The goal of the study is to test the capability of the CareTaker monitor to calibrate its continuous blood pressure readings independently. Currently the device requires another approved blood pressure monitor to provide a starting calibration.
The new control module enables the CareTaker to perform a pressure sweep of the internal pressure in the finger cuff. The resulting data is analyzed using a combination of pulse analysis/oscillometry approach.
The specific objective of this project was to demonstrate the CareTaker's ability to perform self-calibration to the same performance standard as its predicate device. That device demonstrated efficacy by showing substantial equivalence to the performance of classical sphygmomanometry using a Riva-Rocci/Korotkoff (RRK) measurement on the brachial artery, meeting a bias of at least 5 mmHg and a standard deviation of 8 mmHg across 3 paired readings from at least 85 patients, for a total of at least 255 paired readings.
The study was performed at the University of Virginia Medical Center in Charlottesville, Virginia. UVA's Institutional Review Board approved (#18686) and supervised the study for subjects who were University of Virginia hospital patients and staff > 18 years of age and who were able to give verbal consent.
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| non-invasive blood pressure monitor | Device | Equivalence of CareTaker blood pressure to classical sphygmomanometry |
| Measure | Description | Time Frame |
|---|---|---|
| Performance comparison | Comparison with performance of classical sphygmomanometry using a Riva-Rocci/Korotkov (RRK) measurement on the brachial artery, meeting a bias of at least 5 mmHg and a standard deviation of 8 mmHg | 8 minutes for 3 paired readings |
Not provided
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Subjects will be UVA patients/staff, > 18 years of age, who are able and willing to participate and have given verbal assent
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Martin C Baruch, PhD | CareTaker Medical LLC | Principal Investigator |