SureCRIC is intended to be used as an accessory to cricothyrotomy and tracheotomy devices, helping to identify and stabilize anatomical landmarks for establishing an airway.
Cricothyroid membrane identification accuracy has been reported to be 30-62% in the literature [Bai15, Ell10, Kri15, Lam15]. The current study assesses SureCRIC-assisted medic performance in identifying the cricothyroid membrane in a diverse standardized patient population.
The study is a two-arm, open-label study involving users - largely licensed paramedics or combat medics - identifying the cricothyroid membranes on healthy volunteers. The Control Group will perform cricothyroid membrane identification on healthy male and female volunteers of varying heights, weights, and BMI's with the standard, freehand approach. The Experimental Group will perform cricothyroid membrane identification on the volunteers with the help of the SureCRIC. An expert verifier (senior otolaryngologist) will verify correct cricothyroid membrane identification and will not be blinded in order to reduce any error that may be introduced by SureCRIC removal and subsequent skin movement. The primary endpoint is a pass/fail assessment for cricothyroid membrane identification. The main secondary endpoint is time to cricothyroid membrane identification. The study design was recommended by FDA. The study, per FDA guidance, is not designed for statistical comparisons and only descriptive statistics are provided.
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| SureCRIC | Experimental | SureCRIC-aided cricothyroid membrane identification |
|
| Freehand | Active Comparator | Freehand cricothyroid membrane identification |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| SureCRIC | Device | SureCRIC is intended to be used as an accessory to cricothyrotomy and tracheotomy devices, helping to identify and stabilize anatomical landmarks for establishing an airway. |
| Measure | Description | Time Frame |
|---|---|---|
| Number of Paramedics/Combat Medics Who Accurately Identified the Cricothyroid Membrane on Standardized Patients of Different Sizes | pass/fail assessment for CTM identification as gauged by expert (senior orolaryngologist) | 1 day |
| Measure | Description | Time Frame |
|---|---|---|
| Time to Cricothyroid Membrane Identification | time to cricothyroid membrane identification as measured by direct observation | 1 day |
Not provided
Participants (Users):
Inclusion Criteria:
Exclusion Criteria:
Healthy Volunteers (Standardized Patients):
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| National Center for Human Factors in Healthcare | Washington D.C. | District of Columbia | 20008 | United States |
| PubMed Identifier | Type | Citation | Retractions |
|---|---|---|---|
| 26198864 | Background | Bair AE, Chima R. The inaccuracy of using landmark techniques for cricothyroid membrane identification: a comparison of three techniques. Acad Emerg Med. 2015 Aug;22(8):908-14. doi: 10.1111/acem.12732. Epub 2015 Jul 21. | |
| 20645953 | Background | Elliott DS, Baker PA, Scott MR, Birch CW, Thompson JM. Accuracy of surface landmark identification for cannula cricothyroidotomy. Anaesthesia. 2010 Sep;65(9):889-94. doi: 10.1111/j.1365-2044.2010.06425.x. |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
We aimed to include 2 groups of 10 participants. One group for the SureCRIC arm and one group for the freehand arm. 12 participants/group were recruited to fill this quota. Ultimately, 10 participants were included in the SureCRIC arm and 8 in the freehand arm due to no-shows/cancellations. The 18 participants represented novice target end users: paramedics & military medics. Each of the 18 participants tested on 6 healthy standardized patients: M's & F's of <10th, 50th & >90th %ile heights.
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| FG000 | SureCRIC | SureCRIC-aided cricothyroid membrane identification SureCRIC: SureCRIC is intended to be used as an accessory to cricothyrotomy and tracheotomy devices, helping to identify and stabilize anatomical landmarks for establishing an airway. |
| FG001 | Freehand | Freehand cricothyroid membrane identification Freehand: Freehand palpation approach to identification of the cricothyroid membrane |
| FG002 | Standardized Patients | Healthy males and females representing the <10th, 50th and >90th percentile heights. |
| Title | Milestones | Reasons Not Completed | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Overall Study |
|
Participants: Exposure to human cadaver practice (10% of SureCRIC participants vs. 50% of freehand participants); real-world procedural performance (0% of SureCRIC participants vs. 37.5% of freehand participants).
Standardized Patients: Healthy volunteers.
Not provided
| ID | Title | Description |
|---|---|---|
| BG000 | SureCRIC | SureCRIC-aided cricothyroid membrane identification SureCRIC: SureCRIC is intended to be used as an accessory to cricothyrotomy and tracheotomy devices, helping to identify and stabilize anatomical landmarks for establishing an airway. |
| BG001 | Freehand |
| Units | Counts |
|---|---|
| Participants |
|
| Title | Description | Population Description | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Denominator Units Selected | Denominators | Classes |
|---|---|---|---|---|---|---|---|---|---|
| Age, Customized | Count of Participants |
| Type | Title | Description | Population Description | Reporting Status | Anticipated Posting Date | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Time Frame | Units Analyzed | Denominator Units Selected | Arm/Group Information | Denominators | Classes | Analyses | ||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Primary | Number of Paramedics/Combat Medics Who Accurately Identified the Cricothyroid Membrane on Standardized Patients of Different Sizes | pass/fail assessment for CTM identification as gauged by expert (senior orolaryngologist) | paramedic or combat medic identified cricothyroid membranes on standardized patients as judged by an expert otolaryngologist | Posted | Count of Participants | Participants | 1 day |
|
1day
standardized patient bruising, injury
Not provided
| ID | Title | Description | Deaths (Affected) | Deaths (At Risk) | Serious Events (Affected) | Serious Events (At Risk) | Other Events (Affected) | Other Events (At Risk) |
|---|---|---|---|---|---|---|---|---|
| EG000 | SureCRIC | SureCRIC-aided cricothyroid membrane identification SureCRIC: SureCRIC is intended to be used as an accessory to cricothyrotomy and tracheotomy devices, helping to identify and stabilize anatomical landmarks for establishing an airway. |
Not provided
| Term | Organ System | Source Vocabulary | Assessment Type | Notes | Statistical Information |
|---|---|---|---|---|---|
| injury including bruising | Skin and subcutaneous tissue disorders | Systematic Assessment | injury from tools, devices, palpation etc. to either the participant (paramedic/combat medic) or standardized patient |
Not provided
| Title | Organization | Phone | Extension | |
|---|---|---|---|---|
| Chief Medical Officer | InnoVital Systems | 2022862328 | amit@innovitalsystems.com |
Not provided
| Type | Includes Protocol | Includes SAP | Includes ICF | Document Label | Document Date | Document Uploaded Date | Document File Name |
|---|---|---|---|---|---|---|---|
| Prot_SAP | Yes | Yes | No | Study Protocol and Statistical Analysis Plan | Sep 15, 2017 | Oct 26, 2021 | Prot_SAP_001.pdf |
Not provided
Not provided
Not provided
| ID | Term |
|---|---|
| D000402 | Airway Obstruction |
| ID | Term |
|---|---|
| D012131 | Respiratory Insufficiency |
| D012120 | Respiration Disorders |
| D012140 | Respiratory Tract Diseases |
Not provided
Not provided
The study is a two-arm, open-label study involving users - largely licensed paramedics or combat medics - identifying the cricothyroid membranes on healthy volunteers. The Control Group will perform cricothyroid membrane identification on healthy male and female volunteers of varying heights, weights, and BMI's with the standard, freehand approach. The Experimental Group will perform cricothyroid membrane identification on the volunteers with the help of the SureCRIC.
Not provided
Not provided
Not provided
Not provided
| Freehand | Other | Freehand palpation approach to identification of the cricothyroid membrane |
|
| 25991744 | Background | Kristensen MS, Teoh WH, Rudolph SS, Tvede MF, Hesselfeldt R, Borglum J, Lohse T, Hansen LN. Structured approach to ultrasound-guided identification of the cricothyroid membrane: a randomized comparison with the palpation method in the morbidly obese. Br J Anaesth. 2015 Jun;114(6):1003-4. doi: 10.1093/bja/aev123. No abstract available. |
| 25637060 | Background | Lamb A, Zhang J, Hung O, Flemming B, Mullen T, Bissell MB, Arseneau I. Accuracy of identifying the cricothyroid membrane by anesthesia trainees and staff in a Canadian institution. Can J Anaesth. 2015 May;62(5):495-503. doi: 10.1007/s12630-015-0326-y. Epub 2015 Jan 31. |
Freehand cricothyroid membrane identification Freehand: Freehand palpation approach to identification of the cricothyroid membrane |
| BG002 | Standardized Patients | Healthy male and female volunteers corresponding to the 5th, 50th and 95th percentile heights. |
| BG003 | Total | Total of all reporting groups |
| Participants |
|
| Sex/Gender, Customized | Count of Participants | Participants |
|
| Race and Ethnicity Not Collected | Race and Ethnicity were not collected from any participant. | Count of Participants | Participants |
|
| First-hand, real-world human performance | Count of Participants | Participants |
|
| Human cadaver training | Count of Participants | Participants |
|
Freehand cricothyroid membrane identification
Freehand: Freehand palpation approach to identification of the cricothyroid membrane
| OG002 | Standardized Patient | Healthy volunteers - males and females corresponding to the 5th, 50th, and 95th percentile heights |
|
|
| Secondary | Time to Cricothyroid Membrane Identification | time to cricothyroid membrane identification as measured by direct observation | paramedic or combat medic | Posted | Mean | Standard Deviation | seconds | 1 day |
|
|
|
| 0 |
| 10 |
| 0 |
| 10 |
| 0 |
| 10 |
| EG001 | Freehand | Freehand cricothyroid membrane identification Freehand: Freehand palpation approach to identification of the cricothyroid membrane | 0 | 6 | 0 | 8 | 0 | 8 |
| EG002 | Standardized Patients | Healthy volunteers corresponding to the <10th, 50th and >90th percentile heights. | 0 | 6 | 0 | 6 | 0 | 6 |
|
Not provided
Not provided
| Title | Measurements |
|---|---|
|
| Title | Measurements |
|---|---|
|