The aim of this randomized study is conducted is to evaluate the performance and safety of air-Q intubating laryngeal airway as primary airway device in various short surgical procedure in children compared to Ambu® AuraGain™ Laryngeal Mask. Current Supraglottic airway devices (SADs) available in are classic LMA (cLMA), proseal LMA (pLMA), supreme LMA (sLMA) and the newer is Ambu® AuraGain™laryngeal mask. As variety of newer SADs for children have emerged since their introduction in clinical practice, hope the outcomes of this study it help advancing our knowledge and acumen in selecting appropriate devices for paediatric population.
JUSTIFICATION OF STUDY
In view of no clinical data is available evaluating the clinical performance and safety of air-Q and Ambu® AuraGain™ in children, therefore this prospective and randomized study is conducted. Most of previous randomized studies in term of safety did not compare the hemodynamic stability.Hope the outcomes of this study it help advancing our knowledge and acumen in selecting devices for paediatric population.
RESEARCH HYPOTHESES
METHODOLOGY
This is a prospective, single blinded and randomized controlled trial study will be conducted at Operation Theatre (OT), Hospital University Sains Malaysia (HUSM). involving paediatric patients scheduled for various surgical procedures within 2 hours where supraglottic airway management would be appropriate..
Sample size calculation For objective 1, sample size was calculated using power and sample size software and the data based on previous study by V.Darlong et al (Comparison of air-Q and Ambu Aura-i for controlled ventilation in infants: a randomized controlled trial). In this study, the time for successful insertion for air-Q is 16.53±1.53 and Ambu Aura-I is 14.68±2.83. Considering power of 80% and type error-I α of 5%, the sample size was required is 15 participants in each group. 10% is added for drop out sample. Therefore sample size, n 15 + (0.1x15) = 17 participants for each group.
For objective 2, sample size was calculated using power and sample size software and the data based on previous study by V.Darlong et al (Comparison of air-Q and Ambu Aura-i for controlled ventilation in infants: a randomized controlled trial). In this study, the OLP for air-Q is 20.21±4.62 and Ambu Aura-I is 16.21±5.66. Considering power of 80% and type error-I α of 5%, the sample size was required is 22 participants in each group. 10% is added for drop out sample. Therefore sample size, n 22 + (0.1x22) = 25 participants for each group.
Overall sample size is estimated using objective 2 because of larger sample size. Total sample size is 50 patients.
SAMPLING METHOD Patients are randomly assigned into 2 groups: air-Q (A) and Ambu® AuraGain™ (B) using computer-generated randomization. Sequentially-numbered sealed opaque envelopes containing group allocation as determined by computer-generated sequence, will be opened by a nurse in charge for study device preparation.
RECRUITMENT OF SUBJECT & INFORMED CONSENT SEEKING Paediatric patients who meets the study criteria, their parents will be approached to invite patients participate in this study. Each parent will be given a thorough explanation along with copy of Patient Information Sheet. All questions that the participant's parents have, will be answered to their satisfaction before consent been obtained by signing the consent form.
The study is divided into screening, pre-operative, intra-operative and post-operative period.
Screening period After obtaining approval from the Ethics committee, participants will be selected according to inclusion and exclusion criteria during preoperative assessment round. Study procedures will be explained and written consent will be obtained from the participant's parents.
Pre-operative period In all participants, a eutectic mixture of local anaesthetic cream (EMLA) will be applied on both hands 30 minutes before transfer to operation theatre. No premedication will be prescribed in the morning of the surgery. All consented participants will be randomized using computer generated randomization into two groups: Group a (air-Q) and group B (Ambu Aura Gain). All demographic data including types and duration of surgery and diagnosis of procedure is recorded.
Intra-operative Anaesthesia Technique In operation theatre, all of the participants will be monitored with non-invasive blood pressure (NIBP), pulse oximeter (SpO2), electrocardiogram (ECG) and capnography (EtCO2). Anesthesia is induced with sevoflurane in oxygen. The anesthesia plane is gradually deepened by increasing the inspired concentration of sevoflurane (2-8%) till the loss of eyelash reflex. An intravenous (IV) cannula is inserted and intravenous fentanyl 1mcg/kg and rocuronium 0.6mg/kg are administered. Use of neuromuscular blocking agent is to minimise the risk of reflex activation of the airway. After 3 minutes, the appropriate size (depending upon body weight) of SADs (according to the group allocated based on the sequence of computerized randomization) lubricated with water soluble lubricant (KY jelly) is inserted by the investigator by standard technique as per manufacture's recommendation. Then the cuff is inflated and intracuff pressure is then standardized to 60 cmH20 using a digital cuff pressure monitor. Data and timing are collected by independent observers. To prevent biasedness in conducting the study, the same investigator will insert the device for all participants.
Haemodynamic parameters such as blood pressure (BP), heart rate (HR), oxygen saturation (SpO2) and end tidal carbon dioxide (EtCO2) is recorded at different interval namely, pre induction, post induction, post insertion, every 2 minutes for 10 minutes and then 5 minutes thereafter until extubation.
Anesthesia is maintained with sevoflurane (Minimum Alveolar Concentration value of 1.0-1.2) in Oxygen: Air mixture of 50% . Pressure controlled ventilation is set and targeting the minute ventilation to achieve normocarbia (EtCO2 35-40 mmHg). Intra-operative analgesia is supplement by suppository paracetamol 20 mg/kg or diclofenac sodium 1mg/kg.
Measurement of parameters
Post-Operative At the end of surgery, sevoflurane is turned off and 100% oxygen is administered. Muscle relaxant is reversed by neostigmine (50 mcg/kg) and glycopyrrolate (10 mcg/kg). With adequate tidal volume and respiratory rate, oropharyngeal suctioning is done and the device is removed. On removal of the device, any presence of blood stain or tongue-lip-dental trauma is recorded. In the post-operative period, the participant is observed for any complications such as airway reflex activation (laryngospasm, bronchospasm or severe cough), oxygen desaturation (spO2 <90%) and vomiting/aspiration.
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Ambu Aura Gain | Experimental | Ambu Aura Gain supraglottic airway size 2 or 2.5 will be inserted once patients are paralyzed |
|
| Air-Q | Active Comparator | Air-Q supraglottic airway size 1.5 and 2.0will be inserted once the patients are paralyzed |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| Ambu Aura Gain supraglottic airway | Device | Anesthesia is induced with sevoflurane in oxygen. The anesthesia plane is gradually deepened by increasing the inspired concentration of sevoflurane (2-8%) till the loss of eyelash reflex. After intravenous (IV) cannula is inserted and intravenous fentanyl 1mcg/kg and rocuronium 0.6mg/kg are administered. Use of neuromuscular blocking agent is to minimize the risk of reflex activation of the airway. After 3 minutes, the appropriate size (depending upon body weight) of Ambu Aura Gain is lubricated with water soluble lubricant (KY jelly) is inserted by investigator based on manufacturer recommendation with the patient's head in neutral position. The cuff then is inflated to an intra-cuff pressure of 40 cm H20, measured using a digital cuff pressure monitor. |
| Measure | Description | Time Frame |
|---|---|---|
| measure fibreoptic (FO) grade of laryngeal view | Flexible fibreoptic scope is use to view anatomical alignment of the device to the larynx, 1cm proximal to the ventilating orifice | 5 minutes after insertion of the device when good tidal volume is achieved |
| Measure | Description | Time Frame |
|---|---|---|
| oropharyngeal leak pressure (OLP) | OLP is recorded by observing the peak airway pressure which audible leak occur at the first time when fresh gas flow delivered at 3 litre/min and the expiratory valve is completely closed | 5 minutes after insertion of the device when good tidal volume is achieved |
| number of attempt |
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Rhendra Hardy Mohamad Zaini, MD | Universiti of Science Malaysia | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| University of Science Malaysia Hospital | Kubang Kerian | Kelantan | 16150 | Malaysia |
| PubMed Identifier | Type | Citation | Retractions |
|---|---|---|---|
| 25917434 | Result | Darlong V, Biyani G, Baidya DK, Pandey R, Chandralekha, Punj J, Upadhyay AD. Comparison of air-Q and Ambu Aura-i for controlled ventilation in infants: a randomized controlled trial. Paediatr Anaesth. 2015 Aug;25(8):795-800. doi: 10.1111/pan.12663. Epub 2015 Apr 27. | |
| 24961960 | Result | Darlong V, Biyani G, Pandey R, Baidya DK, Punj Ca. Comparison of performance and efficacy of air-Q intubating laryngeal airway and flexible laryngeal mask airway in anesthetized and paralyzed infants and children. Paediatr Anaesth. 2014 Oct;24(10):1066-71. doi: 10.1111/pan.12462. |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
|
| Air-Q supraglottic airway | Device | Anesthesia is induced with sevoflurane in oxygen. The anesthesia plane is gradually deepened by increasing the inspired concentration of sevoflurane (2-8%) till the loss of eyelash reflex. After intravenous (IV) cannula is inserted and intravenous fentanyl 1mcg/kg and rocuronium 0.6mg/kg are administered. Use of neuromuscular blocking agent is to minimize the risk of reflex activation of the airway. After 3 minutes, the appropriate size (depending upon body weight) of Air-Q supraglottic airway is lubricated with water soluble lubricant (KY jelly) is inserted by investigator based on manufacturer recommendation with the patient's head in neutral position. The cuff then is inflated to an intra-cuff pressure of 40 cm H20, measured using a digital cuff pressure monitor. |
|
Measure number of attempts of inserting the devices |
| Throughout the insertion process up to 5 minutes post induction |
| Time of insertion | Measure the time of successful insertion of the devices | The time from which the device first being hold until successful insertion up to 5 minutes post induction |
| Blood pressure | Documented the blood pressure (BP) is recorded at different time interval | Time from pre-induction until 5 minutes post insertion |
| Heart rate | Recorded the heart rate (HR) is recorded at different interval | Time from pre-induction until 5 minutes post insertion |
| Oxygen saturation (SpO2) | Oxygen saturation (SpO2) is recorded at different interval | Time from pre-induction until 5 minutes post insertion |
| end tidal Carbon dioxide (EtCO2) | The EtCO2 is recorded at different time interval | Time from pre-induction until 5 minutes post insertion |
| numbers of adverse effects of the devices as assessed by clinical observation | Presence of adverse effect of the device postextubation | Throughout the stay in recovery bay within 30 minutes after operation |
| 26648173 | Result | Jagannathan N, Hajduk J, Sohn L, Huang A, Sawardekar A, Gebhardt ER, Johnson K, De Oliveira GS. A randomised comparison of the Ambu(R) AuraGain and the LMA(R) supreme in infants and children. Anaesthesia. 2016 Feb;71(2):205-12. doi: 10.1111/anae.13330. Epub 2015 Dec 9. |
| 23519724 | Result | Whyte SD, Cooke E, Malherbe S. Usability and performance characteristics of the pediatric air-Q(R) intubating laryngeal airway. Can J Anaesth. 2013 Jun;60(6):557-63. doi: 10.1007/s12630-013-9918-6. Epub 2013 Mar 22. |