group and will receive the INVSENSOR00027 investigational sensor. INVSENSOR00027: Investigational, noninvasive sensor that will be placed on the subject's chest and/or back. INVSENSOR00027 measures signals regarding subject's movement and posture, heart rate, and respiratory rate. |
| Units | Counts |
|---|---|
| Participants |
|
| Title | Description | Population Description | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Denominator Units Selected | Denominators | Classes |
|---|---|---|---|---|---|---|---|---|---|
| Age, Categorical | Count of Participants |
| Type | Title | Description | Population Description | Reporting Status | Anticipated Posting Date | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Time Frame | Units Analyzed | Denominator Units Selected | Arm/Group Information | Denominators | Classes | Analyses | ||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Primary | Sensitivity of INVSENSOR00027's Posture Change Detection | For each subject's position detection, INVSENSOR00027 output will be compared against observer's notes at each turn of the subject. Correctly detected position changes will be marked as True Positive. Missed detections (if any) will be counted. The sensitivity of INVSENSOR00027's posture change detection will be computed as Sensitivity = True Positive/(True Positive+Missed Detection) | Posted | Number | percent sensitivity for posture change | 1-5 hours |
|
Adverse event data was collected over the course of the study (35 days).
Not provided
Not provided
| ID | Title | Description | Deaths (Affected) | Deaths (At Risk) | Serious Events (Affected) | Serious Events (At Risk) | Other Events (Affected) | Other Events (At Risk) |
|---|---|---|---|---|---|---|---|---|
| EG000 | INVSENSOR00027 Test Group | The subjects will be enrolled into the test group and will receive the INVSENSOR00027 investigational sensor. INVSENSOR00027: Investigational, noninvasive sensor that will be placed on the subject's chest and/or back. INVSENSOR00027 measures signals regarding subject's movement and posture, heart rate, and respiratory rate. |
Not provided
Not provided
Not provided
| Title | Organization | Phone | Extension | |
|---|---|---|---|---|
| Ahmed Alghazi | Masimo | 949-297-7000 | studies@masimo.com |
Not provided
| Type | Includes Protocol | Includes SAP | Includes ICF | Document Label | Document Date | Document Uploaded Date | Document File Name |
|---|---|---|---|---|---|---|---|
| Prot_SAP | Yes | Yes | No | Study Protocol and Statistical Analysis Plan | Aug 7, 2018 | May 15, 2020 | Prot_SAP_000.pdf |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| 1-5 hours |
| Participants |
|
| Sex: Female, Male | Count of Participants | Participants |
|
| Race/Ethnicity, Customized | Count of Participants | Participants |
|
| Region of Enrollment | Number | participants |
|
| Units | Counts |
|---|---|
| Participants |
|
|
| Primary | Sensitivity of INVSENSOR00027's Activity Change Detection | For each subject's activity (e.g., walking and upright) detection, INVSENSOR00027 output will be compared against observer's notes. Correctly detected activity changes will be marked as True Positive. Missed detections (if any) will be counted. The sensitivity of INVSENSOR00027's activity change detection will be computed as Sensitivity = True Positive/(True Positive+Missed Detection). | Posted | Number | percent sensitivity for posture change | 1-5 hours |
|
|
|
| Primary | Sensitivity of INVSENSOR00027's Fall Event Detection | Fall detection was an optional event per protocol. | Posted | Number | percent sensitivity for fall detection | 1-5 hours |
|
|
|
| Secondary | Respiratory Rate Detection | Accuracy will be determined by comparing the respiratory rate measurement of the INVSENSOR00027 to the respiratory rate reference (RRef) and calculating the arithmetic root mean square (Arms) value. In order to obtain the Arms value, the RR measurement from the reference is subtracted from the INVSENSOR00027 RR measurement for a number of samples. The average of this difference is computed as the bias. The standard deviation of the differences is computed as the precision. The square root of the sum of the squares of bias and precision is computed as the Arms value. | 2 subjects were excluded due to data collection issues. | Posted | Number | A(RMS) of breaths per minute | 1-5 hours |
|
|
|