This is an observational, cohort-based, single-site, prospective study conducted in accordance with ISO 14155-1:2003 and ISO 14155-1:2011. The primary objectives are to evaluate the accuracy of the device's measurements/readings in comparison to a gold standard or standard of care patient monitor.
This is an observational, cohort-based, single-site, prospective clinical study enrolling up to 100 patients who are hospitalized in the internal ward. There is no control group.
The primary objectives of the study are:
Evaluate the body temperature measured by the ear unit and validate its accuracy.
Evaluate the accuracy of SPO2 signals and results by comparing the readings to a gold standard device with a finger sensor.
Evaluate the ECG and respiration signals recorded by the patch device by comparing them to the signals recorded by a patient monitor with standard leads.
The secondary objectives of the study are:
Usability - evaluate the ease of use of the device by medical staff and patients (e.g., positioning, activation, replacement of patch, understanding processes and requirements).
Evaluate the signals quality and the patch adherence after 5 - 7 days on the chest.
The study will assess the safety and effectiveness of the Vital Signs Monitoring System (VSMS) for its intended use/proposed indications. Data will be collected while the VSMS is adhered to each patient for the entire hospitalization period, up to 7 days. There is no follow-up required.
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| VSMS with ear unit | application of Vital Signs Monitoring System with ear unit for 6-10 days during hospitalization (throughout their admission period) |
| |
| VSMS without ear unit | application of Vital Signs Monitoring System without ear unit for 6-10 days during hospitalization (throughout their admission period) |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| Vital Signs Monitoring System | Device | The VSMS will be adhered to each enrolled patient for the entire hospitalization period. |
|
| Measure | Description | Time Frame |
|---|---|---|
| Skin Temperature measured in units of Celsius/Fahrenheit | Validation of measurement accuracy of skin temperature as compared to gold standard. | 6-10 days |
| Oxygen Saturation measured in units of percentage peripheral saturated oxygen | Validation of measurement accuracy of peripheral oxygen saturation as compared to gold standard. | 6-10 days |
| .ECG signal recording and Respiration measured in units heart beat per minute | Assess quality of device's recorded ECG and respiration signals as compared to patient monitor with standard leads | 6-10 days |
| Measure | Description | Time Frame |
|---|---|---|
| Quality of recorded signal | Evaluate the quality of recorded signals in time by comparing signal of day 1 with signal in last day of adherence of the patch | after 5-7 days of device usage |
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Not provided
Hospitalized patients of both genders, above 18 years of age, with ASA physical status classified as 1-4. In order to obtain the necessary results, the following shall also define the patient population:
Temperature: 30-50% of subjects shall be febrile, i.e., having:
SPO2: 10-20% of patients shall have a blood oxygen saturation level of 70-90%.
ECG / Respiration: At least 20% of patients shall have some type of arrhythmia.
| Name | Role | Phone | Extension | |
|---|---|---|---|---|
| Anat Alon | Contact | 972 8-977-9999 | alona@asaf.health.gov.il | |
| Peter Spiegel | Contact | psiegel@asaf.health.gov.il |
| Name | Affiliation | Role |
|---|---|---|
| Moshe Tishler, Professor | Assaf Harofeh Hospital/Medical Center | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| Internal Ward B, Aliza Begin Building, Assaf Harofe Hospital (2nd Floor) | Recruiting | P.O. Beer Ya'akov, Zerifin | 70300 | Israel |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| ID | Term |
|---|---|
| D053120 | Respiratory Aspiration |
| ID | Term |
|---|---|
| D012120 | Respiration Disorders |
| D012140 | Respiratory Tract Diseases |
| D010335 | Pathologic Processes |
| D013568 | Pathological Conditions, Signs and Symptoms |
Not provided
Not provided
Not provided
Not provided
Not provided